7.2.1
Additionally for Ambistik CPET
1.
Start the software.
For this:
Start the PC and press the
"BLUE CHERRY" symbol on the
desktop.
The Ambistik CPET with the serial
number and the measured CO
2
concentration of the ambient air is
displayed at the bottom of the
BLUE CHERRY screen [1].
(Example data: 0.096 %).
Therefore, the CO
2
concentration
does not have to be entered
manually for the correct
calibration of the CO
2
measurement.
2.
Open O
2
/ CO
2
calibration.
For this:
Press the "O
2
/ CO
2
Calibration"
button [2] in the left menu line
located above the "Close" button.
3.
Check the O
2
/ CO
2
concentration.
For this:
Press the button "O
2
/ CO
2
Conc." [3] on the right side.
The CO
2
concentration (sample
data 0.09 %) has already been
transferred [4] and the field is
grey.