![GERATHERM 979119 Instructions For Use Manual Download Page 55](http://html1.mh-extra.com/html/geratherm/979119/979119_instructions-for-use-manual_2216238055.webp)
Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Version: 7 | Release date: 02. August 2021
Page 55
7
Operating Instructions
7.1
Checking for Worn Parts
The Ergostik should be checked for defective wearing parts
each day before beginning treatments.
You can get a description in:
•
Chap. 8 “Servicing / Maintenance”
And:
Possible danger to life.
Reason: Misdiagnosis caused by measurement error
due to incorrect components or improper use.
Therefore:
•
Replace the flow sensor and the mouthpiece in
case of a functional error!
•
Always follow instructions (see chap. 8 "Servicing
/ Maintenance" and chap. 9 "Cleaning and
Disinfection")!
7.1.1
Minifilter / Permapure Check
The permapure drying tube and
minifilter must be checked regularly
for discoloration.
Please note that the permapure tube
and minifilter are transparent as
shown in Fig. [1] and are not
severely discoloured as shown in
Fig. [2].
If the permapure tube turns brown
during the course of use, it should
be replaced together with the
minifilter.
1