Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Version: 7 | Release date: 02. August 2021
Page 67
9
Cleaning and Disinfection
The devices of the manufacturer were designed in such a way
that minimal effort is required for cleaning and disinfection.
This is why just a few tasks are necessary to keep Ergostik
functional and clean.
Cleaning and disinfection may cause discolouration of the
components, but without impairing their function.
The following intervals apply:
Component
Interval
Method
Noseclip
After each patient
Dispose of!
Silicone mouthpiece
After each patient
Cleaning and
disinfection bath
Face mask
After each patient
Cleaning and
disinfection bath
Mask adapter
After each patient
Cleaning and
disinfection bath
Network mask
After each patient
Cleaning and
disinfection bath
Reusable flow sensor,
Ergoflow
After each patient
Cleaning and
disinfection bath
Ergostik
Weekly
Wipe / surface
disinfection
Ergoclip
After each patient
Wipe / surface
disinfection
All other touchable parts
Weekly
Wipe disinfection
Possible physical injury.
Reason: misinterpretation of obvious measuring errors
(drift) caused by liquid in the tubes. Therefore:
•
Clean tubes only externally!
•
Replace dirty tubes!