Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Version: 7 | Release date: 02. August 2021
Page 43
4.1.2.2
Sensors
Flow Sensor
The heart of the Ergostik for precise flow measurement is the
Ergoflow flow sensor. It must be calibrated after every cleaning
and before the first examination of a day.
As the flow sensor ensures the precise flow measurements, the
cleaning and calibration requirements must absolutely be
adhered to, regardless of the version selected.
Attention
Ergostik could be damaged
Reason: Sensors are precise and high-resolution
components.
For this: Ensure the respective functions are preserved!
Therefore:
•
Exactly follow instructions for cleaning, disinfection and
calibration!
You will find a detailed description in:
•
Chap. 9 "Cleaning and Disinfection"
•
The separate IFU Volume Calibration