Instructions for Use
Spirostik
Serial no.: xx|8|001|yyy and 2001xxxxx
Version: 6 | Release date: 14. July 2021
Page 43
6.2
Returning to Operation after Servicing / Cleaning Work
The responsible organisation is responsible for returning the
device to operation after servicing / cleaning work for which the
operator / user is authorised.
Servicing work here also includes the necessary checks which
are necessary if Spirostik should fall down.
Before the Spirostik can be used for measurements again, all
components must be properly connected / mounted and
calibrated. Calibration also fulfils the regulatory requirement of
MPBetreibV §7 for functional testing after maintenance work.
Please proceed as described in the respective chapters.
You can get precise instructions on assembly etc. in:
•
Chap. 7.4 “Inserting the Flow Sensor”
•
Chap. 7.5 “Calibrating Spirostik”
Possible physical injury.
Reason: Measurement error due to leakage caused by
the incorrect assembly of components. Therefore:
•
Carefully observe the assembly instructions!
(See chap. 8.2.1 “Checking for Damage”)