Instructions for Use
Spirostik
Serial no.: xx|8|001|yyy and 2001xxxxx
Page 46
Version: 6 | Release date: 14. July 2021
7.4
Inserting the Flow Sensor
1.
Insert Spiraflow [2] into the
handle [1]
For this:
Align the Spiraflow so that the
interrupted guide rib [2a] and the
nose on the handle are on the
same side
and
the pressure tapping holes [2b]
show in the direction of the
handle.
2.
Snap the Spiraflow [2] into the
handle [1]!
For this:
Apply light pressure
Spiraflow not only must be fixed,
but also locked in place!
For this:
Check the correct position in the
handle on the basis of the
interrupted guide rib and the nose
and correct it if necessary.
2b
1
1
2
2a
2
1