Geotech TempHion ISE Instructions Manual Download Page 22

TEMPHION INSTRUCTIONS

Seametrics • 253.872.0284 

Page 22 seametrics.com

and the sensor may read erratically!

Excess filling solution 

spills over, forcing out 

any air bubbles

• 

Once filled, be sure the O-ring is in the upper 

groove to prevent reference solution leakage and 

contamination. 

Before deploying be sure to move 

the O-ring to the lower groove to  expose the 

liquid junction port. 

pH Field Calibration Temperature Chart

The  pH  of  any  buffer  varies  slightly  with  temperature 

change. For this reason, it is important that you determine 

the current pH value of the buffer you are using during 

calibration. From this chart, determine the pH value at 

your current buffer temperature and use this value as the 

“Ref pH” value in the Settings and Calibration window. For 

example, if you are using a pH buffer of 7 and your buffer 

temperature is 20 degrees C, then you would enter 7.02 in 

the Ref pH box.

Calibration Temperature

˚C

˚F

pH 4.00

pH 7.00

pH 10.00

10

50

4.00

7.07

10.19

15

59

4.00

7.05

10.12

20

68

4.00

7.02

10.06

25

77

4.00

7.00

10.00

30

86

4.02

6.99

9.94

35

95

4.02

6.98

9.90

40

104

4.04

6.97

9.85

45

113

4.05

6.97

9.82

50

122

4.06

6.97

9.78

MAINTENANCE and REFERENCE

M & I Range for pH Calibration

There are three variations of TempHion pH configuration. 

Depending on which way yours is configured, your M and I 

values should be between those values shown in the chart 

below when you calibrate the sensor. 

To determine which configuration you have, you will need 

to  create  a  short  data  file,  then  retrieve  and  export  the 

data—it doesn’t matter whether the sensor is deployed, or 

in a buffer, or just in air for this. Once you have exported 

your data file, open it in Excel and note the P and H values 

for the pH channel.

•  If P = 2.97 and H = 0.567, this is the preferred 

configuration. 

•  If P = -2.389 and H = -0.29, this is an older 

configuration. This can still be used, but if you need 

higher accuracy, contact Seametrics for instructions 

on how to reconfigure.

Depending on which version you have, your M and I 

values during calibration of the pH channel should be the 

following ranges. 

How to Store and Recondition

When storing your TempHion, care must be taken to 

properly protect the electrodes, including the reference 

electrode. After storage, before use, the electrodes may 

need reconditioning, depending on the electrode and the 

mode of storage.  Seametrics’ recommendations follow. If 

you have further questions, please contact your Seametrics 

technical support representative.

Storage

Reference Electrode

• 

Best if stored in filling solution. 

•  Alternately, can be stored in KCl.

•  If storing dry, be sure to empty and clean reservoir. 

pH bulb / ORP

•  Best if stored in pH 4 or KCl.

•  Can be stored dry, but may shorten life of bulb. In 

Seametrics’ experience, the bulb should last roughly 

two years if typically stored dry.

ISE

•  Storing dry is best. 

P = 2.97

H = 0.567

P = – 2.389

H = – 0.29

P = 6.76

H = 0.567

Range for M

–53 to –57

–50 to –60

–50 to –60

Range for I

240 to 270

180 to 300

108 to 300

Summary of Contents for TempHion ISE

Page 1: ...SA C e rtified Compa n y ISO 9001 2008 TempHion pH ORP Redox ISE Smart Sensor and Data Logger Instructions TempHion pH ORP Redox TempHion ISE For Sales Service Contact 2650 E 40th Ave Denver CO 80205...

Page 2: ...TEMPHION INSTRUCTIONS Seametrics 253 872 0284 Page 2 seametrics com...

Page 3: ...ng Page 16 Retrieving Data Page 16 Viewing Data Page 16 Exporting Data Page 17 A Word about Units Page 17 Direct Read Modbus SDI 12 Setting Units for Direct Read Page 17 Power Consideration Page 18 Re...

Page 4: ...Alternately it can be powered with an external auxiliary power supply for data intensive applications The unit is programmed using Seametrics easy to use Aqua4Plus or Aqua4Plus Lite control software...

Page 5: ...s a4d and csv also xls in Windows 8 and earlier Output Channels Temperature pH ORP Bromide Element 30K ohm thermistor Epoxy bead external housing Glass combination electrode Platinum ring Ag AgCl soli...

Page 6: ...ge is found immediately file a claim with the carrier Check the label attached to the cable at the connector end for the proper cable length Do s and Don ts Don t drop into well Lower gently to preven...

Page 7: ...ed up for cleaning refilling or other maintenance Reference Electrode Assembly The modified thread provides a continuous liquid path between the solution in which the instrument is immersed and the re...

Page 8: ...nal batteries If auxiliary power is desired you can use a 6 16 VDC supply that can provide 15 mA Connect to Vaux pin 1 white and Ground pin 5 blue or contact Seametrics for auxiliary power supplies Co...

Page 9: ...rics Aqua4Plus or Aqua4Plus Lite software However the TempHion is quite versatile communicating via either Modbus or SDI 12 interfaces allowing you to do the following Read a TempHion via Modbus using...

Page 10: ...t at all times reach high enough to cover the sensing element cutout Be sure the supplied cap is securely placed on the weather resistant connector at the top of the cable Do not install such that the...

Page 11: ...he window Notethatthetemperaturechannelrarelyneedsadjustment If you think your temperature channel needs calibrating contact your service representative Use the calibration calculator to perform one a...

Page 12: ...ance and Reference section From this chart determine the pH value at your current buffer temperature and use this value as the Ref pH value in the calibration instructions below For example if you are...

Page 13: ...roughout the calibration Temperature of the sensor must also be this temperature prior to calibration One Point Calibration Computing Calibration Value Rinse sensor with distilled water and pat dry Pl...

Page 14: ...Point Add 10 cc of the same standard to the water Depending on which solution you are using this will result in a concentration as shown below 0 1 Molar NaBr Bromide 791 8 ppm Stir to distribute stan...

Page 15: ...nel do not use the built in calculator in the Settings and Calibration Window Instead follow the instructions below Rinse sensor with distilled or deionized water and then pat dry with clean paper tow...

Page 16: ...of the current readings on the sensor Setting up Data Recording Click the tool button A Logging Schedule Window will open Refer to the software instruction manuals for details in describing your sched...

Page 17: ...quipment or networks You may need to use Aqua4Plus Aqua4Plus Lite to make a few settings prior to directly reading the TempHion with your equipment These might include the units for the returned value...

Page 18: ...Holding Registers Readings are located in two registers each starting at address 62592 TempHion register addressing is zero based i e starts at zero If your equipment uses one based addressing you wi...

Page 19: ...for temperature is Celsius To change this use the Direct Read Units option in the Aqua4Plus Aqua4Plus Lite Control Software When using the M MC C or CC command all mV channels will report in either p...

Page 20: ...n mV Concurrent Measurement a C a 0024 cr lf Request all measurements a D0 a 21 345 7 181 053 459 cr lf Read temperature pH in pH units ISE in ppm ORP in Eh a C1 a 0024 cr lf Request all measurements...

Page 21: ...Seametrics 0 54 0 Chloride ION Plus B 3 572 0 49 Bromide Seametrics 3 878 0 89 CAUTION Filling solutions are not considered hazardous but they can be irritating to the skin Protective gloves are advi...

Page 22: ...below when you calibrate the sensor To determine which configuration you have you will need to create a short data file then retrieve and export the data it doesn t matter whether the sensor is deploy...

Page 23: ...an be done by holding in your hand using a vise or using pliers as follows Swage Knurling Black square ring Housing Top cap Knurling Black square ring Housing Top cap Battery Type Two 1 5V AA batterie...

Page 24: ...in next page Cableless Sensor By Hand cabled version only 1 Tightly grasp the top cap in one hand 2 Brace your hand against something such as a table or the ground Do not allow the cable to be pinched...

Page 25: ...onnector it only goes in one way so make sure not to force it Pull wires gently to the side to allow battery removal Connector connected properly 4 Hold the top cap assembly at 90 to the housing openi...

Page 26: ...S BELOW Incorrect readings O ring has not been moved to expose the reference reservoir opening Move the O ring from the upper groove to the lower groove See the Installation section Element or referen...

Page 27: ...ltered or repaired by the Buyer without prior written approval by the Seller all warranties are void Equipment and accessories not manufactured by the Seller are warranted only to the extent of and by...

Page 28: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 14315r14 20180306 3 6 2018...

Reviews: