Reference Temperature
(Input during calibration)

15.0 to 25.0°C (59 to 77°F)

Temperature Coefficient
(Input during calibration)

0.0% to 4.0%

TDS Constant
(Input during calibration)

0.3 to 1.00

Cell Constant Range

4.50 to 5.50

ATC Probe

Thermistor, 10KΩ, at 25 °C

Calibration Backup

Yes

Audio Feedback

Yes, on all keys

Power Source

One 9V battery (included with meter)

Real time clock (RTC) on the 
EC300M also powered by 
CR2032 coin battery (3V)

Operating Range 
- Temperature

0 to 50 ºC (32 to 122 °F)

Operating Range - 
Relative Humidity

Up to 95%

Instrument Casing

Waterproof IP-67 with cable 
connector cap installed

Weight (with battery)

272 grams (.6 lbs)

Dimensions (L x W x H)

18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 
in x 1.5 in)

Memory

50 data sets on the EC300A
250 data sets on the EC300M

Auto Shutoff

Automatically powers off after 
30 minutes of inactivity

Recal Timer

EC300M only - Customer 
selects recalibration interval; 
from 0 to 60 days

Real-Time Clock (RTC) 
for Date/Time Stamp 
of Saved Data

EC300M only

Waterproof USB for 
Downloading Data to PC

EC300M only

606042REF-B-EC300A-EC300M-Manual-English.indd   15

12/11/2017   10:07:45 AM

Summary of Contents for EcoSense EC300A

Page 1: ...USER MANUAL English Fran ais Espa ol Portugu s EcoSense EC300A EcoSense EC300M Portable Conductivity Salinity and Temperature Instruments User Manual DOCUMENT 606042REF...

Page 2: ...amage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructions ii abuse or misuse of the product iii failure to maintain the product in accord...

Page 3: ...t Modes 6 Calibration 7 TDS 7 CELL 7 Temperature Coefficient 8 Temperature Reference 8 Conductivity Calibration 8 Conductivity Measurements 8 Saving Viewing and Deleting Data 9 Downloading Data to a C...

Page 4: ...Accessories Part Numbers 14 Technical Services 14 Specifications 14 606042REF B EC300A EC300M Manual English indd 4 12 11 2017 10 07 39 AM...

Page 5: ...l contaminants that may affect probe connections 3 Wait for unit and all connections to dry before reinstalling the battery and resuming operation 4 If the unit does not function correctly after step...

Page 6: ...attery cover and o ring Figure 1 Replace the 9V battery Replace the battery cover and o ring be sure to align the o ring correctly to prevent a bad seal and fasten the two battery cover screws Battery...

Page 7: ...vailable with a single four electrode cell Other features include automatic conductivity ranging automatic temperature compensation long battery life and 50 60 Hz AC noise rejection This meter is univ...

Page 8: ...ction indicator EC300M only 8 Displays when measuring salinity 9 Calibration mode indicator 10 Data set number 11 Temperature display 12 Indicates TDS measurement 13 Indicates conductivity measurement...

Page 9: ...oSense Mode CAL Recal Timer Date Time Up and down arrow keys Increases or decreases the display value as desired 3 DO200M INPUT EcoSense Mode CAL Recal Timer Date Time Calibration key Press to enter t...

Page 10: ...confirms calibration steps and confirms data deletion On the EC300M this key confirms recal timer entry and date time selections 8 YSI EcoSense Handheld Overlay 12 19 2016 DO200M colors Black backgrou...

Page 11: ...quentially display the following sections Note Press Enter to accept any values changes in each section and automatically advance to the next section If there are no changes the unit accepts the curre...

Page 12: ...in a standard of known conductivity preferably a standard in the middle range of the solutions to be measured Completely submerge the probe without touching the sides of the calibration container Shak...

Page 13: ...ice to return to the measurement mode If Each is selected the Up and Down arrow keys will allow you to scroll through the saved data records Press Enter to delete the selected record All records after...

Page 14: ...urn to the run screen Any value between 0 and 60 days can be selected Set the value to 0 to disable the Recal Timer Date Time Settings EC300M Only A short press i e key is not held of the Date Time ke...

Page 15: ...for service Undr Temperature 10 0 C Opening the Data File with Excel Depending on the region and language setting of your PC measurement data might be formatted incorrectly by Excel when the data fil...

Page 16: ...t Wizard choose Delimited 5 Under Step 2 of the Text Import Wizard choose Semicolon 6 Under Step 3 click on the column with the incorrectly formatted data This column should be highlighted in black Ch...

Page 17: ...1 Dip the cell in cleaning solution and agitate for two to three minutes Any foaming acid tile cleaner such as Dow Chemical Bathroom Cleaner should clean adequately For a stronger cleaner use a soluti...

Page 18: ...com S P E C I F I C AT I O N S These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website a...

Page 19: ...ure 0 to 50 C 32 to 122 F Operating Range Relative Humidity Up to 95 Instrument Casing Waterproof IP 67 with cable connector cap installed Weight with battery 272 grams 6 lbs Dimensions L x W x H 18 7...

Page 20: ...2013 IEC 61326 1 2012 IEC 61000 3 2 2005 IEC 61000 3 3 2008 IEC 61000 4 2 2008 IEC 61000 4 3 2006 IEC 61000 4 4 2004 IEC 61000 4 6 2008 IEC 61000 4 8 2009 Item 606042REF Revision B December 2017 For t...

Reviews: