Problem
Possible cause
Solution
No irrigating solution
supply
The pump power
supply is off.
The pump cassette is
installed in the well
incorrectly
Peristaltic pump is
damaged.
Lack of/bad irrigation
tubes connection.
Irrigation tube is
damaged.
The nozzle of SAF
system is damaged.
The solution in the
bottle is over or the
silicone tube is not
lowered to the bottom
of the bottle.
Switch on the pump power
supply (
see section 6, Step
5)
Fix the pump cassette in
the well on the pump unit,
rotating it clockwise to the
typical click.
Apply to the service
department.
Check the tubes
connection
(
see section 6, Step 4)
Replace the silicon tube
and/or the pump cassette
with tubing
Replace the SAF-file
Fill the bottle with the
solution or lower the tube
to the bottom of the bottle,
joining the grey mark on
the tube with the aperture
on the cap.
If you have not found the necessary information, You may consult
the manufacturer on the phone: +7(495)663-22-11 (extension
109), E-mail: [email protected]
or address to the service
department