In all commands, ‘a’ represents the device address. Blue rows signal that parameters can be set using
this command.
Supported SDI-12 commands are given in Table 10.
Table 10. SDI-12 commands
Name
Command
Response
Details
Address query
?!
a<CR><LF>
Device will identify using its SDI-
12 address, default address is 0
Ping sensor
a!
a<CR><LF>
Device will respond if its address
is ‘a’
Address change
aAb!
b<CR><LF>
Device will respond if its address
is ‘a’ with its new address ‘b’
Send identification
aI!
a13GEOLUX LX-80
fffssssss<CR><LF>
a – address
SDI-12 version – 1.3
vendor identification – GEOLUX
sensor model – LX-80
sensor version – fff, where fff is
firmware version
ssssss - device serial number
Start verification
aV!
a00001<CR><LF>
One value is ready imminently
Send data
aD0!
a+d<CR><LF>
d – verification:
0->not ready , 1->ready
Additional data
aD1! .. aD9!
a0<CR><LF>
No values are returned for addi-
tional data.
Start measurement
aM!/aMC!
ammm8<CR><LF>
8 values are ready within mmm
seconds
Send data - aM
aD0!
a+f+f+d+d+d+d+f+f<CR><LF>
a+f+f+d+d+d+d+f+f<CRC><CR>
<LF>
+f – measured relative level
depending on sensor height
+f – measured distance from
sensor to water
+d – measured temperature inside
device
+d – measured water temperature
(on request only)
+d – measured tilt angle of device
in x direction (on request only)
+d – measured tilt angle of device
in y direction (on request only)
+f – SNR of latest measurement
+f - standard deviation of the data
in mm
Additional data
aD1! .. aD9!
a0<CR><LF>
a0<CRC><CR><LF>
No values are returned for
additional data.