4. COMMAND LIST
The commands listed here are to be used if communications between the LC-2 and the host
computer are established via a terminal emulator (e.g., Windows HyperTerminal). If using
LogView, these commands can be ignored. To send commands and receive information from the
Model LC-2 the communications mode must be established between the host computer and the
datalogger (see Section 3.4 and Section 3.5).
Pressing
?
<ENTER> while in the communications mode displays the following list of
commands:
Command
Description
----------------------------------------------------------------
C
View current Clock
CSmm/dd/yy/hh:mm:ss Clock Set
DEFAULT
Load factory DEFAULT settings
DF Date Format (0=Julian, 1=month,day)
Dnnnnn Display nnnnn arrays from pointer
E End communications and go to sleep
Gnn/szzzz/sffff/soooo Gauge information, where;
nn = gauge type
szzzz = zero reading with sign
sffff = gauge factor with sign
soooo = offset with sign
IDdddddddddddddddd View current ID, set to ddddddddddddddd
LC
Select Linear Conversion
Ln/lllll/iii
View Log intervals/change n interval,
lllll=length,iii=iterations of interval
LD,LE Log intervals Disable, Enable
M,MD,ME Monitor status, Disable, Enable
MS
Memory Status
N Display Next time to read
NAddd
Network Address (1-256)
NS,ND,NE
Network Status, Disable, Enable
PC
Select Polynomial Conversion
Pnnnnn Position array pointer to nnnnn
R
Reset memory
RESET
Reset processor
RT
Readings total (8=8000 readings,
16=16000 readings)
SR
Synchronize readings (0=not synch’d,
1=synch’d)
S,SS
Datalogger status, System status
SCnnnnn
View Scan interval/enter nnnnn interval
SPhh:mm
Stop logging, hh:mm = stop time
SThh:mm
Start logging, hh:mm = start time
SV
Software Version
TEST
System Test
TF
Time Format (0=hhmm, 1=hh,mm)