background image

MODEL GK-404 VIBRATING WIRE READOUT 

 |  THERMISTOR TEMPERATURE DERIVATION  | 

 11

B.2 THERMISTOR RESISTANCE FOR 8.2KΩ

Thermistor Type: Thermometrics BR55KA822J

Resistance to Temperature Equation:

EQUATION 2: 

Thermistor Resistance for 8.2kΩ

Where:

T

 = Temperature in °C

LnR

 = Natural Log of Thermistor Resistance

A

 = 1.02569 x 10

-3

 

B

 = 2.369 x 10

-4

 

C

 = 1.019 x 10

-7

Note:

 Coefficients calculated over the -50 to +150 

C span.

TABLE 8: 

Thermistor Resistance for 8.2kΩ

T =

1

– 273.15

A+B(LnR)+C(LnR

3

)

Ohms

Temp

LnR

LnR

3

Calc. Temp Diff F.S. Error Ohms Temp LnR

LnR

3

Calc. Temp Diff F.S. Error

113898

-30 11.643 1578.342

-30.17

0.17

0.06

407.62 120 6.010 217.118

120.00

0.00

0.00

86182

-25 11.364 1467.637

-25.14

0.14

0.05

360.8  125 5.888 204.162

125.00

0.00

0.00

65805

-20 11.094 1365.581

-20.12

0.12

0.04

320.21 130 5.769 191.998

130.00

0.00

0.00

50684.2

-15 10.833 1271.425

-15.10

0.10

0.03

284.95 135 5.652 180.584

135.00

0.00

0.00

39360

-10 10.581 1184.457

-10.08

0.08

0.03

254.2  140 5.538 169.859

140.01

-0.01

0.00

30807.4

-5 10.336 1104.068

-5.07

0.07

0.02

227.3  145 5.426 159.773

145.02

-0.02

-0.01

24288.4

0 10.098 1029.614

-0.05

0.05

0.02

203.77 

150 5.317 150.314

150.03

-0.03

-0.01

19294.6 5 9.868

960.798

4.96

0.04

0.01

183.11 

155 

5.210 141.428

155.04

-0.04

-0.01

15424.2 10 9.644

896.871

9.98

0.02

0.01

164.9 160 

5.105 133.068

160.06

-0.06

-0.02

12423 15 

9.427

837.843

14.98

0.02

0.01

148.83 

165 

5.003 125.210

165.08

-0.08

-0.03

10061.4 20 9.216

782.875

19.99

0.01

0.00

134.64 

170 

4.903 117.837

170.09

-0.09

-0.03

8200K

 25 

9.012

731.893

25.00

0.00

0.00

122.1 

175 

4.805 110.927

175.08

-0.08

-0.03

6721.54 30 8.813

684.514

30.01

-0.01

0.00

110.95 

180 

4.709 104.426

180.07

-0.07

-0.02

5540.74 35 8.620

640.478

35.01

-0.01

0.00

100.94 

185 

4.615 98.261

185.10

-0.10

-0.04

4592 40 

8.432

599.519

40.02

-0.02

-0.01

92.086 

190 

4.523 92.512

190.09

-0.09

-0.03

3825.3 45 

8.249

561.392

45.02

-0.02

-0.01

84.214 

195 

4.433 87.136

195.05

-0.05

-0.02

3202.92 50 8.072

525.913

50.01

-0.01

-0.01

77.088 

200 

4.345 82.026

200.05

-0.05

-0.02

2693.7 55 

7.899

492.790

55.02

-0.02

-0.01

70.717 

205 

4.259 77.237

205.02

-0.02

-0.01

2276.32 60 7.730

461.946

60.02

-0.02

-0.01

64.985 

210 

4.174 72.729

210.00

0.00

0.00

1931.92 65 7.566

433.157

65.02

-0.02

-0.01

59.819 

215 

4.091 68.484

214.97

0.03

0.01

1646.56 70 7.406

406.283

70.02

-0.02

-0.01

55.161 

220 

4.010 64.494

219.93

0.07

0.02

1409.58 75 7.251

381.243

75.01

-0.01

0.00

50.955 

225 

3.931 60.742

224.88

0.12

0.04

1211.14 80 7.099

357.808

80.00

0.00

0.00

47.142 

230 

3.853 57.207

229.82

0.18

0.06

1044.68 85 6.951

335.915

85.00

0.00

0.00

43.673 

235 

3.777 53.870

234.77

0.23

0.08

903.64 90 

6.806

315.325

90.02

-0.02

-0.01

40.533 

240 

3.702 50.740

239.69

0.31

0.11

785.15 95 

6.666

296.191

95.01

-0.01

0.00

37.671 

245 

3.629 47.788

244.62

0.38

0.13

684.37 100 

6.528

278.253

100.00

0.00

0.00

35.055 

250 

3.557 45.001

249.54

0.46

0.16

598.44 105 

6.394

261.447

105.00

0.00

0.00

32.677 

255 

3.487 42.387

254.44

0.56

0.19

524.96 110 

6.263

245.705

110.00

0.00

0.00

30.496 

260 

3.418 39.917

259.34

0.66

0.23

461.91 115 

6.135

230.952

115.00

0.00

0.00

Summary of Contents for GK-404

Page 1: ... 2019 GEOKON All rights reserved Document Revision O Release date 11 15 19 Model GK 404 Vibrating Wire Readout Instruction Manual ...

Page 2: ......

Page 3: ...ON is not responsible for any damages or losses caused to other equipment whether direct indirect incidental special or consequential which the purchaser may experience as a result of the instal lation or use of the product The buyer s sole remedy for any breach of this agreement by GEOKON or any breach of any warranty by GEOKON shall not exceed the purchase price paid by the purchaser to GEOKON f...

Page 4: ......

Page 5: ...INTENANCE 6 4 1 LOW BATTERY INDICATION AND AUTO OFF 6 4 2 BATTERY REPLACEMENT 6 4 3 CLEANING 6 4 4 CALIBRATION 6 5 TROUBLESHOOTING 7 6 LIMITS OF LIABILITY 8 APPENDIX A SPECIFICATIONS 9 A 1 VIBRATING WIRE READOUT 9 A 2 TEMPERATURE READOUT 9 A 3 PHYSICAL 9 A 4 TRANSDUCER PLUG PIN OUT 9 APPENDIX B THERMISTOR TEMPERATURE DERIVATION 10 B 1 THERMISTOR RESISTANCE FOR 3KΩ 10 B 2 THERMISTOR RESISTANCE FOR ...

Page 6: ...II FIGURES FIGURE 1 LEMO CONNECTOR 2 FIGURE 2 NECK STRAP CONNECTOR 2 FIGURE 3 CONNECTOR ON GK 404 2 FIGURE 4 DOWNWARD POINTING CONNECTOR 3 FIGURE 5 NECK STRAP CONNECTIONS 3 ...

Page 7: ...IFICATIONS 4 TABLE 3 VIBRATING WIRE READOUT 9 TABLE 4 TEMPERATURE READOUT 9 TABLE 5 PHYSICAL SPECIFICATIONS 9 TABLE 6 TRANSDUCER PLUG PINOUT 9 TABLE 7 THERMISTOR RESISTANCE FOR 3KΩ 10 TABLE 8 THERMISTOR RESISTANCE FOR 8 2KΩ 11 TABLE 9 THERMISTOR RESISTANCE FOR 10KΩ 12 ...

Page 8: ...IV EQUATIONS EQUATION 1 THERMISTOR RESISTANCE FOR 3KΩ 10 EQUATION 2 THERMISTOR RESISTANCE FOR 8 2KΩ 11 EQUATION 3 THERMISTOR RESISTANCE FOR 10KΩ 12 ...

Page 9: ...apable of operating for more than 20 hours continuously on two AA batteries It is designed for the readout of all Geokon vibrating wire gauges and transducers and is capable of displaying the reading in either digits frequency Hz period μs or microstrain με The GK 404 also displays the temperature of the transducer embedded thermistor with a resolution of 0 1 C ...

Page 10: ...e carrying case To connect the nylon strap to the carrying case clip both ends of the strap onto the plastic loops on the sides of the case Connect the neck strap to the GK 404 as follows 1 Locate the plastic connector on the neck strap see the figure below FIGURE 2 Neck Strap Connector 2 Place the larger circular hole onto the plastic nub on the back of the GK 404 with the bottom of the connector...

Page 11: ... GK 404 The individual leads can be identified as shown in the wiring diagram in Table 1 below Each sensor is read in turn by connecting either the red or black clip to the lead marked common The black or red clip must be connected to the leads marked 1 2 3 etc The blue clip should be connected to the cable shield wire and the green and white clips to the cable leads marked thermistor 1 White s bl...

Page 12: ...ing A F has different preprogrammed parameters that are optimized for certain instruments Each option also has different scaling and processing characteristics The table below details the correct position to use with the various models sold by GEOKON TABLE 2 Position Specifications 3 2 MODE BUTTON Pressing the Mode button will select the format of the vibrating wire reading The available options a...

Page 13: ...select automatic shutoff after five minutes 15 minutes or 30 minutes The Auto Off may be disabled so that the GK 404 will only shut off if the front panel On Off button is pressed The Auto Off will activate only if no front panel buttons are pressed for the specified amount of time The timer will reset each time a button is pressed When the GK 404 powers down all settings are saved to the internal...

Page 14: ...actory To replace the batteries use a 1 Phillips screwdriver to remove the battery compartment cover located at the back of the readout Install a new set of batteries following the polarities indicated inside the battery compartment Reinstall the battery compartment cover Never use two different types of batteries or mix old batteries with new This could cause batteries to leak and permanently dam...

Page 15: ...r of readout SYMPTOM VIBRATING WIRE GAUGE READING UNSTABLE Improper position selected Verify gauge model number against the recommended setting see Table 2 in Section 3 1 If position setting is correct the gauge is either marginal or there may be a strong source of electromagnetic interference nearby SYMPTOM THERMISTOR MEASUREMENT SHOWS DASHES Check alligator clip connections to thermistor leads I...

Page 16: ...lems be double checked using another measurement system In addition the readout is provided as is and GEOKON assumes no responsibility as to results performance or interpretation associated with the GK 404 vibrating wire readout Warranty shall cover parts except batteries and labor for a period of one year from the date of purchase In addition there is a one month grace period to the warranty for ...

Page 17: ...ype Thermistor YSI 44005 Dale 1C3001 B3 Alpha 13A3001 B3 Sensor Accuracy 0 5 Celsius Measurement Range 20 to 50 Celsius Measurement Resolution 0 1 Celsius Measurement Accuracy 1 0 of Full Scale Display 16 columns x two lines LCD Dimensions 120 mm 4 72 x 65 mm 2 56 x 22 mm 0 87 Weight Approx 145 grams 5 1 ounces with two AA alkaline batteries installed Temperature Range 20 to 50 Celsius Battery Two...

Page 18: ... 43 331 5 84 102 5 125 88 46K 38 8417 3 1363 44 321 2 85 99 9 126 82 87K 37 8006 4 1310 45 311 3 86 97 3 127 77 66K 36 7618 5 1260 46 301 7 87 94 9 128 72 81K 35 7252 6 1212 47 292 4 88 92 5 129 68 30K 34 6905 7 1167 48 283 5 89 90 2 130 64 09K 33 6576 8 1123 49 274 9 90 87 9 131 60 17K 32 6265 9 1081 50 266 6 91 85 7 132 56 51K 31 5971 10 1040 51 258 6 92 83 6 133 53 10K 30 5692 11 1002 52 250 9 ...

Page 19: ... 9 216 782 875 19 99 0 01 0 00 134 64 170 4 903 117 837 170 09 0 09 0 03 8200K 25 9 012 731 893 25 00 0 00 0 00 122 1 175 4 805 110 927 175 08 0 08 0 03 6721 54 30 8 813 684 514 30 01 0 01 0 00 110 95 180 4 709 104 426 180 07 0 07 0 02 5540 74 35 8 620 640 478 35 01 0 01 0 00 100 94 185 4 615 98 261 185 10 0 10 0 04 4592 40 8 432 599 519 40 02 0 02 0 01 92 086 190 4 523 92 512 190 09 0 09 0 03 382...

Page 20: ...7 0 138 119 4 170 63 3 202 36 2 234 18 971 11 4 725 43 1 480 75 555 3 107 241 1 139 116 9 171 62 1 203 35 6 235 18 090 12 4 543 44 1 432 76 539 9 108 235 3 140 114 5 172 61 0 204 35 1 236 17 255 13 4 368 45 1 385 77 525 0 109 229 7 141 112 1 173 59 9 205 34 5 237 16 463 14 4 201 46 1 340 78 510 6 110 224 3 142 109 8 174 58 8 206 33 9 238 15 712 15 4 041 47 1 297 79 496 7 111 219 0 143 107 5 175 57...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...GEOKON 48 Spencer Street Lebanon New Hampshire 03766 USA Phone 1 603 448 1562 Email info geokon com Website www geokon com GEOKON is an ISO 9001 2015 registered company ...

Reviews: