5
6. OUTPUT TO DIGITS CONVERSION
Use the following formula to convert the 4-20mA output signal to “Digits”:
Digits = 12250 - [625 x (mA-4)]
Equation 1 - Convert 4-20mA Output to Digits
Where;
mA = 8020-42CPR 4-20mA output signal
For example:
If the 8020-42CPR is outputting a 4-20mA current of 12mA then:
Digits = 12250 – [625 x (12 – 4)]
Digits = 12250 – [625 x 8]
Digits = 12250 – 5000
Digits = 7250