6
3.3 GK-403 Readout Box (Obsolete Model)
The GK-403 can store gauge readings and apply calibration factors to convert readings to
engineering units. The following instructions explain taking gauge measurements using Mode
“B”. Consult the GK-403 Instruction Manual for additional information.
3.3.1 Connecting Sensors
Connecting sensors with 10-pin connectors:
Align the grooves on the sensor connector (male), with the appropriate connector on the
readout (female connector labeled senor or load cell). Push the connector into place, and
then twist the outer ring of the male connector until it locks into place.
Connecting Sensors with Bare Leads:
Attach the GK-403-2 flying leads to the bare leads of a Geokon vibrating wire sensor by
connecting each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
3.3.2 Operating the GK-403
1)
Turn the display selector to position “B”.
2)
Turn the unit on.
3)
The readout will display the vibrating wire output in digits. The last digit may change
one or two digits while reading.
4)
Press the “Store” button to record the value displayed.
If the no reading displays or the reading is unstable, see Section 5 for troubleshooting
suggestions. The unit will turn off automatically after approximately two minutes to
conserve power.