
33
APPENDIX E. - MODEL 4500AR PIEZOMETER
Figure 13 - 4500AR
The Model 4500 AR piezometer is designed to be used with readout systems that can read
frequency but do not have the capability to “pluck” the VW gage. The sensor has built-in
electronics that cause the gage wire to vibrate continuously at its resonant frequency. The output
from the sensor is a five volt DC square wave at this frequency.
A DC input voltage in the range of 6 to 24 volts is required to operate the gage. The current
consumption is approximately 21 mA at 12VDC. The gage output is independent of the input
voltage.
Multiple sensors powered simultaneously can be read at quite fast rates, (up to five sensors per
second,) and dynamic measurements on a single sensor can be made up to approximately 20Hz.
The gage is installed in the field in the same way that the Model 4500 standard piezometer is
installed. (See Section 3 and Section 4.)
Piezometer Wiring
: The three pair cable is wired in pairs, with each pair comprising one
colored and one black lead.
Red +6-24 VDC Power
Red’s black Ground
White Output
White’s black Output Ground
Green Ther
Green’s black Thermistor –
Bare Shield
Upon power up the gage will immediately start to “ring” at the resonant frequency and will
continue to do so until the power is removed. Continuous operation will have no effect on the
gage life.
Note: The sensor is comprised of two transducers: the VW pressure sensor, and a thermistor for
measuring temperature. The signal from the VW transducer is a high level frequency and it will
interfere with the thermistor output if left powered during the period that the thermistor is being
read. If the temperature reading is important the power to the pressure sensor should be switched
off while the thermistor reading is taken.
Summary of Contents for 4500AL
Page 2: ......