3. Production Installation

1) Mounting the Sensor Probe

a. Operation temperature of Sensor Probe is -30 ~ 85 ℃ (-22 ~ 185℉)

b. UV sensor of Sensor Probe and UV light source should be fixed in set distance to do verticality.

Distance is your choice. But optical power should not exceed the 100.00

/

because it is the maximum of Display.

Once a mounting distance has been determined, make sure to use the same distance for any additional sensors used

for other lamps; doing so will provide a more accurate comparison of irradiance among different lamps.

c. In LW series, Sensor Probe connects after close teflon tape 3~4 times in screw page.

d. After mounting and positioning the sensor probe, make sure the window of Sensor Probe is clean and dry.

If it is not clean, then gently wipe it off with a lint free swab.

2) Mounting the Display Panel

a. Mount the Display Panel in a temperature & humidity of less than 50 ℃ (122℉) & 35 ~ 85 %RH.

b. 3 wires of Sensor Probe connect to #18, #19, and #20 of Display.

[18 : Vin (Green), 19 : +5V (Red), 20 : GND (Black)].

c. Power cable connect to #1and #2 of Display. (Power is AC voltage. Be careful!)

d. Voltage Output is 1~5 V DC along relative power, and it is the # 7 (-) and # 8(+) of Display.

e. Current Output is 4-20

DC (300Ω max load) along relative power, and it is the # 9(-) and # 10(+) of Display.

f. Relay can use the 5A / 230VAC, and its criterion is the setting value of C-01.

# 5 is N/O (Normal Open), # 4 is N/C (Normal Close), and # 6 is Comm (Common).

If you want to get the short signal under 70 %, you should set 70 in C-01and connect the each wires

in #5 and #6 of Display.

g. RS 485 communication Terminal is #11 B(-) , #12 A(+).

Fig. 2 Connection Diagram of UV Radiometer 5

Fig. 3 Connection picture of UV Radiometer 5

Power

Sensor Probe

18      19     20

1     2

B(-)    A(+)

RS485

Summary of Contents for MG-05.1

Page 1: ...2 Product Specification and Configuration 3 Product Installation 4 Front panel explanation 5 Function explanation 6 Relative Reaction Curves of UV Sensors 7 A S Request in Case of Product Failure 8 Notes 9 Relay Operation method Instruction Manual MG 05 1 UV Radiometer Ver 10 ...

Page 2: ...cification and Configuration 1 Display Panel Panel Size 97 50 112 Panel cutting size 92 45 Tolerance 0 0 5 Power Supply 85 265 VAC 50 60 Consumption 5 W Operating Temperature 0 50 Operating Humidity 35 85 RH Relay 250AC 3A Current Output 4 20 300 Ω Max Mode Shift Set 1 Display Panel 3 Panel Cutting Size 2 Size of Display Panel 2 Sensor Probe Green Line 18 Red Line 19 Black Line 20 Information of S...

Page 3: ...ry If it is not clean then gently wipe it off with a lint free swab 2 Mounting the Display Panel a Mount the Display Panel in a temperature humidity of less than 50 122 35 85 RH b 3 wires of Sensor Probe connect to 18 19 and 20 of Display 18 Vin Green 19 5V Red 20 GND Black c Power cable connect to 1and 2 of Display Power is AC voltage Be careful d Voltage Output is 1 5 V DC along relative power a...

Page 4: ...cator 5 Function explanation 1 Basic function description Set button Resetting RP Power Reset Max power Shift button Reset dose display Mode button Switchover the display mode you can change display mode to press mode button Hours Max Comm mJ Hours Max Comm mJ Hours Max Comm mJ Hours Max Comm mJ Hours Max Comm mJ Relative power display AT time display Absolute power display Max intensity display m...

Page 5: ...s the Set button over 3 second Accumulate time value will set to zero Reset Max Power In the Max power display mode Press set button for 3 seconds Press the Set button over 3 second Max power display value will set to 0 Measuring mJ value In the Dose mJ display mode press shift button Then the mJ cm2 value will set to zero and calculate the mJ Hours Max Comm mJ Hours Max Comm mJ Hours Max Comm mJ ...

Page 6: ...value you setting will flashing press the Mode button to adjust value f Shift button to change the setting unit g After all setting press the SET button 3s to quit setting mode and return to value shown mode h Press the mode button return to code shown mode i at the code shown mode press Shift Mode button then exit the setting mode Hours Max Comm mJ Hours Max Comm mJ Hours Max Comm mJ Hours Max Co...

Page 7: ...ayed intensity 0 19999 Please Do not change this setting to the factory setting C 04 Decimal Point setting 0000 0 0 0000 Please Do not change this setting to the factory setting C 05 Max intensity display Toggle ON OFF C 06 mJ display Toggle ON OFF C 07 RS485 Device No Setting 0 99 C 08 RS485 Baud rate 2400 4800 9600 C 09 MJ cm2 Measurement time 0 3600 3 Function setting table Page 6 ...

Page 8: ...y user s misuse or carelessness within warrant period or any failure after the warrant period shall be chargeable for it s A S 3 Product inquiry and on line customer service Tel 82 42 862 3982 Fax 82 42 862 2982 E mail uvsensor geni uv com Web site http www geni uv com 8 Notes 1 CAUTION TURN ALL POWER OFF NEVER EXPOSE EYES OR SKIN TO UV LIGHT FROM ANY SOURCE WEAR GLOVES FACE SHIELD GLASSES PER ANS...

Page 9: ...ltage output terminal is 7 and 8 current output terminal is 9 and 10 Please turn off the indicator before connect output signal cable If you want to use relay output please use with 6 and 5 N O Normal open in normal case the terminal status is open but if the signal is ON the terminal will be closed OR 6 and 4 N C Normal Close in normal case the terminal status is close but if the signal is ON the...

Reviews: