GNF - 722 - 12 (0) Genicom Co., Ltd.
A4(297
×
210
㎟
)
Page 2
3. Production Installation
1) Mounting the Sensor Probe
a. Operation temperature of Sensor Probe is -30 ~ 85 ℃ (-22 ~ 185℉)
b. UV sensor of Sensor Probe and UV light source should be fixed in set distance to do verticality.
Distance is your choice. But optical power should not exceed the 100.00
㎽
/
㎠
because it is the maximum of Display.
Once a mounting distance has been determined, make sure to use the same distance for any additional sensors used
for other lamps; doing so will provide a more accurate comparison of irradiance among different lamps.
c. In LW series, Sensor Probe connects after close teflon tape 3~4 times in screw page.
d. After mounting and positioning the sensor probe, make sure the window of Sensor Probe is clean and dry.
If it is not clean, then gently wipe it off with a lint free swab.
2) Mounting the Display Panel
a. Mount the Display Panel in a temperature & humidity of less than 50 ℃ (122℉) & 35 ~ 85 %RH.
b. 3 wires of Sensor Probe connect to #18, #19, and #20 of Display.
[18 : Vin (Green), 19 : +5V (Red), 20 : GND (Black)].
c. Power cable connect to #1and #2 of Display. (Power is AC voltage. Be careful!)
d. Voltage Output is 1~5 V DC along relative power, and it is the # 7 (-) and # 8(+) of Display.
e. Current Output is 4-20
㎃
DC (300Ω max load) along relative power, and it is the # 9(-) and # 10(+) of Display.
f. Relay can use the 5A / 230VAC, and its criterion is the setting value of C-01.
# 5 is N/O (Normal Open), # 4 is N/C (Normal Close), and # 6 is Comm (Common).
If you want to get the short signal under 70 %, you should set 70 in C-01and connect the each wires
in #5 and #6 of Display.
g. RS 485 communication Terminal is #11 B(-) , #12 A(+).
Fig. 2 Connection Diagram of UV Radiometer 5
Fig. 3 Connection picture of UV Radiometer 5
Power
Sensor Probe
18 19 20
1 2
B(-) A(+)
RS485