background image

550-0498

Pg. C-7

While the alarm has been reset, the device will continue to monitor

the air for the presence of CO.

When CO reaches alarm levels, the alarm will sound a repeating

temporal 4 repeating horn pattern - 4 beeps, a pause, 4 beeps, a

pause, etc. Press the reset button until the horn becomes silent.

The initial reset cycle will last approximately 5 minutes.

NOTE:

After initial 5 minute reset cycle, the alarm will re-evaluate

present CO levels and respond accordingly. If CO levels remain

potentially dangerous, or increase to higher levels, the alarm will

sound again.

While the alarm is silenced:
If the CO alarm:

This means:

Is silent for only 5 minutes,

CO levels are still

then the alarm sounds again

potentially dangerous

If the CO alarm:

This means:

Remains silent after the test

CO levels are dropping

button has been pressed

MAINTENANCE

DO NOT open CO alarm for cleaning. IF CO ALARM IS

OPENED, PRODUCT WARRANTY BECOMES VOID.
CAUTION:

If the device does not work properly, do not try and fix it

yourself. This will void your warranty. See "TO RETURN A CO

ALARM" for instructions to return CO alarms that do not operate

properly. DO NOT TRY TO FIX IT YOURSELF.

Gentex recommends CO alarms be tested a minimum of once a

week. The test feature of your CO alarm accurately simulates CO

conditions and tests the CO alarm's functions as required by

Underwriters Laboratories.

NOTICE:

Do not cover, tape, or otherwise block the openings of your

CO alarm. These openings are designed to allow air to pass

through your CO alarm, thus sampling the air around the CO alarm.

NOTICE:

CO alarms are not to be used with detector guards

unless the combination has been evaluated by a nationally

recognized testing laboratory and found suitable for that

purpose.
WEEKLY TESTING

Press the test/reset button on the alarm until the alarm sounds.

During testing the device will simulate a CO condition in the alarm.

The alarm sequence should last 5-6 seconds. If the device does

not go into alarm mode when tested, replace the device immediately.

If the alarm ever fails to test properly, replace it immediately.

Products under warranty may be returned to the manufacturer

for replacement, see “Limited Warranty.”

DO NOT stand close to the device when the alarm is

sounding. Exposure at close range could result in hearing

damage.

Never use exhaust from vehicle to test CO portion of alarm.

Exhaust may cause permanent damage to alarm and voids the

warranty.

TO KEEP THE ALARM WORKING PROPERLY:

Test weekly as described in “Weekly Testing.”

Keep alarm cover clean using soft cloth. DO NOT vacuum or use

compressed air, water, cleaners or solvents to clean alarm.

Replace battery immediately if low battery warning is heard. See

“Battery Installation” section.

NOTICE:

DO NOT spray cleaning chemicals or insect sprays

directly on or near the alarm. DO NOT paint over the alarm.

Doing so may cause permanent damage.

Household cleaners, aerosol chemicals and other

contaminants can affect the sensor. When using any of these

materials near the alarm, make sure the room is well ventilated.

The CO alarm is not washable. DO NOT submerge the alarm in

water. Water can affect the sensor, causing permanent damge.

If your home is being fumigated, disconnect unit

temporarily from wire harness and store where it will not be

exposed to chemicals or fumes. When fumigation is complete

and all traces of fumes clear, re-connect unit back to wire

harness and push the reset button.

UNDERWRITERS LABORATORIES, INC. UL 2034

WHAT LEVELS OF CO CAUSE AN ALARM

UL Standard UL 2034 requires residential CO alarms to sound

when exposed to levels of CO and exposure times as described

below. CO levels are measured in part per million (ppm) of CO over

time (in minutes).

UL 2034 Required Alarm Points

1

:

If the CO alarm is exposed to 400ppm of CO, THE DEVICE MUST

ALARM BETWEEN 4 - 15 MINUTES.

If the CO alarm is exposed to 150 ppm of CO, THE DEVICE

MUST ALARM BETWEEN 10 - 50 MINUTES.

If the CO alarm is exposed to 70 ppm of CO, THE DEVICE MUST

ALARM BETWEEN 60 - 240 MINUTES.

The device is designed not to alarm when exposed to a constant

level of 30 ppm for 30 days.

TROUBLESHOOTING GUIDE

PROBLEM:

THIS MEANS:

ACTION TAKEN:

CO alarm goes back

into alarm 5 minutes

after the reset button

was pressed.

CO levels indicate

a potentially

dangerous situation.

IF YOU ARE

FEELING

SYMPTOMS OF CO

POISONING,

EVACUATE your home

and call 911 or the Fire

Department. If not,

press the reset button

again and

keep

ventilating your

home.

Green light is OFF. Red

light is not flashing.

Unit will not go into test

mode when test button

is pressed.

Device may not be

receiving power.
NO AC and no

battery in device.

Contact licensed

electrical technician for

equipment inspection

service, immediately.

Alarm sounds 2 quick

chirps every 30

seconds.

Device has become

dirty or defective or

large influx of CO

(over 1,000 ppm).

Clean (refer to

Maintenance Section)

or warranty return.

Alarm sounds 3 quick

chirps every 30

seconds.

END OF LIFE

SIGNAL.

Device

needs to be

replaced.

Contact Gentex

Corporation for

replacement

information.

Green light is ON and

red light is not flashing

and Alarm chirps once

every 30 seconds

Low or no battery in

device
AC is powering

device

Replace battery (refer

to Battery Installation

Section) or return to

manufacturer

Reset button is

pressed. LED’s do not

flash and device does

not go into test mode.

Device is not

operating properly.

Contact Gentex

Corporation for

replacement

information.

Any questions that are not answered within this manual, call Gentex

Corporation at 1-800-436-8391.

1

Approximately 10% COHb exposure at levels of 10% to 95%

Relative Humidity (RH).

USING THE RESET FEATURE

The reset feature is for your convenience only and will not

correct a CO problem. ALWAYS check your home for a

potential problem after any alarm. Failure to do so can result

in injury or death.

NEVER remove the battery from your alarm to silence the

horn; use the reset feature. Removing the battery, removes

your protection! See “If Your CO Alarm Sounds” for details

on responding to an alarm.

The reset feature is intended to reset the CO alarm while the

problem is corrected -

IT WILL NOT CORRECT A CO PROBLEM.

Summary of Contents for CO1209 SERIES

Page 1: ...exposed to a critical level of CO measured in parts per million or ppm within a specified time measured in minutes This CO alarm features a permanently installed sensor an indicator light and an 85dBA...

Page 2: ...other appliances may also prevent alert persons from hearing the alarm horn This device is not intended for people who are hearing impaired CO alarms are not a substitute for life insurance Though th...

Page 3: ...cupiable level of a dwelling unit including basements excluding attics and crawl spaces 3 Other locations where required by applicable laws codes or standards The installation of additional CO alarms...

Page 4: ...DO NOT INSTALL THIS CO ALARM In garages kitchens furnace rooms or in any extremely dusty dirty or greasy areas Closer than 15 feet 4 6 meters from a furnace or other fuel burning heat source or fuel b...

Page 5: ...1125 feet between the first and last CO alarm NOTES ON TANDEM INTERCONNECTING MODELS DO NOT connect Gentex CO alarms to other manufacturers CO alarms No more than 12 Gentex models CO1209 may be tande...

Page 6: ...to add the telephone numbers of the emergency service provider and a qualified technician Once CO alarm is installed place one 1 label next to the CO alarm and place the other label near the fresh ai...

Page 7: ...en using any of these materials near the alarm make sure the room is well ventilated The CO alarm is not washable DO NOT submerge the alarm in water Water can affect the sensor causing permanent damge...

Page 8: ...entex postage prepaid You must also pack the CO Alarm to minimize the risk of it being damaged in transit You must also enclose a return address CO Alarms returned for warranty service should be sent...

Reviews: