8
o. In the
Advanced
tab, click on the
M²
button. Two (2) new tabs open and the translation
stage automatically connects to the computer.
p. Click on the right arrow in the
Manual Settings
, this moves the stage to the farthest
possible position.
q. Observe the beam in the
Display
tube) to center the beam on the sensor.
r. Click on the left arrow in
Manual Settings
, this moves the stage to the closest possible
position, then center the beam using mirror M1.
s. Repeat steps q. to s. if needed.
Figure 2-3:
Alignment tube
On the left, the alignment tube mounted with its 2 fluorescent pinholes. On the right, schematics of the
alignment tube with mirrors M1 and M2 aligning a laser beam.
M1
M2
P1
P2
Alignment Tube