SDX-1239 Operating Instructions
Rev - 3/21/2017
Page 16 of 22
FAC
Requests wavelength correction factor. Query only, returns the
correction being applied by Mach 5/6 for the user selected
wavelength. 1.00 if the wavelength is the calibration wavelength or of
wavelength correction is off.
Example.
Send: facCRLF
Reply: 1.000CRLF
RST
Reset the secondary processor to force a probe eeprom read. Returns
the MSP firmware version after the rest is complete.
Example.
Send: rstCRLF
Reply: 1.00.00CRLF
SMD
Sets the mode of operation.
0 is Measure mode
1 is Monitor Mode
2 is Control Mode
Example.
Send SMD2CRLF
Reply: OK
Send SMD
Reply 2CRLF
The SMD command side effects the AOM update rate. Setting the
Mode to Measure will set the AOM update rate to 10Hz. Setting it to
Monitor or Control will set it to 1kHz.The Update rate can then be set
to any desired value with the UPD command
AOM
Sets the AOM control voltage in Measure and Monitor Modes.
Example.
Send AOM0.55CRLF
Reply: OK
Send AOM
Reply 0.55CRLF
UPD
Sets the AOM control voltage update rate in Hz. The range is 10Hz to
10kHz.