4.3  SATURATION  
 

The  maximum  power  varies  with wavelength,  power  density  and  from  one  diode  to another.    When 
making  measurements  near  the  saturation  power,  you  must  verify  the  saturation  effect  with  a 
calibrated Filter. 
 

4.3.1 Procedure with a known transmission value filter 

 

 

Make the measurement with and without the filter. 
Your power ratio should be equal to the transmission value of the filter. 

   

 

 

4.3.2 Attenuator calibration procedure 

 

Make sure your stable power source is far below the saturation point. 
Make the measurement with and without the filter. 
The transmission value is the ratio of the measurement with filter / without filter. 

  

4.4  MEASUREMENT OF THE AVERAGE POWER OF A PULSE LASER BEAM 

 
Conditions to be met: 

The repetition rate must be at least 5 times the analog Bandwidth (see table IV). 

 

The peak power must not saturate the detector. 

  

 

To  know  if  the  detector  is  saturated,  use  procedure  4.3.1.      Be  careful  when  making  the  attenuator 
calibration in pulse mode.  The peak power must be in the linear region of the photodiode.   
 
Peak power = Energy per pulse / pulse width.  Energy per pulse = Average power / repetition rate.

 

 

4.5  WAVELENGTH 

 
The  photodiode  response  varies  with  wavelength.    You  may  select  your  wavelength  with  the 

Settings/Wavelength

  menu  of  the  monitor  or  enter  you  wavelength  in  the 

Settings/Custom

  menu  of 

the monitor. 
 
If  you  decide  to  use  the  photo  detector  without  a  Gentec-EO  monitor,  you  will  have  to  use  the 
sensitivity given by the photo detector calibration certificate to calculate the power on you laser beam.  
If  your  wavelength  is  not  given  by  the  calibration  certificate,  you  will  have  to  make  a  linear 
interpolation  between  two  of  the  available  calibration  values.    Fig. 1-4,    Fig. 1-5  and  Fig. 1-6 shows 
the typical spectral response. 
 
Linear interpolation formula: 
 

Slope

Sens

y

Sensitivit

LOW

desired

*

_

_

 

 

 

 
 

 

 
Sensitivity 

desired_

:  The sensitivity at the desired wavelength.   

LOW

DESIRED

)

(

)

(

_

_

LOW

HIGH

LOW

HIGH

Sens

Sens

Slope

Summary of Contents for PE10B-Si

Page 1: ...to obtain a return authorization number Return the material to the appropriate address below Contacting Gentec Electro Optics Inc To help us answer your calls more efficiently please have the model n...

Page 2: ...idity or surface contaminants exposure to flame solvents or water and connection to improper electrical voltage TABLE OF CONTENTS WARRANTY I First Year Warranty i Lifetime Warranty i TABLE OF CONTENTS...

Page 3: ...TO DETECTOR DIMENSIONS 3 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPON...

Page 4: ...uoted maximum ranges are average power at the 1064 nm wavelength for PH100 Si and PH20 Ge 532 nm for PH100 SiUV and 850 nm for PH100 SiUV with OD 2 quoted minimum ranges are average power at the 980nm...

Page 5: ...ific PH and PE Series head Wavelengths are not available where the sensor physical properties don t allow it The Gentec eo monitors use the data in this connector to adjust their characteristics autom...

Page 6: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 3 1 2 2 Dimensions FIG 1 1 PHOTO DETECTOR DIMENSIONS...

Page 7: ...30 mW 320 mW cm2 1064 nm Minimum Powera 0 6 nW 980 nm 0 6 nW 850 nm 2 nW 1550 nm Damage threshold Maximum Average Power Density 100 W cm 2 100 W cm 2 100 W cm 2 Typical Detector saturation currentb 6...

Page 8: ...W 980 nm 0 45 A W 850 nm 0 98 A W 1550 nm With attenuator PH100 Si OD1 OD2 PH100 SiUV OD1 OD2 PH20 Ge OD1 OD2 Maximum Power typ w OD 1 300 mW 1064 nm 25 mW 532 nm 300 mW 1064 nm Minimum Power a w OD 1...

Page 9: ...measurable power M Link 200 W S Link 175 W Maestro 150 W 633nm M Link 32 W S Link 26 W Maestro 22 W 1047 nm Maximum Average Power Density 65 mW cm2 532 nm 320 mW cm2 1064 nm Noise equivalent power 10...

Page 10: ...Repetition Rate 1000 Hz Max Pulse Width 10 s Maximum Measurable Energy M Link 0 16 J S Link 0 13 J Maestro 0 11 J 633nm M Link 3 2 nJ S Link 2 6 nJ Maestro 2 4 nJ 1047 nm Maximum Energy Density 5 J c...

Page 11: ...ack protective cover on the detector head before making the offset nulling If you want to subtract the room light do the offset nulling with the cover off 2 2 QUICK POWER MEASUREMENT PROCEDURE This se...

Page 12: ...component not related to the laser power This will add a systematic error to absolute power measurements This error may disappear from relative power measurements When you subtract two measurements m...

Page 13: ...0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 10 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE W...

Page 14: ...saturate the detector To know if the detector is saturated use procedure 4 3 1 Be careful when making the attenuator calibration in pulse mode The peak power must be in the linear region of the photo...

Page 15: ...esired Wavelength The difference between the desired wavelength and the inferior wavelength 4 5 1 Example You have a PH100 Si and your laser is at 632 8 nm See your CERTIFICATE OF CALIBRATION for the...

Page 16: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 13 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPONSE FIG 1 5 PH100 SIUV PH10B SI PE10B SI TYPICAL SPECTRAL RESPONSE...

Page 17: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 14 FIG 1 6 PH20 GE PH5B GE PE5B GE TYPICAL SPECTRAL RESPONSE...

Page 18: ...center when the detector reaches its end of life Breaking the calibration seal or opening the monitor will void the detector warranty The complete Detector contains 1 Detector with wires or DB 15 1 i...

Page 19: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 16...

Reviews: