• Inspect the unit for any damage that may have occurred during shipment.
If the equipment is damaged, report it to the carrier and contact your
supplier. Save all packaging material in the event a return is necessary.
• UPHO comes fully assembled, requiring a minimum set-up. Please install
the equipment on a clean, dry-stable surface far away from any compatible
electrical outlet.
• Allow some space around UPHO for proper ventilation. This unit is "FOR
INDOOR USE ONLY". Avoid operating in areas of excessive humidity or
extremes of temperature.
Removal of the Transport Security System
• A transport security system is located on the underside of UPHO:
comprised of 4 bolts (circled on picture below).
THIS SYSTEM MUST BE
REMOVED BEFORE OPERATING THE UPHO.
• Remove the 4 transport bolts from the UPHO by
TILTING THE UPHO
SLIGHTLY TO EXPOSE THE TRANSPORT BOLTS
(as seen in the picture) and
using a 13 mm x 16 mm metal wrench.
DO NOT TURN THE UPHO ON ITS
SIDE - THIS MAY LEAD TO INTERNAL DAMAGE.
• Please retain and carefully store the 4 transport bolts, as they will be
needed to reinstall if UPHO needs to be transported in the future.
• Operating UPHO with transport security system installed, or transporting
UPHO without the transport security device installed, can lead to damage
to the mechanical components of the instrument.