Set the trimmer of the probe so that the signal in Perception resembles the input
signal.
Figure 7.15 below shows how the signal should look. When the trimmer is
positioned incorrectly, undershoot or overshoot is seen in the signal.
A
B
C
Figure 7.15:
Trimming of response
-
Incorrect and correct waveform responses
A
Incorrect
-
Undershoot
B
Correct
C
Incorrect
-
Overshoot
GEN3t