NORMAL OPERATION

The UV254SD can measure and store three different kinds of values:

• The intensity (in mW/cm

2

) of UVA light

• The intensity (in mW/cm

2

) of UVC light

• Temperature (in ºF or ºC), as sensed by a Type K or Type J thermocouple

Pressing and holding the FUNC./HOLD/NEXT button will bring up, in order, three blank 
five-digit displays above the terms “A” (for UVA light), “C” (for UVC light) and “tP” (for
temperature). Each display also automatically shows the correct unit for the parameter.

1. To select the measurement mode

, release the FUNC./HOLD/NEXT button when the

desired parameter appears. This becomes the meter’s default measurement mode until
you change it.

Whenever the UV254SD is in measurement mode, you can turn off the backlight (which is
on by default) by briefly pressing (but not holding) the POWER/ESC button. To reactivate the
backlight, briefly press the button again.

Whenever the meter is in measurement mode, you also can check the current date and
time by briefly pressing the SET 

button. Doing so causes both values to appear briefly 

at the lower left of the display.

2. To measure the intensity of UVA light

, seat the UVA probe (with its cap on) in the

socket at the top of the UV254SD (callout 1-7) and use the FUNC.HOLD/NEXT button to
select UVA measurement mode. If the display shows a value other than zero, press the
LOGGER 

button (which has the word “Zero” stenciled above it) and hold it for at least

three seconds until you hear a beep. Then remove the cap from the probe.

Holding the probe by its handle, point the sensor at its end directly at the light source. The
display will read out the intensity of the source’s UVA light component in units of mW/cm

2

.

3. To measure the intensity of UVC light

, seat the UVC probe (with its cap on) in the

socket at the top of the UV254SD and use the FUNC.HOLD/NEXT button to select UVC
measurement mode. If the display shows a value other than zero, press the LOGGER 

button (which has the word “Zero” stenciled above it) and hold it for at least three
seconds until you hear a beep. Then remove the cap from the probe.

Holding the probe by its handle, point the sensor at its end directly at the light source. The
display will read out the intensity of the source’s UVC light component in units of mW/cm

2

.

4. To measure temperature

, first make sure neither the UVA nor the UVC probe is plugged

into the instrument. Then use the FUNC.HOLD/NEXT button to select temperature
measurement mode.

Obtain a Type K or Type J thermocouple and seat it in the socket (callout 1-8) on the top of
the UV254SD. The display will read out the sensed temperature in degrees F or C,
(depending on the setting chosen during setup), along with the type of thermocouple it is
configured for.

To change the unit of measurement, press the SET 

button to enter setup mode and

perform Step 8 of the setup procedure. To configure the meter for use with a Type J
thermocouple (Type K is the default), press the SET 

button to enter setup mode and

perform Step 7 of the setup procedure. 

7

Summary of Contents for UV254SD

Page 1: ...UV254SD USER S MANUAL UVA AND UVC LIGHT METER WITH DATALOGGING SD CARD Please read this manual carefully and thoroughly before using this product...

Page 2: ...p 5 6 Normal Operation 7 Holding and Storing Measurements 8 Automatic vs Manual Datalogging 8 9 Transferring Data from the SD Card to a Computer 9 Specifications 10 Maintenance Troubleshooting Tips 10...

Page 3: ...h the instrument Because it is microprocessor based the UV254SD can make full use of the portability reliability and large storage capacities that SD memory cards offer Measurements of light intensity...

Page 4: ...tors and other physical features of the UV254SD and its two supplied probes 4 1 1 Liquid crystal display 1 2 POWER ESC button 1 3 FUNC HOLD NEXT button 1 4 REC ENTER button 1 5 SET button 1 6 LOGGER b...

Page 5: ...AtE appears on the display Quickly within three seconds press the REC ENTER button The word dAtE will appear in the center of the display along with the flashing value 00 00 00 at the lower left above...

Page 6: ...e thermocouple type has been selected and stored the display will show a value above the letters SP t Press the SET or LOGGER button to decrease or increase the value until the desired sampling time a...

Page 7: ...for at least three seconds until you hear a beep Then remove the cap from the probe Holding the probe by its handle point the sensor at its end directly at the light source The display will read out t...

Page 8: ...datalogging press the REC ENTER button once The top line of the display will then show the term REC Pressing the LOGGER button at this point will make REC flash and add the flashing term LOGGER at th...

Page 9: ...press the REC ENTER button and hold it for at least three seconds This action causes the REC message to disappear from the top line of the display TRANSFERRING DATA FROM THE SD CARD TO A COMPUTER Aft...

Page 10: ...saving data 30 mADC with backlight on Dimensions of meter 6 97 x 2 68 x 1 77 in 177 x 68 x 45mm Weight of meter 12 38 oz 351g Dimensions of UVA probe head 1 77 in diagonal x 1 26 in L 45 x 32mm Dimens...

Page 11: ...zed personnel or misuse alterations normal wear and tear or accidental damage The defective unit must be returned to General Tools Instruments or to a General authorized service center freight prepaid...

Page 12: ...31 6100 FAX 212 431 6499 TOLL FREE 800 697 8665 e mail sales generaltools com www generaltools com UV254SD User s Manual Specifications subject to change without notice 2010 GENERAL TOOLS INSTRUMENTS...

Reviews: