General DPH230SD User Manual Download Page 5

SETUP

1. Choose the power source.

Before using the DPH230SD, be sure it is powered by fresh

batteries or an optional 9VDC AC adapter plugged into the bottom jack on its right side
(callout 1-15).

To remove the battery compartment cover, remove the two screws holding it in place
(callout 1-11). Then install six “AA” batteries in the correct orientation, using the polarity
marks on the inside of the compartment as a guide. Replace the cover by replacing the two
screws.

2. Install an SD card. 

To prepare for setup, also install the supplied 2 GB SD memory card

or another card with a capacity from 1 GB to 16 GB in the socket on the bottom of the
DPH230SD (callout 1-18). When installing the card, make sure its gold contacts are
facing front and push the card into the socket until you hear a click. To remove the card,
push it in until you hear a click and the card pops out.

Now power on the instrument by pressing the POWER button (callout 1-2) to generate a
short beep. (To power off the DPH230SD, press the POWER button and hold it until the
instrument responds with a long beep.) 

When the meter powers on, a series of transient startup screens will briefly appear. Once
the display has stabilized, perform the following eight setup steps in the order presented.

3. Format the SD card.

LOGGER SET button (callout 1-8) and hold it for at least five

seconds, until “Sd-F” appears in the lower half of the display. Quickly (within three
seconds) press the REC/ENTER button.

(If you press buttons too slowly in setup mode, the screen will revert to the normal display.
To return to the setup sequence, press and hold the LOGGER SET button again. To move
ahead to the next field in the sequence or to the next parameter, press the HOLD/ESC
button.)

Pressing the REC/ENTER button causes the word “no” or “yES” to appear over the term
“Sd-F”. Press the 

button to make a selection. Choose “yES” whenever a new SD card is

being used, or when a used card is being repurposed (from use with another SD card
instrument or a camera, for example) and all data on it is to be erased. Choose “no” to
preserve any data on a card previously used with this instrument.

5

1-1 Liquid-crystal display 

1-2 POWER button 

1-3 HOLD/ESC button

1-4 REC/ENTER button

1-5 

button 

1-6 

button 

1-7 TIME button

1-8 LOGGER SET button

1-9 Kickstand

1-10 Battery compartment cover

1-11 Battery compartment cover screws

1-12 Tripod attachment nut

1-13 Socket for ATC temperature probe

1-14 Socket for pH electrode

1-15 Socket for 9VDC AC adapter 

1-16 Reset button

1-17 RS232 output jack

1-18 SD card socket

Summary of Contents for DPH230SD

Page 1: ...METER WITH DATALOGGING SD CARD Please read this manual carefully and thoroughly before using this product 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at...

Page 2: ...Calibrating the pH Electrode 7 8 Normal Operation 8 9 Holding and Storing Measurements 9 Automatic vs Manual Datalogging 10 Transferring Data from an SD Card to a Computer 11 12 Specifications 13 Mai...

Page 3: ...enables rapid switching between the two measurement modes Measurement accuracies are 0 02 pH units 2 digits for pH readings and 0 5 of the reading 2 digits for ORP readings Because it is microprocess...

Page 4: ...obe that enables automatic temperature compensation ATC of pH electrodes an ORP electrode pH 4 and pH 7 buffer solutions and a 9VDC adapter for a 110V power supply See the Optional Accessories section...

Page 5: ...splay has stabilized perform the following eight setup steps in the order presented 3 Format the SD card LOGGER SET button callout 1 8 and hold it for at least five seconds until Sd F appears in the l...

Page 6: ...time Once the date and time have been set and stored the display will show a value below the letters SP t Press the or button to decrease or increase the flashing hour value until the desired number...

Page 7: ...setup procedure and enter normal operating mode CALIBRATING THE pH ELECTRODE The optional pH electrode you are using must be calibrated before making pH measurements with it for the first time The ele...

Page 8: ...en repeating the above steps at least twice more 11 To recalibrate the electrode in the future you must clear the existing settings To do so press the or button until this screen appears Then press th...

Page 9: ...solution under test Gently shake the head of the electrode to help any air bubbles escape and to ensure an accurate reading The display will show the ORP of the solution in mV Whenever the DPH230SD is...

Page 10: ...ep 5 of the setup procedure Then press the REC ENTER button once The display will show the term REC on the top line a value in the middle and below it the letter P on the left and a number from 1 to 9...

Page 11: ...either directly if it has an SD card slot or through an SD card reader Because the files containing time stamped data logs have the file extension xls they open in Microsoft s Excel application Figur...

Page 12: ...12 Fig 5 Typical Excel mixed data graphics screen Fig 4 Typical Excel graphics only screen...

Page 13: ...probe Storable recallable readings Maximum minimum Datalogging sampling time Settable between 1 second and 8 hours 59 minutes 59 seconds in automatic datalogging mode SD card capacity 1 GB to 16 GB S...

Page 14: ...requires removing the two screws that hold the battery compartment cover in place as explained in Step 1 of the setup procedure on page 5 After inserting fresh batteries in the correct orientation ti...

Page 15: ...ed Acceptance of the exclusive repair and replacement remedies described herein is a condition of the contract for purchase of this product In no event shall General be liable for any incidental speci...

Page 16: ...ut notice 2010 GENERAL TOOLS INSTRUMENTS NOTICE WE ARE NOT RESPONSIBLE FOR TYPOGRAPHICAL ERRORS MAN DPH230SD 12 07 10 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431...

Reviews: