Model IR4000M
27
1.5.3 Status Error (Read-Only)
A Read returns the bit map for any error that is presently occurring. The following
table shows the errors that are represented by each bit in the register.
Table 16 Bitmap for Status Error Delayed
Bit Position
3 2 1 0
Error
IR High
IR Low
Clean Window
(Negative Gas
Reading)
IR Close to Low
(External Gain)
Bit Value
8 hex / 8 dec
4 hex / 4 dec
2 hex / 2dec
1 hex / 1 dec
Bit Position
7 6 5 4
Error
Failed to Zero
Calibration Mode
Fail
Low Line
Condition
Wire Shortage
Bit Value
80 hex / 128 dec
40 hex / 64 dec
20 hex / 32 dec
10 hex / 16 dec
Bit Position
11 10 9 8
Error
Heater problem
Ref. Lamp Problem
Active Lamp Prob.
Gas Left
Bit Value
800 hex / 2048 dec
400 hex / 1024 dec
200 hex / 512 dec
100 hex / 256 dec
Bit Position
16 15 14 13
Error
EEPROM Error
Excess Neg. Drift
Misc Fault (Needs
repair)
unused
Bit Value
8000 hex / 32768 dec 4000 hex / 16384 dec
2000 hex / 8192 dec 1000 hex / 4096 dec