17
4.3
Descriptions of Working Parameters
C od e
Parameter
Default
Description
F1
Null
Null
The first major term of working parameter.
F1.1
4-20/0-20/
0-24/0-5/
0-10/-5-5/
-10-10/ USEr
4-20
The analog output form
:
4-20
:
4-20mA
;
0-20
:
0-20mA
;
0-20
:
0-24mA
;
0-5
:
0-5V
;
0-10
:
0-10V
;
-5-5
:
-5-5V
;
-
10-10
:
-10-10V
;
USEr:
user-defined analog output
F1.2
ON/
OFF
OFF
Switch for Auto-Zeroing when power-on,
OFF
: disabled
ON
: enabled
F1.3
0
~
9
0
Zero-tracking Range
(
0
~
9d optional
)
.
This parameter is for automatic calibration, disabled
when is set
”0”
.
F1.4
1
~
9
1
Motion Detecting Range
(
1
~
9d optional
);
F1.5
00
~
9 9
50
Zeroing
Range
(
00%
~
99%
of
Maximum
capacity
);
F1.6
0
~
9
5
Digital filtering parameter:
0:
without filtering
9:
strongest digital filtering
F1.7
0
~
9
0
Stable filter parameter
(
the second filter based on
the first filter
)
:
0:
without filtering
9:
strongest digital filtering
F2
Null
Null
The second major term of working parameter.
F2.1
ON/OFF
OFF
Password Switch
F2.2
000000
Password Setting;
Enabled when parameter
F2.1
is set
“ON”
.
F3
Null
Null
user-defined parameters of analog
quantity
(
Display when the parameter
F1.1
is
chosen
”USEr”.
F3.1 IOUT/ VOUT
IOUT
Chose the analog output mode
Summary of Contents for Tuxon-A
Page 1: ...Tuxon A User s Manual Ver 1 2012 10...
Page 4: ......
Page 7: ...3 1 5 Technical Specifications...
Page 8: ...4 1 6 Dimensions of Indicator...
Page 9: ...5 2 Installation and Wiring 2 1 How to Install Indicator...
Page 13: ...9 3 2 Flow Chart of Calibration 3 2 1 Flow Chart of Calibration for Tuxon A...
Page 14: ...10...
Page 20: ...16 4 2 Parameter Setting Method 4 2 1 Data Input Method 4 2 2 Option Selecting Method...