10
11
POLY-TUBE CLEANING
Cleaning the poly-tube
Slowly insert the cleaning filament end of the
Simplicity multi-tool into the sound port of the
dome flex-tip. Push the filament through the
dome completely to ensure all debris has been
removed. If you are unable to clear the blockage,
please contact the Readywearexpress customer
service dept. (page 21).
With the dome removed, insert the filament line
into the opening of the dome. Push the filament
through the dome completely to ensure all debris
has been removed.
Reinsert the flex dome on to the poly-tube, then
insert tubing connector into opening of the device
with a slight back and forth twisting motion and
push into place. Rotate the tube turn it back to its
original position.
POLY-TUBE REMOVAL
Removing the poly-tube
Turn the tubing (at connector base) left or right
ninety degrees from its original position.
Be sure to turn the tubing at the base where the
tube inserts in to the device, otherwise damage to
the tubing could occur.
The tubing position should change from directly
behind the device to parallel to its side. With a
slight back and forth twisting motion, pull apart
the device and poly-tubing.
After the poly-tube has been removed, the
dome flex tip will need to be separated from the
tube. Remove the dome flex-tip from the poly-tube
by gently twisting the poly-tube and pulling the
dome tip. Do not pull the dome by the outer edge,
hold the dome at the base where it connects to the
tubing.