Problem
Cause
Solution
______________________________________________________________________
No Functions operate
No power to unit
Check power cord.
Is it plugged into receptacle?
Green Power-on light
Electrical power cord not
does not glow
properly plugged into
Check cord for breaks.
control module
Power cord plugged
Check for power to receptacle.
In but no functions
operate
Bad handset
Replace handset.
Bad control module
Replace control module.
_____________________________________________________________________________
Three actuators
Bad control module
ISOLATE THE PROBLEM
operate “OK” one
Interchange connections on
actuator will
Bad handset
control module so that problem
not operate
actuator is plugged into an
Bad actuator motor
actuator connection on the
control module that is working
satisfactorily. If problem
actuator now operates “OK”
but actuator that was operating
“OK” now has a problem, either
the control module or handset is
faulty. Replace as required.
_____________________________________________________________________________
Actuator will not
Defective wiring
Check wiring between control
function properly
and actuator. Reverse wires.
or not at all
Defective actuator motor
Replace actuator
_____________________________________________________________________________
Battery will not charge
Defective wiring
Check wiring between control
and battery.
Defective battery
Replace battery
_____________________________________________________________________________
Bed not grounded
Unit is not grounded between
Check ground (green wire)
power cord and frame or
between frame and power cord.
between power cord and wall
receptacle
Is power cord plugged into a
grounded wall receptacle?
_____________________________________________________________________________
Caregiver control LED’s
Controller error
Reset controller- hold knee up
flashing – no functions
and knee down buttons on
operate
handset simultaneously for 5
seconds until beep sequence
starts
Diagnostics & Repair (Power Drive System)