Gendex Dental Systems

E-8

4519-190-70943_4

ORALIX AC SYSTEM

PREVENTIVE  MAINTENANCE

4.2.6. System adjustments

The system adjustments shall be performed with the arms and Tubehead completely assembled.

In standby position the arm sections must be perpendicular, laying respectively horizontal and vertical to the
floor.

When properly adjusted the arms shall return in standby position, both horizontally and vertically, when slightly
displaced by hand.
On the contrary the arms shall remain steady in work position when a larger displacement is applied.
If not, proceed first to an adjustment of the friction screw: when this is not sufficient adjust also the spring
tension of the arm section involved (for the adjustment procedures see the Section F of this manual).

The Tubehead shall balance in all positions and the torque required to rotate it shall not exceed 1.8 Nm in the
horizontal plane or 1.5 Nm in the vertical plane.
If an adjustment of the frictioning of the vertical rotation is required, provide it as described in the Section F of
this manual.

The Extension Arm is provided of friction screws for the adjustment in the horizontal rotation of the Folding Arm.
Adjust them if required according to the Section F of this manual.

In case of Mobile System, adjust if necessary the verticality of the Mobile Stand as described in the Section F
of this manual.

NOTE

There in no facility on the Mobile Stand for the frictioning of the horizontal rotation of the
Folding Arm around the Mobile Stand column

4.3.

ELECTRICAL  CHECKS

4.3.1. Power supplies

Remove the Timer cover, check with a digital DC voltmeter that the supplies are correct on PCB TC1, according
to the table below:

voltage

measure point

reference point (0V)

+5V

X4-8

X4-7

+9V

X4-9

X4-7

+16 to 22V

X4-10

X4-7

Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER

Summary of Contents for ORALIX AC

Page 1: ...contains descriptive information on the Oralix AC intraoral dental X Ray system Disclosure to third parties or reproduction in any form without written consent is prohibited Gendex Dental Systems Pri...

Page 2: ...XAC A 5 3 4 FOLDINGARM A 5 3 5 EXTENSIONARM A 5 3 6 WALL SUPPORT A 5 4 TECHNICAL DATA A 6 4 1 MAINS SUPPLY A 6 4 2 TUBEHEAD A 6 4 3 COLLIMATOR A 4 4 TIMER A 4 5 WEIGHTS A 5 IEC CLASSIFICATION A 6 APPL...

Page 3: ...logist who wants to make use of the parallel tecnique The stable and strong mechanical construction ensures an optimal image quality allowing accurate and simple patient positioning WARNING X ray unit...

Page 4: ...ralixAC 230 V 9869 000 00101 Densomat Timer 9869 002 00102 Secondent Timer 9869 002 00202 Wall Support 9869 005 00001 Extension Arm 90 cm 9869 001 00201 45 cm option 9869 001 00001 FoldingArm 9869 001...

Page 5: ...onsistently providing films with optimal density The voltage compensation can be inhibited by means of a dedicated jumper It also incorporates other circuits dedicated to safety functions such as back...

Page 6: ...ix AC 230 V consists of a high voltage and filament transformer and 3 elements grid control X ray tube in a lead shielded oil filled metal case covered by a plastic housing A collimator positioning de...

Page 7: ...l mains Voltage 10 Tube current see section D 3 3 6 mA 10 Nominal mains Voltage 10 mA 10 Nominal mains Voltage mA 10 Nominal mains Voltage 10 Radiation output 5 mGy sec 30 at 20 cm from the focus Tube...

Page 8: ...7 1 7 HQGH HQWDO 6 VWHPV 25 SYSTEM QRGH 7XEH 5HIHUHQFH LV DQG QRGH 7DUJHW QJOH 5HIHUHQFH LV QRGH LV 5HIHUHQFH LV Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 9: ...Actual exposure time For 50 Hz is the number of X ray pulses correspond ing to the selected time For 60 Hz is the number of X ray pulses with mini mum deviation respect to the selected exposure time t...

Page 10: ...UD JHQHUDWRUV 3DUWLFXODU UHTXLUHPHQWV IRU WKH VDIHW DQG SHUIRUPDQFH RI UD WXEH DVVHPEOLHV IRU PHGLFDO GLDJQRVLV 3DUWLFXODU UHTXLUHPHQWV IRU WKH VDIHW RI DVVRFLDWHG UD HTXLSHPHQW RXQFLO LUHFWLYH XURSHD...

Page 11: ...25 6 67 0 7 1 7 HQGH HQWDO 6 VWHPV Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 12: ...D B 14 2 7 CLOSING THE UNIT B 16 3 MOBILE SYSTEM MOUNTING B 17 3 1 ROOM PREPARATION B 17 3 2 UNPACKING B 17 3 3 ASSEMBLING THE MOBILE STAND B 18 3 4 INSTALLINGTHE FOLDINGARM B 19 3 5 INSTALLINGTHETIME...

Page 13: ...specified that each attachment point is able to support pull and thrust load of 288 Kg 636lb REMARK It is task of the installation engineer to choose the proper fixation hardware in relation to the t...

Page 14: ...formation it is recommended to use the THROUGH THE WALL method of attachment which makes use of an anterior rein forcement steel plate 2 mm thick and a steel coun ter plate 2 mm thick of dimensions la...

Page 15: ...ed sizes of the power cables wires and the relevant resistances meter are shown below recommended cable wires size resistance AWG 16 1 35 mm2 2 x 0 015 Ohm meter AWG 14 2 11 mm2 2 x 0 0094 Ohm meter I...

Page 16: ...a 2 INSTALLATION PROCEDURE 2 1 OPENING WALL SUPPORT AND TIMER Remove cover away from the Wall Support by unscrewing the screw a Unscrew the two locking screws b and remove cover of the Timer b 2 1 Pri...

Page 17: ...onotusetheTimeras a template Electronic components could be irreparablydamaged Drill the four pilot holes for the Wall Support with a 9 mm drill bit for the Timer with a 8 mm drill bit Position the Ti...

Page 18: ...ury to the patient or operator Install and tighten all screws with lock wash ers see page B 2 Wall mounting methods NOTE Ensure that Timer will be aligned with Wall Support Connect line cable on termi...

Page 19: ...terminal blocks See BZ 4 System electrical connections Verify the jumpers and dip switches setting as described in Section C Jumpers and dip switches setup of this manual NOTE Do not reinstall covers...

Page 20: ...nd verify the com pleteness according to items supplied a ring b ground wire with nut and lock washer c friction screws d snap out covers Cut the guide string WARNING do not pull string completely out...

Page 21: ...pplied con nect one side of the protective ground wire b to the ground point of the Extension Arm then con nect the other side to the earth screw of Wall Support Loosen two friction screws c of Extens...

Page 22: ...move binding of Folding Arm until indi cated in the instructions below Slide the chromium ring h over support pin b fitting the ring hole into locking pin k If necessary set sector i not supplied only...

Page 23: ...with the help of the guide cord Lightly grease the support pin By moving vertically carefully insert the support pin of the Folding Arm into the Extension Arm secure finally the screws c of the Extens...

Page 24: ...on mains supply and verify that the full line voltage is measured If no voltage is measured proceed as fol lows turn off the mains reverse the line wires connection turn on the mains again repeat the...

Page 25: ...rm is released without Tubehead and without opposing by hand to the springs action it is possible to damage the springs themselves Partially loosen the screw a and remove cover plate b Shift upwards h...

Page 26: ...ION Depressing ground contact f with a screwdriver Insert Tubehead pin into the Folding Arm Secure with U shaped retaining clip d ensur ing it is flush in slots f Tubehead Folding Arm d c 5 6 7 Printe...

Page 27: ...ehead Verify the integrity of the locking teeth h and the presence of the lock washer Finally install the collimator 2 7 CLOSING THE UNIT Before closing the unit perform the Setting to work actions as...

Page 28: ...socket shall be prepared within 1 5 m from the chair with the protective earth pin connected to the equipotential earth point of the room and with the line wires of the proper size NOTE The mains resi...

Page 29: ...are shaped in different thick Set the column on the pedestal insert the two exagonal screws M8 x 70 a and lock washers supplied WARNING The proper fixation of the column to pedestal is fondamental fo...

Page 30: ...locking pin e Pull down through the column the cable c of the Folding Arm to the mounting plate of the Timer Prepare the sector f on top of the column in the position indicated in the drawing WARNING...

Page 31: ...ns power of the column is securely connected at block terminal pin X3 3 the protective earth lead y g from the stud of the column is securely connected at block terminal pin X3 3 Verify the jumper and...

Page 32: ...ow Carefully plug in the option card on X10 and X11 connectors in the Densomat printed circuit board Set the dipswitch S1 2 from the OFF position single tubehead to the ON po sition double tubehead to...

Page 33: ...INSTALLATION B 22 Gendex Dental Systems ORALIXAC SYSTEM Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 34: ...ns voltage DVM routine C 5 5 2 4 Exposure Counter routine C 5 5 3 TESTAND DIAGNOSTIC ROUTINE SECONDENT C 6 5 3 1 Test of LEDs functionality Test mode routine C 6 5 3 2 Measure of mains voltage DVM rou...

Page 35: ...ited W9 Opened Not used Opened Not used W10 Closed Not used Closed Not used W11 Closed F3 shunted Opened F3 working Closed F3 shunted Opened F3 working S1 DENSOMAT SECONDENT 1 not used see table Tubeh...

Page 36: ...all balance in all positions and the torque required to rotate it shall not exceed 1 8 Nm in the horizontal plane or 1 5 Nm in the vertical plane If an adjustment of the frictioning of the vertical ro...

Page 37: ...all the test and the diagnostic routines X RAY emission is inhibited All segments of display light up then software version is visualized Ux x When the Test mode routine is active all LEDs of keyboard...

Page 38: ...the DVM routine Digital Voltmeter is initiated and UAC VAC will be displayed for a second The DVM routine is a test function used to check the mains voltage value measured by the Timer This value is s...

Page 39: ...e meas ured by the Timer This value is indicated on the operator panel as follows 1 If the voltage readout is 230 VAC the lower LED and the higher LED are lit Fig 1 2 If V 230 VAC upper LEDs are lit w...

Page 40: ...4 Press exposure button and verify that amber X ray ON indicator lights up and X ray buzzer signal is audible during exposure 5 Verify that audible and visible X ray ON indicators go OFF when exposure...

Page 41: ...the wait time the indication appears on the display Dead man function test 7 Press S button see figure 3 until the indication 2 5 appears on dis play 8 Press and immediately release exposure button v...

Page 42: ...5 RADIATION LEAKAGE D 3 3 6 EARTH RESISTANCE D 3 3 6 1 Wall mounted systems D 3 3 6 2 Mobile systems D 3 3 7 EARTH LEAKAGE D 4 DRAWINGS ACCEPTANCECHECKLIST DZ 1 1 INTRODUCTION This section contains i...

Page 43: ...preheating time Direct measurement can be performed with a digital mAmeter at the midpoint of the high voltage transformer secondary by shunting Tubehead resistor GA R5 Use a precision mAmeter f s 10m...

Page 44: ...n leakage measurements at 100 cm from the focus in any direction and verify that the dose rate values for the combination Tubehead and collimator are lower than 45 GY h 5 mR h as specified See Section...

Page 45: ...mA Mobile Wall mount off off on 2 0 5 off on on 2 0 5 on off on 2 0 5 on on on 2 0 5 off off off 2 2 off on off 2 2 on off off 2 2 on on off 2 2 V DIGITAL VOLTMETER INPUT IMPEDENCE 10 MOhm R 1 KOhm 1...

Page 46: ...1 3 Labels E 3 4 1 4 System mechanical performance E 3 4 2 MECHANICAL CHECKS E 4 4 2 1 Maintenance of the Tubehead E 4 4 2 2 Maintenance of the Extension Arm E 5 4 2 3 Maintenance of the Mobile Stand...

Page 47: ...ggested frequency is 24 months for each inspection of the Oralix system One man is needed for the execution of the preventive maintenance inspection 2 2 EQUIPMENT REQUIRED Digital DC voltmeter Standar...

Page 48: ...sible to damage the springs themselves 4 1 VISUAL INSPECTION 4 1 1 Cables connectors and plugs Verify that all cables are not damaged and properly fixed that no conductor lead is exposed or stressed C...

Page 49: ...le cover a of the Folding Arm is properly fixed and there is no access possibility to the Tubehead retaining clip b Partially loosen the screw g and remove the plate a and inspect the wires connection...

Page 50: ...at the side of the Wall Support and grease the spigot n Inspect for wear ring o and replace it if required Inspect for wear the cabling and check the protective earth connection 4 2 3 Maintenance of...

Page 51: ...hat the Wall Support is firmly connected to the wall by four fixation points the applied torque shall be 10 N m Check also that it is properly levelled Check the power connections at the terminal bloc...

Page 52: ...by the holes pro videdandthespringsthroughthe holes q provided Lift slightly the Folding Arm from its support and grease the spigot s Inspect the earth connections at points i and the respective bondi...

Page 53: ...e torque required to rotate it shall not exceed 1 8 Nm in the horizontal plane or 1 5 Nm in the vertical plane If an adjustment of the frictioning of the vertical rotation is required provide it as de...

Page 54: ...EDS indication over each push button are working properly 4 Check the adjustment of the mains measuring circuit according to Section C Ajustment of the mains voltage measuring circuit of this manual 4...

Page 55: ...Gendex Dental Systems E 10 4519 190 70943_4 ORALIXAC SYSTEM PREVENTIVE MAINTENANCE Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 56: ...RNING Always disconnect the equipment from the mains supply before starting any maintenance action the mains voltage may be present on parts of the system such as electronic board Tubehead Folding Arm...

Page 57: ...FOLDING ARM FRICTION For the adjustment of Folding Arm friction operate on the screws a shown in the figure below For the adjustment of friction of the horizontal rotation of the FoldingArm operate on...

Page 58: ...ushing d clockwise to increase spring tension counter clockwise to decrease tension Use special drift Replace covers pins and caps 3 3 TUBEHEAD FRICTION To adjust Tubehead friction at pivot axis proce...

Page 59: ...el by unscrewing the wheel nut a with special 30 mm open wrench Record the position of nut b on the old wheel Screw the nut b on the new wheel ensuring the same position of the nut In this way the lev...

Page 60: ...utine Densomat Run the DVM Digital Voltage Measure routine until the mains voltage is monitored on the display Adjust the trimmer R26 on the p c b in order to obtain on the display the same value meas...

Page 61: ...p l l i t s e r u l i a f e h f I d e y a l p s i d d r a o b 1 C T g n i h s a l f U u 0 1 e g n a r f o t u o e g a t l o v s n i a M g n i h s a l f g n i l o o c e b u t g n i r u d t s e u q e r...

Page 62: ...h t f I d e y a l p s i d d r a o b 1 C T g n i h s a l f d e l y d a e R g n i l o o c e b u t g n i r u d t s e u q e r e r u s o p x E h t o b e g a s s e m m r a l a e m i t g n i t i a w y b r o...

Page 63: ...e of charge basis Alternatively in some countries you may be able to return your products to your local retailer upon the purchase of an equivalent new product Disposing of this product correctly will...

Page 64: ...Gendex Dental Systems ORALIXAC SYSTEM TECHNICALDATA AZ 1 Appendix Appendice Appendice Ap ndice Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 65: ...ENSOMAT 9869 002 00102 SECONDENT TIMER MINUTERIE SECONDENT 9869 002 00202 FOLDING ARM BRACCIO ARTICOLATO BRAS ARTICULE BRAZO ARTICULADO 9869 001 00301 MOBILE STAND STATIVO MOBILE STATIF MOBILE SOPORTE...

Page 66: ...dimensioni sono espresse in millimetri NOTE toutes les dimensiones sont exprim es en millim tres NOTA todas las dimensiones son expresas en m limetros 615 DIMENSIONS FOR MOBILE STAND DIMENSIONIDELLOST...

Page 67: ...s con cono de 20 cm brazo de prolongacion de 90 cm DIMENSIONS FOR WALL MOUNTED SYSTEM DIMENSIONIPERILSISTEMAAPARETE DIMENSIONESPOURLESYSTEMEMONTEAMUR DIMENSIONESDELSISTEMADEPARED Recommended height Al...

Page 68: ...impulsions N di impulsi N d impulsions Tiempo de exposiciones N de impulsos N de impulsos Sec 50 Hz 60 Hz 0 03 2 3 0 05 3 4 0 06 4 5 0 08 5 6 0 1 6 7 0 12 7 8 0 16 9 10 0 2 11 13 0 26 14 17 0 32 17 2...

Page 69: ...ORALIXAC SYSTEM INSTALLATION Gendex Dental Systems BZ 1 TEMPLATE FOR WALL SUPPORT 184 mm 90 mm Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 70: ...INSTALLATION BZ 2 Gendex Dental Systems ORALIXAC SYSTEM TEMPLATE FOR TIMER DENSOMATAND SECONDENT Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 71: ...XTENSION BRAZO DE PROLONGACION 2 Snap out cap Coperchi Cabouchons pression Capuchones 1 Screw Viti Vis Tornillo FOLDING ARM BRACCIO ARTICOLATO BRAS ARTICULE BRAZO ARTICULADO 1 Sector Settore Secteur S...

Page 72: ...ems ORALIXAC SYSTEM WALL MOUNTED SYSTEM ELECTRICAL CONNECTIONS SISTEMA A PARETE COLLEGAMENTI ELETTRICI SYSTEME MURAL CONNEXIONS ELECTRIQUES EQUIPO DE PARED CONEXIONES ELECTRICAS Printed on 22 M r 2012...

Page 73: ...endex Dental Systems BZ 5 MOBILESYSTEM ELECTRICALCONNECTIONS SISTEMA MOBILE COLLEGAMENTI ELETTRICI SYSTEME MOBILE CONNEXIONS ELECTRIQUES EQUIPO MOVIL CONEXIONES ELECTRICAS Printed on 22 M r 2012 09 28...

Page 74: ...osure time setup lower teeth Tasti per l impostazione tempo esposizione denti inferiori Touches pour la s lection du temps d exposition dents inf rieures Teclas para la disposicion del tiempo de expos...

Page 75: ...g diagnosis indicator Indicatore programmazione diagnosi Indicateur de programmation diagnostic Indicator de programacion diagnosi Tubehead being enabled Monoblocco abilitato Monobloc habilit Monobloq...

Page 76: ...rete Indicateurs du temps d exposition s lectionn tension de r seau Indicadores del tiempo de exposicion seleccionado tension de red System ON indicator Indicatore di sistema acceso e in funzione Indi...

Page 77: ...Indicadores Advertencias R 3 2 ACCURACY OF kVpk Verifica dell alta tensione di picco Precision de kVpk Precision de la kVpk R 3 3 ACCURACYOFTUBECURRENT Verifica della corrente del monoblocco Precisio...

Page 78: ...THE WALL SUPPORT S 4 2 5 MAINTENANCE OF THEARM 4 2 6 SYSTEMADJUSTMENT 4 3 1 POWER SUPPLIES 4 3 2 INSPECTION OF THE TIMER 4 3 3 EXPOSURE COUNTER READING 5 1 PERFORMANCE TEST 5 2 CUSTOMER TEST LABELS IN...

Page 79: ...E DELBRACCIOARTICOLATO 4 2 6 REGOLAZIONI MECCANICHE 4 3 1 ALIMENTAZIONI 4 3 2 CONTROLLO DEL TIMER 4 3 3 LETTURADELCONTATORE DI ESPOSIZIONI 5 1 PROVADELLE PRESTAZIONI 5 2 PROVE DIMOSTRATIVEAL CLIENTE T...

Page 80: ...RASARTICULE 4 2 6 REGLAGES DE SYSTEME 4 3 1 ALIMENTATION 4 3 2 INSPECTION DE LA MINUTERIE 4 3 3 RELEVE DU COMPTEUR D EXPOSITIONS 5 1 ESSAI DES PERFORMANCES 5 2 ESSAIS DEMONSTRATIFSAU CLIENT Etiquettes...

Page 81: ...ICULADO 4 2 6 AJUSTES MECANICOS 4 3 1 ALIMENTACION ELECTRICA 4 3 2 INSPECCION DEL TIMER 4 3 3 LECTURADELCONTADOR DE EXPOSICIONES 5 1 PRUEBA DE FUNCIONAMIENTO 5 2 PRUEBAS PARAEL USUARIO Etiquetas indic...

Page 82: ...il in the tubehead Therefore disposal of discarded spare parts must be done via a company specialized in industrial wastes Oralix contiene delle sostanze nocive all ambiente in particolare circa 1Kg d...

Page 83: ...chous pression Casuetes a presi n Q2 d Coupling with pin Accoppiatore Coupleur C 4519 100 71772 Acoplador Q2 e Friction ring Anelli di frizione D 4519 100 27862 Bagues d embrayage Aros de rozamiento Q...

Page 84: ...largo 1 2 3 4 5 DESCRIPTION DESCRIZIONE CODE CODICE DESCRIPTION DESCRIPCION CODE CODIGO Q4 a GA Handle Manicotto Poign e Empunadura 4519 120 78511 Q4 c1 Cover Coperchio Couvercle Tapa 4519 124 00881...

Page 85: ...TUBEHEAD ORALIXAC SYSTEM Q 4 Gendex Dental Systems b1 b2 e1 e2 b3 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 86: ...E CODE CODICE DESCRIPTION DESCRIPCION CODE CODIGO Q5 a Snap out cap Tappo Couvercle tapas a presion 2x 4519 120 73852 Q5 b Special screw Vite speciale Vis special 2x 4519 120 47863 Tornillo de rozamie...

Page 87: ...RT SUPPORTO MURALE SUPPORT MURAL SOPORTE DE PARED Type 9869 005 00001 a 1 2 3 4 5 DESCRIPTION DESCRIZIONE CODE CODICE DESCRIPTION DESCRIPCION CODE CODIGO Q6 a Cover Coperchio Couvercle Tapa 4519 128 0...

Page 88: ...089 01036 Q7 b Flat cable Cavo piatto Cable plat cable chato 4519 103 01551 Q7 c Exposure handswitch without cable 4519 190 01461 Comando raggi senza cavo Interrupteur manuel rayons sans fil Commutado...

Page 89: ...thout cable 4519 190 01461 Comando raggi senza cavo Interrupteur manuel rayons sans fil Commutador de mano sin cable Q8 d Coiled cable Cordone spiralato Cordon spiral 4519 190 01511 Cable enrollado Q8...

Page 90: ...am Secondent Z 4 Schemi di principio Secondent Principle diagramme Secondent Esquema funcionale Secondent TC1 Power Densomat Z 5 TC1 Scheda Densomat TC1 Minuterie Densomat TC1 Densomat TC1 Power Secon...

Page 91: ...ACCEPTANCE ORALIXAC SYSTEM Gendex Dental Systems Z 2 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 92: ...ELECTION TUBE 1 MAINS1 MAINS 2 200 2 1 4 OPTO COUPLER X RAY ON TUB E 1 COU PLER OPTO COU PLER OPTO READY T UB E1 X RAY ON T UB E2 READY TUB E2 FUS E FUS E FUSE FUS E 24 48V N L N X RAY ON T UB E 1 22...

Page 93: ...LIXAC SYSTEM Gendex Dental Systems Z 4 PRINCIPLE DIAGRAM SECONDENT SCHEMA DI PRINCIPIO SECONDENT PRINCIPLE DIAGRAMME SECONDENT PRINCIPLE DIAGRAM SECONDENT 5 Printed on 22 M r 2012 09 28 10 am Printed...

Page 94: ...Z 5 ORALIXAC SYSTEM ACCEPTANCE Gendex Dental Systems TC1 Power Densomat 4519 101 02203 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 95: ...ACCEPTANCE ORALIXAC SYSTEM Gendex Dental Systems Z 6 TC1 Power Secondent 4519 101 02903 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 96: ...Z 7 ORALIXAC SYSTEM ACCEPTANCE Gendex Dental Systems TC5 Tubehead 2 4519 101 02701 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Page 97: ...ACCEPTANCE ORALIXAC SYSTEM Gendex Dental Systems Z 8 TA3 Encoder control unit 4519 101 02801 Printed on 22 M r 2012 09 28 10 am Printed by TOBIAS BAUER...

Reviews: