Mechanical Separation /
GEA Westfalia Separator
4.6
Alarm #08 - LAHH sludge tank
LAHH – SLUDGE TANK
Sludge tank - full
Separation “Stop”
Alarm#08 ENTER
Text display
Possible causes
Action
Sludge pump does not function.
•
Check sludge function.
Bowl open.
•
Check bowl hydraulics.
Product overflow due to excessively
high clean oil discharge pressure.
•
Check valves in the clean oil line
and position V7.
Closing water pulse T06 too long.
•
Check T06.
Polyamide gasket in bowl top defec-
tive.
•
Check polyamide gasket in bowl
top; replace if necessary.
Gaskets in bowl defective.
•
Check gaskets.
Level switch LSHH soiled or defective.
•
Check level switch LSHH.
Help text – Information on control system
E N T E R
Step 1
Acknowlede fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
LSHH – SLUDGE TANK
Please check:
Sludge pump
Bowl hydraulics
->
Seite 2
Valves in the
clean oil line
Position V7
Closing water pulse
->
Page 3
Polyamide gasket
Switch LSHH
Back with ENTER
F 1