GE H
EALTHCARE
RAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
6-6
Section 6-4 - Monitor Arm Adjustment
Section 6-4
Monitor Arm Adjustment
1.) If the “outdated” monitor arm (KTZ154817) with mounted monitor moves self-acting, retighten the
Allen screw on the upper part of the monitor arm.
2.) If the monitor tilts downward, retighten the cross-head screw on the swivel head of the monitor arm.
Figure 6-6 retighten screws
1
2
Attention:
Do not damage the
plastic with the tool.