Connections
6.2.3
Power Supply (for Auxiliary Equipments)
For auxiliary equipment
ST1-ST5
Mains OUT for auxiliary equipment
These outlets are fed from an isolation transformer which is switched with the F2 switch of the rear
side. This outlet voltage is not depending on the mains voltage.
CAUTION!
The sum of the power consumption of equipment connected to these outlets must
not exceed 350 VA (incl. the color video monitor)!
CAUTION:
The outlets ST1 – ST5 may be set to 230V or 115V
Modification of voltage setting only by an authorized service person!
Voluson
®
730 - Instruction Manual
105838 Rev. 3
6-7