Chapter 2 - Site Preparations
2-7
P R E L I M I N A R Y
2-3-2
Mandatory Site Requirements
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in section
(see
below).
•
A dedicated “hospital-grade” single branch power outlet of adequate amperage (see
page 2-3.
) that meets all local and national codes and is located less than 2.5 m (8.2 ft) from the
unit’s proposed location. Refer to the
Electrical Requirements section on page 2-3.
•
A door opening of at least 0.54 m (1.77 ft) in width.
•
The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
•
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•
Power outlets for other medical equipment and gel warmer.
•
Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
•
Clean and protected space for storage of probes (either in their case or on a rack).
•
Material to safely clean probes (performed using a plastic container, never metal).
•
In the case of a network option:
•
An active network outlet in the vicinity of the ultrasound unit.
•
A network cable of appropriate length (regular Pin-to-Pin network cable).
•
An IT administrator who will assist in configuring the unit to work with your local network. A fixed
IP address may be required when using DICOM. Refer to the form provided in
page 2-13
for network details that are required.
Note:
All relevant preliminary network outlets installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
2-3-3
Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
•
Door opening of 92 cm (3 ft) in width.
•
Accessible circuit breaker for a dedicated "hospital-grade" power outlet.
•
Sink with hot and cold running water.
•
Receptacle for bio–hazardous waste, for example, used probe sheaths.
•
Emergency oxygen supply.
•
Storage area for linens and equipment.
•
Nearby waiting room, lavatory, and dressing room.
•
Dual-level lighting (bright and dim).
•
Lockable cabinet for software and manuals.
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......