Section 8-5 - Electronic Cage Components - Replacement Procedures
P R E L I M I N A R Y
8-5-12
CFE Release Arm Replacement Procedure
8-5-12-1
Tools
Appropriate Phillips screwdriver.
8-5-12-2
Time Required
15 minutes
8-5-12-3
Preparation
Shut the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
.
8-5-12-4
CFE Board Release Arm Removal Procedure
The lever-shaped CFE Release Arm is located on the PSB Board - see
.
1) Remove the CFE Board:
2) Uninstall the CFE Release Arm (Lever) by unscrewing the two securing screws as indicated
in
Figure 8-142, above
.
8-5-12-5
CFE Release Arm Installation Procedure
1) Install a new CFE Release Lever in position on the PSB and secure with the two securing screws
previously removed.
2) Refit the CFE Board:
FRU Part #
Figure 8-142 Location of CFE Release Arm
•
CFE Board Removal Procedure on page 8 - 112
•
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......