Section 8-5 - Electronic Cage Components - Replacement Procedures
P R E L I M I N A R Y
8-5-8
CPU Partition Replacement Procedure
NOTE:
The CPU partition is located on the rear side of the BEP- see
8-5-8-1
Tools
Appropriate Phillips and flat screwdrivers.
8-5-8-2
Time Required
20 min
8-5-8-3
Preparations
Shut down the Vivid S60/Vivid S70 ultrasound unit, as described in
.
8-5-8-4
CPU Partition Removal Procedure
1.) Remove the BEP:
2.) Place the BEP on a clean work surface with the CPU Partition facing
upwards
.
3.) Remove the two securing screws (shown in
Figure 8-129
)
that attach the CPU Partition to the BEP.
4.) Gently slide the CPU Partition upwards and remove it from the BEP.
Figure 8-129 CPU Partition on BEP
FRU Part #
•
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......