Chapter 3 - System Setup
3-5
Section 3-3
Receiving the
Vivid q N
3-3-1
Examine All Packages
Examine all packages closely at time of delivery, as described in the procedure below.
Table 3-4
Examine All Packages
STEP
TASK
ILLUSTRATIONS
1
Is damage apparent?
•
If
yes
; continue with the instructions in subsection
3-3-2 - Damage in Transportation
.
•
If no; continue with step 2.
2
Is the Shock Indicator red colored inside the middle of the
indicator?
•
If
yes
: The Shock Indicator has been triggered.
Make a remark on the Post Delivery Checklist about the
triggered indicator before you continue with step 2.
•
If
no
: Continue with step 2.
3
Is the Tilt Indicator red colored inside the middle of the
indicator?
•
If
yes
: The Tilt Indicator has been triggered.
Make a remark on the Post Delivery Checklist about the
triggered indicator and then follow the rest of the
instructions in subsection
•
If
no
: Continue with the instructions in
.
WARNING
WARNING
For Vivid q N portable ultrasound scanners supplied from the factory with a dummy
battery, when unpacking the unit, do not discard the dummy battery.
When charging the Vivid q N lithium ion battery on an external charger, since the
battery in fact forms one of the four legs of the scanner, removing it will leave the unit
standing unbalanced. For this reason, the dummy battery should be kept, so it may be
inserted in position to provide stability to the scanner while the lithium ion battery is
being charged.
In addition, use of the dummy battery is recommended during transportation or
long-term storage of the Vivid q N portable ultrasound scanner.
RED COLOR
RED COLOR