GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-3
3.) Using both hands, pull the Bearing Handle slightly apart to release it from the system.
4.) Remove the Bearing Handle.
8-2-1-5
Bearing Handle Installation Procedure
1) Place a new Bearing Handle in position, using both hands to pull the Bearing Handle slightly apart
in order to slot the ends into the handle holes on the Vivid q N casing.
2.) Carefully fasten each captive screw on either side of the Bearing Handle to secure the handle in
position, by turning 1/4 turn in a
clockwise
direction.
NOTE:
When tightening the captive screws, do not use excessive force do not apply more than
1/4 turn; overtightening will damage the screw thread.
3.) Proceed to perform the
"Bearing Handle Test" on page 4-19
Figure 8-3 Bearing Handle Removed