Section 8-13 - Loading Base Image Software
Section 8-13
Loading Base Image Software
FRU part num for Base Image software.
5344308-3 Vivid P3 Application Software with Base Image for Expert.The software DVD is kept in the
Rear EMI cover of the system.
NOTE:
While it is believed to be unnecessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE:
To Enter BIOS Press F2 (Arrow) Key immediately after power on and enter BIOS Password for
LP3 as Lakshya.
NOTE:
Please ensure AC adapter is connected during system upgrade!
1.) Before loading the Software, backup the database. For reference see
and
after loading the Software, restore the backup database.
2.) Insert the disk labeled “System & Application Software” into the DVD drive.
3.) Properly turn off the scanner by momentarily pressing the
Power On/Off
Switch. Select “Full
Maintenance Reboot” from the System Exit menu.
4.) If the system will not shutdown normally, hold down the
Power On/Off
Switch until the light
turns from green to amber.
5.) Turn on the scanner. System will detect the DVD-RW automatically.
Figure 8-72 Shutdown Dialog Box