background image

GE H

EALTHCARE

D

IRECTION

 FQ091013, R

EVISION

 1 

 

V

IVID

 

I

 

N

 

AND

 V

IVID

 

Q

 N S

ERVICE

M

ANUAL

Chapter 7 - Diagnostics/Troubleshooting

 7-151

7-7-4

Image Quality Page

When the 

Image Quality

 page is selected, as shown below, you can verify and calibrate image quality.

Figure 7-118   Image Quality Page

7-7-5

Calibration Page

Figure 7-119   Calibration Page

Note:

This page is not populated in this version.

When the 

Calibration

 page is selected (as shown in 

Figure 7-119

), although not populated, you can 

perform different calibration procedures, as described in 

Chapter 6 - 

.

For additional calibration procedures, refer to the following sections:

DC Offset Calibration

-

see 

DC Offset Calibration on page 7 - 25

.

DC Offset Verification

-

see 

DC Offset Verification on page 7 - 28

.

Summary of Contents for Vivid in

Page 1: ...VQN_SVC_FRNT_CVR FM P R E L I M I N A R Y Vivid i n and Vivid q N Service Manual Vivid q N with software version SW 11 x x BT11 Vivid i n with software version SW 9 x x BT 09 Vivid i n with software v...

Page 2: ......

Page 3: ...CE MANUEL DANS UNE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS...

Page 4: ...A DO T CNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A G...

Page 5: ...MUUTA KUIN ENGLANNINKIELIST MATERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T...

Page 6: ...KA POSKYTOVATELE SLU EB OBSLU N HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS...

Page 7: ...ERATORIUS AR PACIENTAS GALI B TI SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET KUNDENS...

Page 8: ...CE IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUP...

Page 9: ...RVISERA OPERATERA ILI PACIJENTA PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK A...

Page 10: ...IR E ER M TER TEKN SYEN KILAVUZUN NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU...

Page 11: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL ix 2 NOV 2011 FRONTMATTER FM P R E L I M I N A R Y ZH CN KO...

Page 12: ...ecially trained field engineers All of GE s electrical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qual...

Page 13: ...may not be copied or duplicated in any form in whole or in part without prior written permission of GE Healthcare GE Healthcare may revise this publication from time to time without written notice TR...

Page 14: ...History List of Effected Pages LOEP REVISION DATE REASON FOR CHANGE 1 2011 NOV 02 Initial release Pages Revision Pages Revision Pages Revision Title Page 1 1 1 to 1 26 1 7 1 to 7 176 1 Warnings i to...

Page 15: ...tant Conventions 1 5 Conventions Used in this Manual 1 5 Safety Considerations 1 7 Introduction 1 7 Human Safety 1 7 Mechanical Safety 1 7 Electrical Safety 1 9 Dangerous Procedure Warnings 1 11 Produ...

Page 16: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xiv Contact Information 1 25 System manufacturer 1 26...

Page 17: ...Requirements 2 2 Cooling Requirements 2 2 Lighting Requirements 2 2 Time and Manpower Requirements 2 2 Electrical Requirements 2 3 EMI Limitations 2 4 Probe Environmental Requirements 2 5 Facility Nee...

Page 18: ...ation box Contents 3 12 Physical Inspection 3 12 EMI Protection 3 12 Preparing for Installation 3 13 Confirming Customer Order 3 13 Verifying the Transportation Box Contents 3 13 Component Inspection...

Page 19: ...ty Configuration 3 98 Set Up Connection to a DICOM Server in a Network 3 110 Setup Connection to HL7 Server in a Network 3 117 Query Retrieve Q R Setup 3 120 MPEGVue Export Configuration 3 127 eVue Da...

Page 20: ...w 4 3 Power ON Boot up 4 3 Power Shut Down 4 7 Log On to the System as ADM 4 10 Using Removable Media 4 12 Labeling Removable Media 4 15 Formatting Removable Media 4 15 Verifying Removable Media 4 16...

Page 21: ...iver 5 15 Probe and MUX P M Board 5 18 Back End Processor 5 21 Introduction 5 21 ETX SBC Central Processing Unit CPU 5 23 ETX Base Board 5 23 Hard Disk 5 25 Fan Control 5 25 RTC BIOS Battery 5 25 Back...

Page 22: ...Q N SERVICE MANUAL xx ECG Patient I O Module Specifications 5 40 Isolation 5 40 ECG Respiratory Module 5 41 Internal ECG Respiratory Board 5 41 External ECG Respiratory Interface 5 43 Peripherals 5 46...

Page 23: ...alibration 6 2 Contrast Adjustment Procedure 6 2 Brightness Adjustment Procedure 6 3 Testing Your Contrast and Brightness Settings 6 3 Blue Tint Adjustment Procedure 6 3 Blue Tint 2 Adjustment Procedu...

Page 24: ...ing the Front End Diagnostic Options 7 25 Calibration Tests 7 25 Radio Frequency Interface RFI Diagnostic Tests 7 31 TR 32 Diagnostic Tests 7 46 Probe and MUX Diagnostic Tests 7 61 Monitoring Tests 7...

Page 25: ...8 38 BIOS Battery Replacement 8 45 HVPS Replacement Procedure 8 46 Fan Replacement Procedure 8 49 Speaker Assembly Replacement Procedure 8 52 LCD Display Replacement Procedure 8 53 LCD Flex Cable Repl...

Page 26: ...etting the BIOS 8 108 BIOS Firmware Update 8 118 Formatting and Partitioning the Hard Disk Automatic Procedure 8 121 Formatting and Partitioning the Hard Disk Manual Procedure 8 123 Full Re ghost and...

Page 27: ...e of Chapter 9 9 1 List of Abbreviations 9 2 Mechanical Hardware Parts 9 3 Electronic Parts 9 4 Cables 9 7 Software 9 8 Probes 9 9 Peripherals 9 11 SafeLock Cart Parts 9 12 Optional Modo Cart 9 14 Viv...

Page 28: ...ional Checks 10 10 Input Power Checks 10 11 Cleaning 10 12 Physical Inspection 10 15 Optional Diagnostic Checks 10 16 Probe Maintenance 10 17 Probe Checks 10 17 Probe Handling 10 17 Basic Probe Care 1...

Page 29: ...RE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxvii Excessive Current Leakage 10 33 Possible Causes of Excessive Current Leakage 10 33 Vivid i n Vivid q N Inspection Certific...

Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii...

Page 31: ...q N portable ultrasound scanner The service provider must read and understand all the information presented here before installing or servicing a unit Table 1 1 Contents in Chapter 1 Section Descripti...

Page 32: ...ck diagrams and functional explanations of the electronic circuits Chapter 6 Service Adjustments Contains instructions for performing service adjustments to the Vivid i n and Vivid q N portable ultras...

Page 33: ...Covered in this Manual The Vivid q N models documented in this manual is shown in Table 1 3 below Table 1 2 Vivid i n Models Model H Cat No Part Number Description Comments Vivid i n Portable Ultrasou...

Page 34: ...The Vivid q N ultrasound imaging scanner is similar to the Vivid i however newly designed to offer additional features including the ability to support use of M4S RS probes Also weighing only 5 Kgs 11...

Page 35: ...s of safety precautions are found on the equipment and throughout this service manual Different levels of severity are identified by one of the following icons which precede precautionary statements i...

Page 36: ...chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific pr...

Page 37: ...IS TO BE MOVED ALONG ANY INCLINE USE EXTREME CAUTION MAKE SURE THAT THE VIVID I n or Vivid q N SCANNER AND ALL PERIPHERALS ARE SECURELY MOUNTED ON THE CART BEFORE ATTEMPTING TO MOVE IT DANGER ULTRASO...

Page 38: ...n or Vivid q N system and SafeLock cart if applicable in a vehicle ALWAYS Eject any disks from the MOD if installed Ensure that the Vivid i n or Vivid q N system is well prepared and packed in its or...

Page 39: ...e cables to become entangled or to be caught in the SafeLock Cart wheels NOTE For detailed information on handling endocavity probes and invasive intracardiac probes refer to the appropriate supplemen...

Page 40: ...s carried out It is very difficult to attach unique dimensions to the PATIENT ENVIRONMENT In practice a distance of 2 5 m 8 2 ft above the floor on which the medical personnel stand and a horizontal d...

Page 41: ...THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIR...

Page 42: ...cons These represent pertinent information regarding the operation of the unit 1 4 1 Universal Product Labels The following diagrams illustrate the labels found on the Vivid i n and Vivid q N ultrasou...

Page 43: ...LTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 1 Introduction 1 13 First version of Main Label Table 1 7 Main Label for Vivid i n cont d 2 of 2 DESCRIPTION ILLUST...

Page 44: ...Consult accompanying documents is intended to alert the user to refer to the User Manual or other instructions when complete information cannot be provided on the label Various CAUTION Dangerous volt...

Page 45: ...njury that may be caused by the weight of the machine if one person attempts to move it considerable distances or on an incline Used in the Service and User Manual which should be adjacent to equipmen...

Page 46: ...SION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 1 16 Section 1 4 Product Labels and Icons Date of manufacture On the base of the unit Table 1 8 Label Icons and Symbols Description and Location cont d 3...

Page 47: ...not heat the battery or discard it in a fire Do not expose the battery to temperatures over 60o C 140o F Keep the battery away from fire and other heat sources Do not charge the battery near a heat s...

Page 48: ...ze Li lon smart packs Important For Vivid i n and Vivid q N portable ultrasound scanners supplied from the factory with a dummy battery when unpacking the unit do not discard the dummy battery When ch...

Page 49: ...labels described in the previous section additional labels may be found on the Vivid i n and Vivid q N ultrasound unit as described in the following section Rating Labels section see below 1 4 4 1 Rat...

Page 50: ...ID I N AND VIVID Q N SERVICE MANUAL 1 20 Section 1 4 Product Labels and Icons 1 4 5 SafeLock Cart Labels The labels shown in the example below are located on the rear side of the cart Figure 1 2 Vivid...

Page 51: ...ances in the electrical power supply 1 5 2 CE Compliance The Vivid i n and Vivid q N portable ultrasound scanner conforms to all applicable conducted and radiated emission limits and to immunity from...

Page 52: ...Used Scope Standard Directive Product safety requirements for Electrical Medical devices Class I Type B with BF and CF applied parts ICE catheter and ECG leads IEC 60601 1 EN 60601 1 UL 60601 1 C 22...

Page 53: ...gout Devices 5 Control all stored and residual energy 6 Verify isolation All potentially hazardous stored or residual energy is relieved NOTICE Energy Control and Power Lockout for Vivid i n and Vivid...

Page 54: ...ids must be properly removed from any part or equipment prior to shipment GE Healthcare employees as well as customers are responsible for ensuring that parts equipment have been properly decontaminat...

Page 55: ...ere applicable save the appropriate log files by pressing Ctrl F or Alt D Remember to save the log files for each day on a separate floppy disk labelled accordingly Note Restart the application before...

Page 56: ...VID Q N SERVICE MANUAL 1 26 Section 1 8 Customer Assistance 1 8 2 System manufacturer Table 1 11 System manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE VINGMED ULTRASOUND A S STRANDPROMENADEN 45...

Page 57: ...d i n or Vivid q N ultrasound unit Included are descriptions of the electrical and facility requirements that must be met by the purchaser A worksheet is provided at the end of this chapter see Figure...

Page 58: ...dued and indirect when a scan is being performed Therefore a combination lighting system dim bright is recommended Keep in mind that lighting controls and dimmers can be a source of EMI which could de...

Page 59: ...s Electrical specifications for the Vivid i n Vivid q N system are as follows 2 2 5 2 Inrush Current Inrush current is not a factor for consideration due to the inrush current limiting properties of t...

Page 60: ...uencies magnetic fields and transients in the air or wiring They also generate EMI The Vivid i n and Vivid q N ultrasound units comply with limits as stated on the EMC label However there is no guaran...

Page 61: ...external connection requires a magnet wrap at each end Install the shield over the front of the card cage Loose or missing covers or RF gaskets allow radio frequencies to interfere with the ultrasound...

Page 62: ...nd calibration and testing must also be performed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence...

Page 63: ...in their case or on a rack Material to safely clean probes performed using a plastic container never metal In the case of a network option An active network outlet in the vicinity of the ultrasound u...

Page 64: ...low shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements Figure 2 1 Recommended Floor Plan 4 3m x 5 2m 14 ft x 17 ft 36 IN 9...

Page 65: ...ferring images in this manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier an...

Page 66: ...ices Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Vivid i n SN Type Floor Room REV Comments CONTACT INFORMATION Name Title...

Page 67: ...Network Information Vivid i n or Vivid q N Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM AP...

Page 68: ...ence of any variances from the basic installation Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying ope...

Page 69: ...or checking and testing the unit probes and external peripherals for electrical safety are also provided NOTE Depending on the customer s specific requirements the Vivid i Vivid q N portable ultrasoun...

Page 70: ...Only qualified service personnel should carry out servicing and troubleshooting 3 2 2 1 System Acclimation Time Following transport the Vivid i n Vivid q N system may be very cold or hot Allow time fo...

Page 71: ...e lifted CAUTION If the unit is very cold or hot do NOT turn ON power to the unit until it has had sufficient time to acclimate to its operating environment CAUTION To prevent electrical shock connect...

Page 72: ...ckage even if the package itself is undamaged To make it easier to detection if the handeling during transportation has been unproper a set of Tilt Shock indicators have been attached to the transport...

Page 73: ...ator has been trigged Make a remark on the Post Delivery Checklist about the trigged indicator and then follow the rest of the instructions in subsection 3 3 2 Damage in Transportation If no Continue...

Page 74: ...very is accepted or signed for by a GE representative or hospital receiving agent 2 Report the damage to the carrier Whether noted or concealed damage MUST be reported to the carrier immediately upon...

Page 75: ...q N Transportation Box Label details Figure 3 2 Label mounted Vivid i n label is sshown in this example Figure 3 3 Vivid q N Transportation Box Label details HANDLE WITH CARE FRAGILE HANDLE WITH CARE...

Page 76: ...N 3 3 2 3 The Vivid i n Transportation Box Label details Figure 3 4 Vivid i n Transportation Box Label details HANDLE WITH CARE FRAGILE HANDLE WITH CARE KEEP DRY PROTECT FROM MOISTURE TOP UPRIGHT TRAN...

Page 77: ...L Chapter 3 System Setup 3 9 Section 3 4 Unpacking the Equipment CAUTION Please read this section fully before unpacking the Vivid i n and Vivid q N ultrasound unit Figure 3 5 Shipping Box for the Viv...

Page 78: ...ckages Examin the Transportation box closely at time of delivery as described in the procedure that follow Figure 3 7 Transportation box lid removed Table 3 6 Shipping Carton Dimensions and Weights De...

Page 79: ...ady to mark the check list 3 Remove the surface packing material paper pad silica gel etc Continue to remove additional packing material as applicable during the following procedure steps 4 Carefully...

Page 80: ...voltage The Voltage settings are 220 240V AC 3 4 5 EMI Protection The Vivid i n Vivid q N Ultrasound Unit has been designed to minimize the effects of Electro Magnetic Interference EMI Many of the co...

Page 81: ...ist the contents of the shipping carton that are additional to the Vivid i n and Vivid q N These include external accessories and probes as well as a regional language support kit and optional periphe...

Page 82: ...ged items Verify the proper illumination of all the control panel buttons 3 Probes Check all probes for wear and tear on the lens cable and connector Look for bent or damaged pins on the connector and...

Page 83: ...lean the peripherals in accordance with the manufacturer s directions To prevent EMI or system overheating dress the peripheral cables inside the peripheral cover 5 AC System Check the AC board connec...

Page 84: ...are visible from the front and side of the ultrasound unit 1 Display Monitor tilts up and down 2 Speakers Two loudspeakers for Doppler sound 3 Control Panel Contains the alphanumeric keyboard and the...

Page 85: ...ultrasound unit rear panel and external peripheral accessory connectors 1 Two interchangeable USB ports digital printer CD RW and other peripherals 2 Docking connector currently not in use 3 Port for...

Page 86: ...i n Vivid q N SafeLock Cart Components Label Item Label Item 1 Top Shelf Assembly with handle 8 Bottom Cover Rear AC Assembly Cover 2 Probe Shelf with probe holders 9 Rear Lower Cover 3 LAN Isolation...

Page 87: ...afety rating plate step 1 3 5 4 1 Connection and Usage of the optional Modo Cart Please refer to the Vivid i n Cart User Manual Supplement to Vivid i n User manual Direction Number FL092096 3 5 5 Syst...

Page 88: ...d unit will have noisy images Check the grounding of the power cord and power outlet Be aware of RF sources Keep the unit at least 5m 16 4 ft away from other EMI sources Special shielding may be requi...

Page 89: ...i n Vivid q N ultrasound scanner as follows Printer Categories and Outline of Procedure Instructions below Configuring Hot Keys to Activate Printing Direct from the Control Panel on page 3 22 Configur...

Page 90: ...re either the Print button or Alt Print to activate a variety of outputs 5 Use the left or right arrows below to add any of the available options listed on the left as Outputs listed on right 6 From t...

Page 91: ...first accesses the system at the Applications level from the Additional Outputs screen By selecting the Advanced option under Selected Devices the user accesses the Printer Properties dialog box and...

Page 92: ...eting the Hardware Installation 3 Click Configure in the Printer Properties dialog to enter Print Setup 4 In the Orientation area select Portrait or Landscape as needed 5 From the Paper Size drop down...

Page 93: ...6 Click Properties The printer properties screen opens Figure 3 17 7 Verify that the selected orientation and paper size are correct 8 Click OK twice and then once more to return to the Advanced Outpu...

Page 94: ...al non isolated power source WARNING Do not attempt to use a different type of DeskJet Color Printer brand or model other than the DeskJet Color Printer provided by GE Medical Systems The ultrasound s...

Page 95: ...attempt to use a different type of DeskJet Color Printer brand or model other than the DeskJet Color Printer provided by GE Medical Systems The ultrasound system is an extremely sensitive and complex...

Page 96: ...ion procedure for mounting the Vivid i q system on the SafeLock cart as described in Mounting the Vivid i n Vivid q N on the SafeLock Cart optional on page 3 65 NOTE When using the Vivid i q system mo...

Page 97: ...d with the specific B W Printer supplied by GE Medical Systems WARNING Do not attempt to use a different type of B W Printer brand or model other than the B W Printer provided by GE Medical Systems Th...

Page 98: ...ick the Tcpip tab on the Connectivity screen see Figure 3 14 on page 3 23 2 Select Advanced Settings see Figure 3 22 3 The Network Connection screen opens Figure 3 23 Click the Up button to open the W...

Page 99: ...D Q N SERVICE MANUAL Chapter 3 System Setup 3 31 5 From the menu select Printer Properties 6 The Sony UP D897 Properties screen opens Select Printing Preferences Figure 3 24 Sony UP D897 printer confi...

Page 100: ...the Hardware Installation 7 From the Printing Preferences dialog under Orientation A Under Paper click the scroll down arrow and select the required paper size B Under Orientation choose the required...

Page 101: ...ed to be used with the specific Digital Color Printer supplied by GE Medical Systems WARNING Do not attempt to use a different type of Digital Color Printer brand or model other than the Digital Color...

Page 102: ...ed for additional power The modem should be connected to the lower USB port on the rear connectors panel When connecting the modem there is no preference as to which cable to connect first One end of...

Page 103: ...e Vivid i n Vivid q N system The installation procedure requires connecting the DVD CD drive The procedure includes installing the Compact Wireless G USB Adaptor software and then the configuration pr...

Page 104: ...vent the system will stop recognizing the CD DVD drive or any other device connected to this USB port A complete system re boot will then be required to make the USB port operational again The Samsung...

Page 105: ...ivid i n Vivid q N system To perform the Database Backup procedure in order to store back up copies of patient Exams data Archive and also to create free space on the local hard disk To export Exams i...

Page 106: ...msung DVD CD RW drive rear panel b Disconnect the 5V power supply output cable from the power port on the Samsung DVD R RW drive rear panel 2 Remove the Samsung DVD CD RW drive from the system Install...

Page 107: ...ollowing tests a Insert a CD R disk in the DVD CD RW Drive b Eject and remove the CD R disk to check the eject mechanism functions properly NOTE In the event that problems are experienced when exporti...

Page 108: ...feLock Cart main column 4 Disconnect both ends of the 5V power outlet one end from the SafeLock Cart main column power and the other from the Samsung DVD R RW drive rear panel 5 Release the locking ar...

Page 109: ...to the USB ports located below the upper shelf of the SafeLock Cart see Figure 3 35 and Figure 3 36 Figure 3 34 Cable for Installation on Vivid i n Vivid q N Cart Figure 3 35 USB Cable Plugged Into th...

Page 110: ...the USB port located at the rear of the Vivid i n Vivid q N system Figure 3 31 on page 3 39 6 Re connect power to the SafeLock Cart and to the Vivid i n Vivid q N system 7 Switch on the circuit breake...

Page 111: ...by GE Medical Systems WARNING Do not attempt to use a different type of mass storage device other than the MOD Drive supplied with the system The ultrasound system is an extremely sensitive and comple...

Page 112: ...asound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage Refer to the External USB Hard Disk Installation Procedure section pa...

Page 113: ...ality No SafeLock Cart connection via USB hub As above If applicable the USB hub must be powered via the isolation transformer Vivid i n Vivid q N mounted on SafeLock Cart As above 3 7 2 7 Footswitch...

Page 114: ...are Installation Vivid i n Vivid q N mounted on SafeLock Cart As above NOTE After connecting the Footswitch to the system define the required functionality settings as follows 1 Press Config to access...

Page 115: ...IVID Q N SERVICE MANUAL Chapter 3 System Setup 3 47 3 Using the Footswitch Functionality drop down menus select the required functionality for the Left Middle and or Right Footswitch see the example i...

Page 116: ...cated on the Vivid i n Vivid q N rear connector panel see note below Note When performing step 1 it may be necessary to use the VGA adaptor P N 2420728 see Figure 3 44 to connect the end of the VGA ca...

Page 117: ...correct positions as shown in Figure 3 47 6 When all power and signal cables are connected to the system VCR and video converter proceed as follows in the order stated a Press the power ON switch on...

Page 118: ...Erase simultaneously The following screen will appear Figure 3 48 Intel Dual Display Clone 9 Select the Intel Dual Display Clone icon and press OK see Figure 3 48 NOTE If the VGA cable is not properl...

Page 119: ...e on the cart It is possible to use a different type of VCR brand and model However it is highly recommended to use a medical graded VCR as this will ensure that both safety and image quality levels m...

Page 120: ...N 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 52 Section 3 7 Completing the Hardware Installation 3 7 2 9 Connecting the VGA Splitter Figure 3 51 VGA Splitter Figure 3 52 Cables and Bracket for Wall Mo...

Page 121: ...d Service Engineer must be present when connecting the VGA cable top right in Figure 3 52 between the Vivid i n Vivid q N system and the GE Video Splitter Isolation box 1 Connect the VGA cable Figure...

Page 122: ...for further connection to other DVI devices Refer to Figure 3 54 above 4 When connecting the DVI output cable to an external monitor press V Out on the Vivid i n Vivid q N system console The screen s...

Page 123: ...ollowing safety precautions Do not expose the charger or power supply to water or conductive liquids this is not a sealed case Do not open the charger or power supply case There are no user serviceabl...

Page 124: ...e faster if the battery is fully discharged to begin with The recalibration time is approximately 9 hours for an empty battery and approximately 14 hours for a fully charged battery Note Calibration i...

Page 125: ...the battery charger unit No action is required if only a recharge is required as the charger will automatically begin to charge the battery The blue calibration LED will flash to indicate that the ba...

Page 126: ...side of the unit Probes can be connected or changed any time as described below regardless of whether the system is powered ON or OFF CAUTION Handle the probe gently while connecting and disconnecting...

Page 127: ...cord so that it is free to move and is not resting on the floor When the probe is connected it is automatically activated CAUTION Do not bend the probe cable acutely Fault conditions can result in el...

Page 128: ...TION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 60 Section 3 7 Completing the Hardware Installation 3 7 4 1 Available Probes For a list of the available probes please see Section 9 7...

Page 129: ...xternal ECG is to be used See also information on Ext ECG Module Connector provided in ECG Module on page 5 39 Note For optimal ECG operation use only electrodes that meet universal standards see Tabl...

Page 130: ...er plug should not under any circumstances be altered to a configuration rated less than that specified for the current DO NOT use an extension cord or adaptor plug Refer to 2 2 5 Electrical Requireme...

Page 131: ...wer cable connector from the AC input socket on the SafeLock Cart or the AC DC converter of the system 3 Unplug the mains power cable from the AC wall outlet socket 4 Unplug the AC DC power connector...

Page 132: ...stem was shut down If the probe has been removed since the system was last used the currently connected probes and their available applications are displayed and selected by default 3 7 7 2 Switching...

Page 133: ...the sliding guides on either side of the shelf as shown below NOTICE Only the SafeLock Cart supplied by GE Ultrasound and or any medical grade cart which has been approved as suitable for use with th...

Page 134: ...of one finger between the SafeLock Cart arm rest and the front of the scanner 3 If necessary perform the alignment procedure as described in 3 8 2 Adjusting the SafeLock Cart Locking Mechanism on page...

Page 135: ...i n and Vivid q N on the SafeLock Cart cont d 6 Connect the USB Cable from the cart to the USB port on the Vivid i n and Vivid q N rear panel Figure 3 64 Vivid i n and Vivid q N Rear Panel on page 3...

Page 136: ...t and the device see Figure 3 67 on page 3 68 2 Working from the underneath the main shelf use a Phillips screwdriver to loosen the two screws half turn counter clockwise holding the middle locking to...

Page 137: ...ant sub sections Table 3 6 Configuration Procedures SUB SECTION DESCRIPTION PAGE NUMBER 3 9 2 Vivid i n Vivid q N Configuration 3 69 3 9 2 2 Enter Location 3 70 3 9 2 3 Date and Time Adjustments 3 71...

Page 138: ...n Window see Preparations If needed select SYSTEM The System Settings Window is displayed 2 Select the Hospital field see a in Figure 3 68 and type the name of the hospital max 64 characters The 24 fi...

Page 139: ...New time is displayed 4 Select the preferred Time Format see c in Figure 3 69 24 the 24 hour format is used 12 the 12 AM PM hour format is used 5 Select the preferred Date Format see d in Figure 3 69...

Page 140: ...page 4 10 Select SYSTEM if needed The System Settings Window is displayed 2 Use the Language drop down dialog see Figure 3 70 to select your preferred language for the on screen interface The followi...

Page 141: ...gional Settings The Regional and Language Options dialog box opens 4 Under Standards and formats in the language field scroll down to the desired language e g Russian and click OK Reboot the system Th...

Page 142: ...XPECTED RESULT S 1 Open the Configuration Window see Log On to the System as ADM on page 4 10 Select SYSTEM if needed The System Settings Window is displayed 2 Use the drop down Units dialog see Figur...

Page 143: ...ntrast and Backlight Intensity Keyboard Setups Printer Setup Disable TCPIP filtering NOTE The Disable TCPIP Filtering option should only be used for Troubleshooting procedures after which it it is rec...

Page 144: ...gs as required For instructions see LCD Display Adjustments and Calibration on page 6 2 3 9 3 5 Keyboard Setup 1 Press F2 select Service Tab The Select Keyboard Setup dialog is displayed 2 Click Keybo...

Page 145: ...RE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 77 3 9 3 5 1 Regional Options Select appropriate Format and Location then click Apply Figure 3 75 Regio...

Page 146: ...IVID Q N SERVICE MANUAL 3 78 Section 3 9 Configuration 3 9 3 5 2 Languages 1 Under the Languages tab click Details to select appropriate Language 2 If necessary activate applicable Supplemental Langua...

Page 147: ...3 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 79 3 9 3 5 3 Advanced 1 Under the Advanced tab select a language version of the non unicode programs you want to use then c...

Page 148: ...s where there is a need follow the instructions provided by GE Medical Systems or the instructions in the respective printer installation manual 1 Select ADD PRINTER to start the Add Printer Installat...

Page 149: ...r Vivid q N system refer to the External USB Hard Disk Installation Procedure on page 8 151 DVD CD R Drive via USB Printer Deskjet HP 450 via USB Virtual printer any of the DeskJet printers can be use...

Page 150: ...e spooled print jobs will be printed out automatically Sometimes this feature is device dependent Therefore in order to initiate the manual print of all spooled jobs process proceed as described below...

Page 151: ...und scanner is mounted on the SafeLock Cart always use the lower USB port for the Hub 5 The Vivid i n and Vivid q N ultrasound scanner and all connected devices working with scanner should be powered...

Page 152: ...age 4 10 2 Select ADMIN lower part of window 3 Select the SYSTEM ADMIN tab 4 Select NEW to open the New Key dialog where you type the Software Option Key Serial String 5 Type the Password Software Opt...

Page 153: ...way A Vivid i n Vivid q N scanner configured for eVue Images or loops may be transmitted to a designated PC e Vue PC on the network in compact MPEG format and will appear at the review PC in almost re...

Page 154: ...HD dataflow The local database is used for patient archiving Images are stored to internal hard drive LocalArchive MOD dataflow The local database is used for patient archiving Images are stored to a...

Page 155: ...m Vivid i n Vivid q N to an EchoPAC PC Workstation In this configuration the data is transferred from the Vivid i n Vivid q N to a dedicated EchoPAC PC Workstation over an Ethernet connection The conn...

Page 156: ...onnect the Vivid i n Vivid q N system to an EchoPAC PC and a network printer Figure 3 86 Ethernet Switch Identifications Figure 3 1 Position Label Function X To PC To connect an end station or a Serve...

Page 157: ...AC PC Workstation Figure 3 88 Hospital Network Connections See the EchoPAC PC Service Manual Part Number EP091298 for details about the EchoPAC PC and the required set up procedures EchoPAC PC Vivid7...

Page 158: ...the GE label in the DVD drive 3 Boot up the system in maintenance mode and navigate to Windows Explorer 4 Create a new folder on the system s D drive and name it WIFI DRIVER 5 Copy the contents of th...

Page 159: ...stall from a list or specific location Advanced see Figure 3 90 and click Next 10 Select Include this location in search Figure 3 91 and browse to D WIFI DRIVER Figure 3 92 11 Click Next The following...

Page 160: ...vity Setup 12 Select the WIFI DRIVER folder then click OK The following screen opens 13 Click Next The installation process continues displaying wizard screens until installation is complete 14 Click...

Page 161: ...q N scanner screen press Config Select the Connectivity tab and from the Connectivity screen click TCP IP 2 Click Advanced Settings to open the Network Connections screen see Figure 3 94 3 From the W...

Page 162: ...ireless network and click Connect In this example Figure 3 95 the wireless network connection is Vivid_S Testing The Wireless Network Connection dialog box opens see Figure 3 96 5 Type the required ne...

Page 163: ...creen indicating that the wireless adaptor is connected see Figure 3 98 6 Return to the instructions for Installing the Compact Wireless G Adaptor Software on page 3 90 and repeat the entire procedure...

Page 164: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 96 Section 3 10 Connectivity Setup Figure 3 99 CISCO_350 Sub directory...

Page 165: ...to the network connector to the EchoPAC PC Workstation see Chapter 3 in the EchoPAC PC Workstation Service Manual Connection via a Peer to Peer Network You will need a network hub and one network cab...

Page 166: ...ge 2 10 Typically the source for this information is the network administrator Follow the steps below to prepare the scanner for use on the network 3 10 3 2 Select TCP IP Set up Screen 1 Press CONFIG...

Page 167: ...IP address 0 0 0 0 4 In addition the scanner s AE Title must be entered in the DICOM server s setup 3 10 3 4 Set the Remote Archive s Network Information In the Remote Archive Setup area of the screen...

Page 168: ...mote Path used for Save As Export from Q Analysis and for exporting error logs with Alt D 1 From the system keyboard press Config and log on as an ADM user with the appropriate password 2 Click Connec...

Page 169: ...log on the default user and password Under Configurable Remote Path User refer to Figure 3 103 type the desired user name and password as configured on the destination PC This is the password usually...

Page 170: ...Vivid i n Vivid q N control panel press the Config button 2 From the System Configuration window click the Connectivity tab at the bottom of the window The Connectivity dialog box opens 3 Select the...

Page 171: ...mputer 12 Type in the Subnet mask 255 255 255 0 Note The Subnet mask should be the same for both the network computer and the Vivid i n Vivid q N 13 Click OK 14 Press the Config button on the control...

Page 172: ...q N system 8 Type in the Subnet mask 255 255 255 0 Note The Subnet mask should be the same for both the network computer and the Vivid i n Vivid q N 9 Click OK close all windows and approve all chang...

Page 173: ...ID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 105 a Select Start Run b Type CMD then press Enter c Type Ping and IP 10 0 0 1 for Vivid i n or Vivid q N then press Enter If return timing...

Page 174: ...s the Set button on the control panel 1 On the Vivid i n Vivid q N control panel press the Config button 2 From the System Configuration window click the Connectivity tab at the bottom of the window T...

Page 175: ...vid q N IP address 10 0 0 1 or the one provided by the local network administrator Note The IP must be different from the IP address used for network computer 12 Type in the Subnet mask 255 255 255 0...

Page 176: ...he network computer access the Control Panel as follows a Double click on My Computer icon on desktop b Double click on Control Panel icon 2 Double click on Network Connections icon 3 Double click on...

Page 177: ...8 Connect the Vivid i n Vivid q N and network computer LAN connection using a regular Network cable The Vivid i n Vivid q N and the network computer are now ready for export import of files Note One c...

Page 178: ...work against the DICOM server the following information has to be entered in the scanner DICOM server IP address subnet mask and eventually the gateway DICOM server port number DICOM server AE title t...

Page 179: ...VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 111 The selected flow is shown as seen in the example in Figure 3 116 below Worklist Local Archive DICOM Server Int HD Figure 3 116 Work...

Page 180: ...ct Properties to display the Properties dialog 7 Click on Search Criterias Figure 3 117 The Search Criterias window opens Figure 3 117 8 Highlight the modality filter in the list and click on Remove s...

Page 181: ...dd to List The Search Criterias window is updated see Figure 3 118 11 Click OK to close the window 12 From the Worklist Properties Dialog select the IP Address down arrow to choose the Worklist Server...

Page 182: ...below to change the IP Address settings a From the IP Address pull down menu select Modify to display the IPs dialog box b From the IPs dialog box select the server you want to modify then select Mod...

Page 183: ...dialog to return to the Properties dialog box f Enter the DICOM server AE Title This entry is case sensitive and must match exactly g Enter the DICOM server port h For some DICOM Servers the default T...

Page 184: ...the connection to the device verification process The verification process may require up to one minutes When done a sign prefix displayed in front of the listed device indicates whether the test pass...

Page 185: ...tation and Vivid i n Vivid q N being configured to the Mitra support personnel 3 10 5 3 Configuration of Modality Worklist Services Identify Dataflow to be Used 1 Identify which dataflow is to be used...

Page 186: ...it number for the Value 3 Click the Add to List button 4 Click the OK button Assign DICOM Worklist name to Dataflow 1 Select the Dataflow tab 2 From the Dataflow name pull down menu select the Dataflo...

Page 187: ...POLL_DIR b Verify that Text is select for the Export file format Test Connectivity with the Vivid HL7 Gateway 1 Start a new study and acquire an image or use an existing study 2 In the Measurements mo...

Page 188: ...rver for further analysis on the Vivid i n Vivid q N 3 10 6 2 Query Retrieve Setup on the Vivid i n Vivid q N 1 Press CONFIG F2 and log on as Adm as described in Log On to the System as ADM on page 4...

Page 189: ...System Setup 3 121 5 Select QueryRetrieve so it is highlighted and then select Properties to display the Properties dialog box Figure 3 128 Query Retrieve Workflow Figure 3 129 Select Query Retrieve s...

Page 190: ...e pull down menu and edit the setup for one of the predefined servers 7 Enter the correct AE Title and Port Number for the DICOM Query Retrieve server in the respective fields in the Query Retrieve sc...

Page 191: ...ANUAL Chapter 3 System Setup 3 123 2 Select the correct tag from the Select Tag pull down menu 3 If needed type in the value 4 Select Add to List 5 Click OK to close the Search Criterias window Figure...

Page 192: ...onnection 1 Select the Smiley button to Ping the server 2 If the network connection to the server is OK it will be illustrated by a smiling Smiley A sad Smiley indicates that the network connection is...

Page 193: ...h Ping and DICOM Ping 1 Select QueryRetrieve from the Selected devices list 2 Select Check If the test passes a pop up dialog is displayed see left illustration in Figure 3 134 Select OK to continue A...

Page 194: ...SERVICE MANUAL 3 126 Section 3 10 Connectivity Setup If the test fails a pop up warning is displayed see left illustration in Figure 3 135 Select OK to continue A red X to the left of the Selected de...

Page 195: ...ML MPEG and JPG formats for all of the images in the exam It also contains measurements and reports generated and saved during the exam the reports are all saved as files in CHM format Any media gener...

Page 196: ...GVue Export 1 Press the Config or F2 button and log on as Adm as described in Log On to the System as ADM on page 4 10 2 Select Connectivity 3 Select Dataflow 4 From the drop down menu select Misc Exp...

Page 197: ...mputer name or the remote computer s ID address in the Remote Path field After the path has been typed it is available for selection from the Dataflow Name drop down menu or you can select an existing...

Page 198: ...l open and select Patient List The Patient List will be displayed with the following header 4 Trackball to the Export softkey in the bottom menu shown below and press SET The Export dialog box opens 5...

Page 199: ...displayed 9 If required click OK then select any additional Patient Files for export 10 When export complete click OK then remove the media containing the exported MPEGVue data from the Vivid i n Viv...

Page 200: ...here consultation with another doctor in the hospital in real time is required 3 10 8 2 Preparations Before configuring the Vivid i n Vivid q N for eVue dataflow it is necessary to decide upon a speci...

Page 201: ...drop down menu 7 From drop down menu select the Dataflow named Local Archive Int HD eVue 8 Trackball to the Selected devices window and select highlight eVue see Figure 3 140 9 Trackball to the Prope...

Page 202: ...transfer of eVue files This step is not needed for Windows XP After initial transfer of eVue it is recommended to disable this check box The Vivid i n Vivid q N dataflow is set and ready for export im...

Page 203: ...select Local Archive Int HD eVue as shown in Figure 3 142 4 Using the soft key buttons select CREATE PATIENT 5 Now start scanning Each image you would like to store press the STORE button A few second...

Page 204: ...igure 3 143 and Figure 3 144 requesting the user to download and install some drivers The requested drivers may be downloaded from the Internet or if the DVD CD was pre configured those drivers will b...

Page 205: ...0 9 1 1 Manually Setting the Network PC for eVue Import 1 On the network computer create a New User by accessing the Control Panel as follows a Double click on My Computer icon on desktop b Double cli...

Page 206: ...Connectivity Setup The User Accounts dialog box opens 3 Select the Advanced Tab 4 Click on the Advanced button The Local Users and Groups dialog box opens 5 Select the Users folder and right click to...

Page 207: ...Password never expires check boxes e Click Create to confirm setting and Close f Close all windows The new user account has been set and is ready for activation Create a new folder for import export a...

Page 208: ...on and select Rename Rename the folder as eVue _Import 9 Setting the Sharing properties for the new folder as follows a Right click on eVue_Import icon and select Properties The eVue_Import Properties...

Page 209: ...ing Tab c Click on the Permissions button The Permissions for eVue Import dialog box opens d Click the Add button Note It is recommended to remove the User Everyone from Share Permissions Figure 3 148...

Page 210: ...log box select all to be active activate all the check boxes as shown below and press OK g From the Select Users Computers or Groups dialog box Figure 3 150 click on the Locations button h In the Loca...

Page 211: ...etup 3 143 j Click the Check Names button The name and path will be detected by underlining as shown below k Now click OK l From the Permission for eVue_Import dialog box highlight and select E1c2h3o4...

Page 212: ...e network computer using the browsing option to select the appropriate path For an example of Vivid i n Vivid q N dataflow settings refer to Figure 3 140 on page 3 133 3 10 9 1 2 Manually Stopping the...

Page 213: ...ID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 145 c Select Control Panel from the Menu 2 Double click on Network Connections as shown above 3 Select your network connection and right cl...

Page 214: ...Uncheck the Internet Connection Firewall check box as shown in Figure 3 153 7 Click the OK button to confirm changes and close all windows Stopping an External Service Firewall 1 On the network comput...

Page 215: ...vices dialog box select your firewall program for example Black Ice 5 Right click and select Stop from drop down menu The following message will be displayed indicating that the Firewall is being stop...

Page 216: ...e network computer access the Control Panel as follows a Press Start b Select Settings c Select Control Panel from the Menu 2 Double click on Windows Firewall The Windows Firewall dialog box opens 3 S...

Page 217: ...may need to obtain these from the local IT person 2 Close all Internet Explorer windows if open 3 From the C Program Files GEULS eStorePlayer Version1_0 folder double click on the UnRegisterAll bat ic...

Page 218: ...AND VIVID Q N SERVICE MANUAL 3 150 Section 3 10 Connectivity Setup 5 From the C Program Files GEULS eStorePlayer Version1_0 folder delete all files as shown in Figure 3 160 on page 3 151 Figure 3 158...

Page 219: ...DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 151 6 Restart Windows then confirm that the MPEGVue application is no longer installed Figure 3 160 Delet...

Page 220: ...lay To read an MPEGVue exam stored on a DVD CD proceed as follows Insert the MPEGVue CD in the computer DVD CD drive The MPEGVue Patient List is displayed see Figure 3 161 below Note In the event that...

Page 221: ...played in full size in the MPEGVue Screen view pane There are several ways to select a thumbnail image from the clipboard Use the mouse pointer With the pointer on the required image click the left mo...

Page 222: ...As function an image may be saved to any of the following file formats BMP WMV MPEG4 JPEG To save an image click on the diskette icon 8 in Figure 3 162 A Save As dialog box will appear allowing you to...

Page 223: ...ssage is displayed prompting the user to first logon to Outlook It is possible to send images loops or full exams from any eVue PC An eVue PC is a PC which contains the MPEGvue player and an e mail cl...

Page 224: ...4 Select an existing recipient address from the drop down menu or type a new address in the Send To field Note Up to 10 addresses can be stored for future selection 5 Click Next The user is prompted t...

Page 225: ...il message has been sent from the Outlook Outbox Notes Examinations sent by E mail as attachments are compressed encrypted and password protected If the size of the exam is too large to be sent as a s...

Page 226: ...to send the MPEGVue Player software to the recipient first complete this procedure then proceed to Sending the MPEGVue Player Software by E mail on page 3 159 The E mail address window is displayed 3...

Page 227: ...6 Check that the message has been sent from the Outlook Outbox When the message has been sent from the Outbox and received by the E mail recipient the image can now be opened in Windows Media Player S...

Page 228: ...as been sent from the Outbox and has arrived at the addressee the recipient should follow the instructions included in the body of the email in order to install the MPEGVue Player First Time Receipt o...

Page 229: ...ed on a remote computer as described in First Time Receipt of MPEGVue Player Software and Images by E mail on page 3 160 the user is able to manage and review images or exams received via email using...

Page 230: ...From the MPEGVue Startup Menu click on the Get new exams by email button The Start MPEGVue screen is displayed as shown in Figure 3 175 below 2 Click the Outlook E mail button shown in Figure 3 175 Th...

Page 231: ...ail Inbox is being scanned for studies When complete all studies found in the Inbox are listed in the Found Studies dialog box as shown in Figure 3 176 below 4 Select the required Study from the list...

Page 232: ...ls Note The Password is case sensitive In the event that an incorrect password is typed the user is prompted with a warning message as shown here Click OK and enter the correct password 7 Click Next T...

Page 233: ...ail examination If you were previously viewing different MPEGvue Patient exams you will need to switch over to the previous instance of the MPEGvue Receiving and Viewing a Partial Examination In the e...

Page 234: ...n the example above to select and display images of that exam The MPEGVue screen is displayed as shown in the example in Figure 3 162 on page 3 153 Activate e Vue 1 From the MPEGVue Startup Menu click...

Page 235: ...rom Email tab is activated 2 Manage Exams from Email as follows a Make sure the Manage Exams from Email tab is activated as shown in Figure 3 181 The Patient List displays details in table format of a...

Page 236: ...Only when both these tasks have been performed will these two buttons become enabled c To copy or move the selected exams first click the Browse button and select the required destination folder then...

Page 237: ...The Patient List displays details in table format of all patient exams received via export from E Vue Patient ID Patient Name Exam Date etc The current path information is displayed below the Patient...

Page 238: ...bove The current path information is displayed below the Patient List together with details of the removable media e g CD DVD ROM By default the destination path Other Folder radio button is enabled a...

Page 239: ...L Chapter 3 System Setup 3 171 e Click the Copy Selected Patients button During the copy operation the following message and progress bar is displayed When done the currently selected exam s are avail...

Page 240: ...eadings Item Value Minimum Requirements Below this information the eVue Operation Utility buttons enable the user to enable or disable the eVue mode as required An additional button Configure eVue is...

Page 241: ...ired eVue folder destination drive If applicable enable the peer to peer connection check box then type the correct current IP address and Subnet mask details Click Save to save the new configuration...

Page 242: ...id system configuration inconsistencies before using the Enable eVue or Disable eVue functions make sure that your PC is connected to the appropriate network and that all network cables are properly c...

Page 243: ...Note On most computers when inserting the DVD CD media with MpegVue files the MpegVue Player will start automatically due to the autoplay option of DVD CD media In the event that any updates are requi...

Page 244: ...The MpegVue Player can be used as a regular player using the Play Stop Forward and Rewind buttons In addition the user can forward images by e mail save images and perform measurements as previously...

Page 245: ...imation Time section on page 3 2 3 11 1 Safety Precautions for Moving the Vivid i n Vivid q N Unit CAUTION Whenever moving the Vivid i n Vivid q N ultrasound unit while mounted on the SafeLock Cart al...

Page 246: ...rt of the original system documentation This will ensure that all relevant safety and user information is available during the operation and service of the complete system 1 Verify that all details ha...

Page 247: ...rd NOTE The Product Locator Installation Card shown may not be the same as the provided Product Locator card From the factory a sheet with four Product Locator cards for transportation and one for Ins...

Page 248: ...the installation Specific language versions of the Vivid i n Vivid q N User Manual may also be available Check with your GE Sales Representative for availability During and after installation suppleme...

Page 249: ...t be performed either as a part of the installation procedure or as required during servicing and periodic maintenance of the Vivid i n Vivid q N ultrasound scanner Note Refer to Chapter 7 Diagnostics...

Page 250: ...nt Required An empty blank MO Disk P N 066E0490 1 3 GB P N 066E0519 5 4 GB P N 066E0514 8 6 GB ECG Pads ECG Cable IEC EUROPE ASIA ROW P N 2378938 2 At least one probe ideally you should check all the...

Page 251: ...urrent DO NOT use an extension cord or adaptor plug Refer to 2 2 5 Electrical Requirements on page 2 3 for more details DANGER ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE PROTECTIVE...

Page 252: ...ng sufficient slack so that the unit can be moved slightly 3 Verify that the power cable is without any visible scratches or any sign of damage 4 Verify that the on site mains voltage is within the li...

Page 253: ...ing a normal boot you may observe that a The unit s ventilation fans starts on full speed but slow down after a few seconds listen to the fan noise b Power is distributed to the peripherals Operator P...

Page 254: ...initiates and sets up the Front End electronics and the rest of the scanner 4 The Keyboard backlight is lit 5 As soon as the software has been loaded a 2D screen is displayed on the monitor as shown b...

Page 255: ...This is not recommended under normal circumstances Closing the lid will place the system in Standby Mode Note The system will remain in Standby Mode for as long as battery power will allow The System...

Page 256: ...em is unplugged for a longer period of time a full shutdown is automatically performed As described above the battery is automatically charged when the system is plugged to the wall outlet and the pow...

Page 257: ...in the LCD display Then system shut down commences The time required to power down the unit including the Back End Processor may vary from 10 seconds up to approximately 1 minute The last component th...

Page 258: ...dm usr It is possible for the administrator adm to establish unique passwords for each user but don t do it now 3 Select the user adm From the factory the password has been left blank so you may try t...

Page 259: ...E MANUAL Chapter 4 General Procedures and Functional Checks 4 11 5 Select System to access the System Setup window The System Settings window is displayed 6 Proceed to select the required settings Tab...

Page 260: ...ther side to the rear of the MOD drive When connecting the MOD drive directly to the system it is necessary to use an additional power source As a safety precaution this must be done via an isolation...

Page 261: ...e it in its cover box for storage in a safe place 4 2 5 4 Connecting External USB Hard Disk to USB Port The External USB Hard Disk can be connected to either of the USB ports upper or lower on the rea...

Page 262: ...nly high grade media types NOTE For BT06 systems it is not possible to use DVD R media only CD R or DVD R 1 Press the Eject button on the right side of the DVD CD Drive once to open the disk tray 2 Pu...

Page 263: ...r the label name Remember that MO Disks have two sides Side A and Side B Each Side must have a unique name Example MO Disks for storing images may be labeled IMAGE_1A Side A on MO Disk 1 IMAGE_1B Side...

Page 264: ...16 Section 4 2 General Procedures 4 2 8 Verifying Removable Media 1 Insert the removable media to be verified 2 Select the pre configured removable media from the Config Connectivity Tools pop up men...

Page 265: ...OTICE Presets should not be saved on the same MO disk as images because the Archive Menu lists the images but not the presets stored on a MO disk Table 4 3 Saving Presets STEP TASK FUNCTION 1 Insert a...

Page 266: ...resets STEP TASK FUNCTION 1 Insert the MO disk with the archived presets into the MOD 2 Press Config or F2 select Admin then select Restore The Restore window is displayed 3 Enable the User Defined Co...

Page 267: ...e the handle from the maximum lowest position to the maximum upper position Verify that the handle is able to reach the maximum lowest and maximum highest positions smoothly 5 Pick up the system using...

Page 268: ...tion The position value is 31 2 for all TGC Sliders 7 Perform the ROTARY TEST as follows Rotate the rotary in both directions and press the button at the center of the rotary The rotary is activated p...

Page 269: ...heck the Battery green check box Check the Battery yellow check box Press Set Battery yellow LED is On c Uncheck the Battery yellow check box Check the Extern DC green check box Press Set Extern DC gr...

Page 270: ...e good control and that the x and y coordinate values displayed at the bottom of the screen move digit by digit up and down Movement is smooth no jumping and the trackball has no dead areas i e when t...

Page 271: ...hind the unit 6 At the WAV window press the Right Channel button Make sure that the right internal speaker is working Verify that there is clear sound from both speakers at all frequencies Note that r...

Page 272: ...are no missing pixels in the Full Screen Green or display in another color 8 Select Blue Make sure there are no missing pixels in the Full Screen Blue or display in another color 9 When done press Clo...

Page 273: ...aintenance Mode using the Service Dongle and load the application by double clicking the Echoloader icon on the desktop 2 Connect the ECG cable to the input on the ECG port on the right side of the Vi...

Page 274: ...ory signal is displayed in the Data area as shown in the example below 9 Verify that a single clear Respiratory signal is displayed 10 When done trackball to the Exit button and press Set 11 When prom...

Page 275: ...and press Start PASS 8 Select MUX Test then select 3S RS Test and press Start PASS 9 Select TR32 Test then select RX Test Select RSTest with 3S RS and press Start PASS 10 When done press the Esc butt...

Page 276: ...all the procedure instructions described in Connecting the VCR on page 3 48 2 Connect an external monitor to the VCR S Video or Video output 3 Connect a 3S RS probe Reset the VCR counter and press th...

Page 277: ...zen on the screen 3 Press the Print button Evaluate the print quality Adjust the brightness and or contrast of the B W printer if required The image displayed on the screen is printed on the B W or co...

Page 278: ...r is installed in a system that is already configured proceed as described below 4 3 2 4 1 Wireless USB Network Functionality Test 1 With the system up and running press Config 2 Click Connectivity an...

Page 279: ...has been successfully performed as indicated by the green confirmation tick see Figure 4 9 7 Under Selected Devices Figure 4 9 proceed to the next device in this example Remote Database that is under...

Page 280: ...e on the Vivid i n or Vivid q N scanner Save several images in the patient file 3 Export the images to the External USB Hard Disk 4 Delete the dummy patient file including images from the Vivid i n Vi...

Page 281: ...t the shelf movement is smooth 3 Press the release handle down while pushing the main shelf back downwards to return it to its lowest position The main shelf returns to its lowest position Table 4 21...

Page 282: ...matted successfully is displayed 8 Select the GetVolInfo button The message Media in drive G has following properties Media Label MY_LABEL Media File System Name FAT is displayed 9 Select the Get BTPs...

Page 283: ...t Patient and press Export The Export pop up is displayed 8 Place a blank DVD media into the DVD CD device Note For information on types of media see DVD CD RW Drive on page 3 36 9 Select To CD DVD Ar...

Page 284: ...to confirm delete must confirm three times The Test Patient is deleted from patient list Table 4 25 Plug and Scan Test Recommended Task Expected Result s 1 Press the On Off button on the control conso...

Page 285: ...Date and Time Adjustments on page 3 71 2 Check the language settings Refer to Language Selection on page 3 72 3 Ensure that all of the software options are configured correctly Refer to Software Opti...

Page 286: ...HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 4 38 Section 4 5 Site Log Section 4 5 Site Log Table 4 1 Vivid i n Vivid q N Site Log Date Service Person Problem Commen...

Page 287: ...RE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 4 General Procedures and Functional Checks 4 39 Table 4 1 Vivid i n Vivid q N Site Log cont d Date Service Person Proble...

Page 288: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 4 40 Section 4 5 Site Log This page was intentionally left blank...

Page 289: ...ystem functions for both the Vivid i n and Vivid q N systems It also describes the power distribution system and probes Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1...

Page 290: ...Gray Scale imaging Color M Mode Doppler Different combinations of the above modes 5 2 1 Introduction The Vivid i n and Vivid q N systems are equipped with a highly efficient power supply resulting in...

Page 291: ...or Panel to the Front End FE Electronics to the Back End Processor BEP and finally are displayed on the monitor and peripherals In addition the Vivid i n and Vivid q N systems enable acquisition of th...

Page 292: ...ator Panel that has ultrasound specific buttons mode dependent softkey buttons and an alphanumeric keyboard 5 2 2 3 Electronics Vivid i n and Vivid q N internal electronics are divided into three main...

Page 293: ...5 6LJQDOV 56 3UREH 3RZHU 0RGXOH 7 9 6 9 YROWDJHV 9 9 9 9 9ROWDJHV 9 67 9 67 9 9 DWWHU 30 WHUQDO 0RQLWRU 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7...

Page 294: ...into two subsystems 1 Scanner Subsystem Consists of the RFI Board Scan Controller and tissue processing For a detailed description see RFI Board on page 5 11 2 Front End Subsystem Includes the followi...

Page 295: ...EH 3RZHU 0RGXOH 7 9 6 9 YROWDJHV 9 9 9 9 9ROWDJHV 9 67 9 67 9 9 DWWHU 30 WHUQDO 0RQLWRU 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7 6 5 0 0 17 3 02...

Page 296: ...IHU 3XOVHU 3 86 6 9 08 3UREH 08 7 5 6LJQDOV 56 3UREH 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7 6 5 0 0 17 3 02 3 4 5 N 86 5 DWD Q 0 86 7 7 67 6 LD...

Page 297: ...the Bottom assembly on the right side as shown in Figure 5 5 Figure 5 5 Bottom Assembly showing Location of Front End WARNING When opening and closing the Keyboard Assembly take care not to damage th...

Page 298: ...091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 5 10 Section 5 3 Front End Unit Figure 5 7 Bottom Assembly showing Position of Boards RFI TR32 P M Probe and MUX Board TR32 Boards located belo...

Page 299: ...Signal Generator Tx voltage Enable Disable Probe Sensing Control of TR32 and P M Boards Ultrasound data acquisition and processing Figure 5 8 RFI Block Diagram 5 3 3 2 Watchdog and Power ON Reset For...

Page 300: ...A Download proceeds as follows 1 The actual FPGA contents bit streams are written into DSP memory from the Host 2 The DSP configures serial port 0 to aid configuration 3 The DSP writes the bit streams...

Page 301: ...ction Theory 5 13 5 3 3 6 RFI Board Connectors Figure 5 9 RFI Board Connector Locations Table 5 2 RFI Board Connectors Connector Description P5 To P M P6 PCI Flex to RFI P14 To power supply J1 To TR32...

Page 302: ...eading falls below minimum temperature the RFI stops scanning and disables Tx voltage The minimum temperature references are set to 15 o C A temperature reading as low as 15 o C indicates a probe temp...

Page 303: ...Boards TR32A and TR32B see Figure 5 10 Each TR32 Board is a complete 32 channel ultrasound Front End that includes the following features and functions 32 Transmitters 32 receiving channels T R switc...

Page 304: ...es Input Description Connector PIN Connected From 5 VA DC voltage distributed via RFI Board A indicates analog circuit J2 80 81 82 RFI 1 8 V DC voltage distributed via RFI Board J2 85 96 RFI 3 3 V DC...

Page 305: ...wo TR32 Boards reside in the RFI Board The two TR32 Boards are linked together in parallel The way in which they operate depends upon the probe currently in use and also on the current type of Transmi...

Page 306: ...he P M enables up to 64 receive transmit channels and performs the following functions Receiving the channels and connecting them to the TR32 Boards Probe connectors High voltage MUX that selects 64 o...

Page 307: ...ts and Function Theory 5 19 Figure 5 15 Probe and MUX Board Connector Locations Table 5 6 Probe and MUX Board Connectors Connector Description J1 Pencil Probe Connector J2 RS Probe Connector J3 To Upp...

Page 308: ...12 V DC voltage distributed in P M Board J5 16 21 RFI Board 3 3 V DC voltage distributed in P M Board J5 5 8 RFI Board 5 VA DC voltage distributed in P M Board A indicates analog circuit J5 13 15 RFI...

Page 309: ...the main controller for the system A block diagram of the BEP is provided in Figure 5 17 on page 5 22 The various components that comprise the Vivid i n Vivid q N BEP are described in the following se...

Page 310: ...TION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 5 22 Section 5 4 Back End Processor Figure 5 17 Vivid i n Vivid q N Back End Processor Block Diagram DFN QG 7 6 DVHERDUG 3 0RQLWRU 0RQLW...

Page 311: ...Base Board provides interfaces to the following components HDD Front End RFI Board USB ports Docking port speaker phones and speaker jacks Keyboard Battery VGA Fans Power supply The Power Management...

Page 312: ...the ETX Base Board Figure 5 18 ETX Base Board Block Diagram 5 4 3 2 VGA Interface The interface to the VGA is via the VGA connector on the rear side of the ETX Base Board The VGA port is connected to...

Page 313: ...e space on the hard disk F Drive Swap file partition The partition size is 1GB The following additional partitions are hidden Grub 32MB Repository 10GB Note When connected to the system a DVD device o...

Page 314: ...Back End Processor 5 4 7 Back End Processor Location of Components Figure 5 20 Views of the Back End Processor BEP To Fan Board From DC DC Power Supply KBD LAN VGA PCI Cable to RFI Internal Battery D...

Page 315: ...n the BEP Table 5 9 BEP Connectors Connector Description J7 VGA J19 EXT DC INPUT J18 DOCKING J20 Speakerphones J16 RFI P1 P4a a All connections to P1 P2 P3 and P4 according to ETX specs ETX J5 LCD PAN...

Page 316: ...nnected to the Fan board see Figure 5 19 on page 5 25 Table 5 10 BEP Input DC Voltages Input Description Connector PIN Connected From EXT_DC 20 V DC from PS J17 1 2 3 4 5 PS Battery Varies from 14 V t...

Page 317: ...al monitor of VGA to S Video adaptor RJ 45 connector for LAN connection Two USB ports for external peripherals 20V DC inlet main DC power to the system Figure 5 22 Vivid i n Vivid q NSystem Rear Panel...

Page 318: ...igure 5 22 DC DC Power Supply Unit located inside and mounted to the bottom assembly provides the system with both Low Voltage and High Voltage power and also enables recharging of the battery The Pow...

Page 319: ...q N Power Supply Unit Components Block Diagram 5 6 3 1 Critical Temperature Sensor The PS Unit includes temperature sensors When the monitoring application measures the maximum temperature limit or ab...

Page 320: ...battery operation to AC operation and vice versa When shutting down the system leave the main power cable connected to keep the battery fully charged NOTE The lithium ion technology used in the system...

Page 321: ...t solder a battery Do not connect the battery to an electrical outlet CAUTION If the Vivid i n or Vivid q N portable ultrasound system is not being used for periods of one month or longer the battery...

Page 322: ...rates the layout of the keyboard and operator panel The buttons and controls are grouped together for ease of use the different areas of the keyboard and operator panel include the following Soft menu...

Page 323: ...Vivid i n Vivid q N Keyboard Block Diagram 5 7 2 Keyboard and Operator Panel Components 5 7 2 1 Power ON OFF Button The Power On Off button receives 3 3 V from the BEP Once shorted this voltage is rou...

Page 324: ...nd Menu Buttons These are equivalent to the normal PC mouse left and right click Set left click Menu right click The operation of these buttons is supported via the PS 2 Board Zoom Trackball and Store...

Page 325: ...standby for driving the power On Off and status LEDs 2 5V BE Back End is used for the remaining circuits The keyboard operates with 5 voltage deviation on each supply The keyboard has current limiting...

Page 326: ...VICE MANUAL 5 38 Section 5 7 Monitor and Operator Panel 5 7 2 12 LCD Display Unit The display monitor is a 15 LCD and is connected via a flex cable to the BEP as shown in Figure 5 28 Figure 5 28 LCD D...

Page 327: ...bipolar ECG channels leads I II The External ECG module supports acquisition of a single bipolar ECG input channel ranging from 1 to 1 volt allowing amplification of 1000 The minimal allowed impedanc...

Page 328: ...eakage and auxiliary currents apply the values listed in Table 5 12 below taken from table IV in IEC601 1 section 19 3 The ECG signals have double reinforced isolation B a from live parts Hence creepa...

Page 329: ...The Vivid i n and Vivid q N Internal ECG Respiratory module is designed to enable acquisition of the following signals ECG Respiratory and external ECG The Internal ECG Respiratory Board is located i...

Page 330: ...width 3db 1 98 31 2 Hz for Monitor use only Lead Fault Indicator Yes A2D resolution 12 bit 2 complement Sample Rate 1000 Table 5 15 Internal ECG Respiratory Board Specifications External ECG Character...

Page 331: ...ratory Monitor output type being used as follows Option 1 Monitor with single ended Positive output Connect the signal lead to pin 2 on the connector Connect the ground shielding to pins 5 and 7 on th...

Page 332: ...n Input Range ac 2v peak to peak A2D resolution 12 bit 2 complement Sample Rate 500 samples second Ground Isolation 4 kV rms Table 5 18 Connector Pin Out J5 ECG Respiration External ECG Pin Number Spe...

Page 333: ...tion Theory 5 45 Table 5 19 Connector Pins Out J3 J4 USB Pin Number Specification J3 1 J4 28 J4 30 VUSB 5V DC Input from USB Bus J3 2 J4 8 D USB differential Data signal J3 5 J3 shell D USB differenti...

Page 334: ...olor Inkjet Printer is used to print out reports and exams and is controlled in one of two ways depending on the system configuration Directly via the USB port Via LAN connection The printing operatio...

Page 335: ...ss Network Adapter The Vivid i n and Vivid q N supports a unique type of Wireless Network Adapter For details regarding type and installation refer to Wireless USB Network Adaptor on page 3 35 The Wir...

Page 336: ...s has been designed to provide the best thermal conduction allowing the maximum amount of heat to be removed from the system It is crucial that the system ventilation access is not blocked a distance...

Page 337: ...Service Adjustments Section 6 1 Overview 6 1 1 Purpose of Chapter 6 This section provides references to the procedures required for testing and adjusting the Vivid i n Vivid q N portable ultrasound s...

Page 338: ...r as stated in the following procedures NOTE Incorrect calibration of the Vivid i n Vivid q N monitor may also result in less than optimal image quality when reviewing images on a DICOM or EchoPAC sta...

Page 339: ...left numbers 1 through 6 see Figure 6 1 and the white bars on the right numbers 1 through 12 are all just visible 6 2 3 Testing Your Contrast and Brightness Settings Observe the V watermark on the low...

Page 340: ...ts and Calibration To adjust the Blue Tint hold down ALT then use the and keys as required Repeatedly pressing will decrease the blue tint repeatedly pressing will increase the blue tint To assist in...

Page 341: ...installations System Vivid i n Monitor BOE HYDIS 1 NOTE On older Vivid i q systems where the IDTECH 1 monitor is supplied the Blue Tint 2 option is not applicable NOTE The Blue Tint 2 option is incomp...

Page 342: ...s the Display Calibration Screen 2 To adjust the blue tint hold down ALT then use the and keys as required Repeatedly pressing will decrease the blue tint repeatedly pressing will increase the blue ti...

Page 343: ...ID Q N SERVICE MANUAL Chapter 6 Service Adjustments 6 7 6 2 5 2 Turning Off the Blue Tint 2 Option 1 Press Config Service The Service Screen opens see Figure 6 9 2 Under LCD Setup de select Blue Tint...

Page 344: ...Scenario B The user prefers to use the External Monitor as the Primary monitor but does not have access to it and therefore cannot make separate adjustments to it using the controls on the External M...

Page 345: ...of the following solutions 1 Purchase a standard VGA isolated splitter Do not use the GE Healthcare VGA isolated splitter P N 5390022 as this has a special pin configuration dedicated for use with Ca...

Page 346: ...t Intensity on page 3 76 6 2 8 Configuring Print Orientation For information on Print Orientation Portrait or Landscape refer to Configuring Printing Orientation Portrait or Landscape and Paper Size o...

Page 347: ...he end of the working day then allow it to continue to run automatically overnight For details see Figure 6 10 below IMPORTANT During the Battery Calibration procedure no scanning can be performed Not...

Page 348: ...the procedure to be completed overnight as outlined in Figure 6 10 1 Click on the battery icon at the bottom of the scanning screen OR from the message window shown below click Recalibrate The Batter...

Page 349: ...ce Adjustments 6 13 3 Wait for the confirmation window to be displayed showing that the battery is fully charged See Battery Calibration Wizard Page 2 in Figure 6 12 4 Click Next The Battery Calibrati...

Page 350: ...splay Adjustments and Calibration 5 Disconnect the AC power cable from the Vivid i n Vivid q N then click Next The second phase of the process Full Discharge commences and the Battery Calibration Wiza...

Page 351: ...charge process The Vivid i n Vivid q N system automatically shuts down at the end of this process the battery calibration procedure is now complete 7 The following morning re connect the AC power cabl...

Page 352: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 6 16 Section 6 2 LCD Display Adjustments and Calibration This page was intentionally left blank...

Page 353: ...e system problems and failures The Vivid i n Vivid q N ultrasound unit includes built in diagnostic tools that can be accessed from the scanner application Table 7 1 Contents in Chapter 7 Section Desc...

Page 354: ...can be achieved When the Back End is operating you can verify proper voltages and temperatures in the Front End and in the Power Supply To check the Front End parts the tests must follow a logical seq...

Page 355: ...lly updated information The Data Flow Map area on the right is designed to provide an instant visual representation of the system components the color illumination of graphic buttons light blue light...

Page 356: ...on page 7 10 Note When performing a specific test for an intermittent problem it is highly recommended to use the Loop option use the combo box or type the required number of loops as appropriate Loo...

Page 357: ...ormed Diagnostic Test Tree Displays a menu of diagnostic test sequences and tests available for selection Execute Displays the name of the currently selected diagnostic test or test sequence Controls...

Page 358: ...cluded in the currently selected test sequence Result Lists test results as either PASS or FAIL Loop Lists name of loop and the sub tests within the loop Date Date test was performed Tests Scroll Bar...

Page 359: ...one by one Erase Button Clears all test sequence log files for that specific currently selected test listed in the Tests window passed and failed tests It is not recommended to clear the test sequence...

Page 360: ...l amplitude first harmonic is displayed in the Report window When there is more than one report listed the first report is displayed by default Double clicking on another report for example Test signa...

Page 361: ...Left clicking on a currently selected channel marks a red square on the report graph and also evokes the display of the following information relative to that specific point Average Deviation standard...

Page 362: ...Test requires approximately 10 minutes for completion 7 2 4 1 1 Full System Test FE and BEP Performs a full sequence of complete Front End and Back End tests automatically that do not require user int...

Page 363: ...a complete test sequence and includes all the RFI board tests described below For test procedure instructions see the Full RFI Test section page 7 31 RFI FE Cache Test Tests the cache memory The cach...

Page 364: ...ructions see the System BF Digital Test section page 7 49 ADC Test The ADC Test is used to check that 1st harmonic 2nd harmonic 3rd harmonic and background noise are within pre defined limits In this...

Page 365: ...end Pass filters A test signal is shifted from the mixer frequency and then the test checks the filter response Attenuation of the filters in different frequencies should be within pre defined limits...

Page 366: ...upplied by the battery Source of the power being provided to the system i e external power via the AC DC or via the battery indicating the battery is either being charged or discharged For test proced...

Page 367: ...without Hard Drive This test performs a full system check as well as a Hard Disk scan The test sequence includes all automatic tests including the Hard Disk For test procedure instructions see the Fu...

Page 368: ...MX Logical Tests the MXX commands PAND PANDN POR and PXOR If errors are detected the failed commands are listed CPU MMX Shift Tests the MXX commands PSLL and PSRL If errors are detected the failed com...

Page 369: ...video buffer with 18 test patterns one pattern at a time These tests fill the entire screen with a single color Video Data Transfer Tests the graphics acceleration part of the video controller This te...

Page 370: ...e Network test logs as FAILED For test procedure instructions see the Network Test section page 7 130 Interactive Audio Test The Audio Test is performed to check for proper operation of the audio syst...

Page 371: ...pecially if the Vivid i n Vivid q N system shows signs of an intermittent problem As described in this section the testing sequence can be paused and restarted at any stage providing a quick and easy...

Page 372: ...iagnostic Test window trackball to Full System Test FE and BEP in the list of tests available for selection from the Diagnostic Test Tree see Figure 7 5 below and press Set The Full System Test FE and...

Page 373: ...ndicators in the Data Flow map corresponding to the various system components will be highlighted in the appropriate color to indicate the current test status as follows Green All recommended tests fo...

Page 374: ...N 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 22 Section 7 3 Performing Full System Diagnostics 8 If necessary use the scroll bars and or arrows to scroll through the list of Test results Figure 7 6 Vi...

Page 375: ...then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as descr...

Page 376: ...he proper sequence will help to derive the correct conclusions regarding the source of the problem As described in this section the testing sequence can be paused and restarted at any stage providing...

Page 377: ...ure 7 7 below and press Set The DC Offset Calibration name is displayed under Execute the relevant system components to be tested in this case all components are highlighted light blue in the Schemati...

Page 378: ...us as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Y...

Page 379: ...s were very close to the upper or lower limits repeat the DC Offset Calibration test again and review the newly acquired results If errors were identified results exceeding acceptable upper or lower l...

Page 380: ...t any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required num...

Page 381: ...Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the DC Offset Verification Test sequence is complete the Diagnostic Test window dis...

Page 382: ...e upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indica...

Page 383: ...tic Test Tree see Figure 7 11 below and press Set The Full RFI Test name is displayed under Execute the relevant system component to be tested is highlighted light blue in the Schematic View RFI as sh...

Page 384: ...s have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Fu...

Page 385: ...nd then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as de...

Page 386: ...probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number o...

Page 387: ...ave been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the RFI F...

Page 388: ...25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly...

Page 389: ...onnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the require...

Page 390: ...e button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the RFI Analog Circuits Test sequence is complete the Diagnostic...

Page 391: ...nd then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as de...

Page 392: ...the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failu...

Page 393: ...the RFI Demodulator Short Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing detai...

Page 394: ...25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly...

Page 395: ...probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of...

Page 396: ...formed When the upper progress bar reaches approximately midway in the Status area a loud high pitched sound will be heard coming from the speakers Note If no sound is heard it is necessary to perform...

Page 397: ...riginal problem persists contact your local On Line Center for assistance If errors were identified but the results were very close to the upper or lower limits perform calibration as described in DC...

Page 398: ...uld be touching the probe connectors 1 In the Diagnostic Test window trackball to TR32 BF Digital Test in the list of TR32 BF tests available for selection from the Diagnostic Test Tree see Figure 7 2...

Page 399: ...map corresponding to the various system components will be highlighted in the appropriate color to indicate the current test status as follows Green All recommended tests for these components have bee...

Page 400: ...ists contact your local On Line Center for assistance If errors were identified but the results were very close to the upper or lower limits perform calibration as described in DC Offset Calibration o...

Page 401: ...probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of...

Page 402: ...d 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the System BF Digital Test...

Page 403: ...on on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioni...

Page 404: ...obes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of L...

Page 405: ...s Set When the ADC Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of wh...

Page 406: ...list and the corresponding results are shown in the Report window below as shown in Figure 7 29 10 Use the tools provided to scroll through the channels and view the results as previously described 11...

Page 407: ...through the channels and view the results as previously described 14 When done trackball to the Close button and press Set to close the View Report window 15 Select the fourth report the report is hi...

Page 408: ...ower limits perform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all...

Page 409: ...ny probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number...

Page 410: ...ssed but not all recommended tests have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start...

Page 411: ...rresponding results are shown in the Report window below as shown in Figure 7 29 10 Use the tools provided to scroll through the channels and view the results as previously described 11 When done trac...

Page 412: ...faulty or that another board that supports the operation of that board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persi...

Page 413: ...ng the probe connectors 1 In the Diagnostic Test window trackball to ACW IQ Symmetry Test in the list of ACW tests available for selection from the Diagnostic Test Tree see Figure 7 32 below and press...

Page 414: ...ide the progress bar in the Status area As the testing sequence progresses the progress bar will advance to reflect the test progress Status indicators in the Data Flow map corresponding to the variou...

Page 415: ...r for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described...

Page 416: ...hat nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively...

Page 417: ...mmended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Start button and press Set...

Page 418: ...ea the dotted line represents the Average as shown in Figure 7 38 7 When done trackball to the Close button and press Set to close the View Report window 8 Select the second report Signal RMS the repo...

Page 419: ...nce If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Component...

Page 420: ...e that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternative...

Page 421: ...tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Start button and press Set When th...

Page 422: ...the dotted line represents the Average as shown in Figure 7 42 7 When done trackball to the Close button and press Set to close the View Report window 8 Select the second report Signal RMS the report...

Page 423: ...e If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components...

Page 424: ...ue in the Schematic View P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe con...

Page 425: ...ents have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have p...

Page 426: ...ain Test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Comp...

Page 427: ...iew P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Adjacent to...

Page 428: ...components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components...

Page 429: ...Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 49 9 Use the tools provided to scroll through the channels and view the results as previously desc...

Page 430: ...hrough the channels and view the results as previously described 13 When done trackball to the Close button and press Set to close the View Report window 14 Select the fourth report Signal SNR the rep...

Page 431: ...sistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Comp...

Page 432: ...tic View P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Adjace...

Page 433: ...omponents have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components h...

Page 434: ...7 25 If the test passed but the original problem persists contact your local On Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all...

Page 435: ...y probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number...

Page 436: ...been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Pencil RX...

Page 437: ...ble upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indi...

Page 438: ...s displayed under Execute all system components to be tested are highlighted light blue in the Schematic View as shown below 2 As indicated in the Special Instructions area disconnect any probes curre...

Page 439: ...these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 6 At any stage trackball to the P...

Page 440: ...ssed and that result values are within the pre defined Max and Min limits as shown in Figure 7 59 10 Select the second report Temperatures the report is highlighted in the Reports list and the corresp...

Page 441: ...test passed and that result values are within the pre defined Max and Min limits as shown in Figure 7 60 12 Select the third report Fans Speed the report is highlighted in the Reports list and the co...

Page 442: ...ssed and that result values are within the pre defined Max and Min limits as shown in Figure 7 59 14 Select the fourth report Battery Parameters the report is highlighted in the Reports list and the c...

Page 443: ...passed and that result values are within the pre defined Upper and Lower limits as shown in Figure 7 62 16 Select the fifth report Tx Voltages the report is highlighted in the Reports list and the cor...

Page 444: ...rform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables...

Page 445: ...lternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Fixed Voltages Test commences While the test proceeds t...

Page 446: ...7 65 Note For this test there is one report Voltages which is listed in the Reports window on the right By default this report is displayed in the Reports window below 7 Check that the test passed and...

Page 447: ...and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue troubl...

Page 448: ...ly you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Temperatures Test commences While the test proceeds the name of t...

Page 449: ...omplete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown i...

Page 450: ...en repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as describ...

Page 451: ...atively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Fans Speed Test commences While the test proceeds the name o...

Page 452: ...is one report Fans Status which is listed in the Reports window on the right By default this report is displayed in the Reports window below 7 Check that the test passed and that result values for eac...

Page 453: ...and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue troub...

Page 454: ...te the Loops checkbox and select the required number of Loops from the combo box 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed tes...

Page 455: ...o resume testing from the point where you paused trackball to the Start button and press Set When the Battery Parameters Test sequence is complete the Diagnostic Test window displays the Finished mess...

Page 456: ...that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the test fails again repeat the Fan Speed Test If the problem p...

Page 457: ...ate the Loops checkbox and select the required number of Loops from the combo box 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed te...

Page 458: ...6 To resume testing from the point where you paused trackball to the Start button and press Set When the Tx Voltages Test sequence is complete the Diagnostic Test window displays the Finished message...

Page 459: ...upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicate...

Page 460: ...on page 7 25 7 6 1 1 Extended Keyboard Test NOTE For a description of this test refer to the information in the Extended Keyboard Test on page 7 14 1 In the Diagnostic Test window trackball to Extende...

Page 461: ...fluence on any of the other ATGC sliders above or below i e all other sliders remain stationary 4 Repeat step 3 to test each of the other ATGC sliders 5 Check all keyboard keys by pressing each in tur...

Page 462: ...re activated illuminated green Patient CW Application etc alternatively check each one one by one individually b trackball to the Backlights button under Lights Control and press Set Make sure that al...

Page 463: ...ed as shown in Figure 7 76 below 17 Trackball to the Exit button and press Set to close the View Test Log dialog box 18 Proceed as follows If the test passed but the original problem persists contact...

Page 464: ...see Figure 7 77 below and press Set The ECG Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View BEP Peripherals Keyboard a...

Page 465: ...stics Troubleshooting 7 113 7 Make sure that the resolution scale is set to 1 Sec as shown in Figure 7 78 above 8 Trackball to the Open Device button and press Set The ECG signal is displayed in the D...

Page 466: ...nal is displayed on the screen as shown in Figure 7 79 10 When done trackball to the Exit button and press Set 11 When prompted by the following message select Yes or No as applicable 12 Proceed as fo...

Page 467: ...ription of this test refer to Full Back End Test without Hard Disk on page 7 15 1 In the Diagnostic Test window trackball to Full Back End Test w o Hard Disk Scan in the list of PC Doctor tests availa...

Page 468: ...ar on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you...

Page 469: ...t supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On...

Page 470: ...ended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so fa...

Page 471: ...or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists con...

Page 472: ...test status as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of...

Page 473: ...Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly a...

Page 474: ...oment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Hard Drive Full Test w o Scan commences...

Page 475: ...uired Linear Seek Random Seek Funnel Scan 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems we...

Page 476: ...d test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Full Memory Test commences...

Page 477: ...d Memory Pattern Test Memory Parity Test 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems wer...

Page 478: ...test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Full Video Test commences Wh...

Page 479: ...mory Video Data Transfer Video Text Output 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems w...

Page 480: ...led test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The PCI Test commences While...

Page 481: ...sults proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local...

Page 482: ...ailed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Network Test commences...

Page 483: ...results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your loc...

Page 484: ...ee Figure 7 96 below and press Set The Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View BEP Peripherals Keyboard as sho...

Page 485: ...er reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem per...

Page 486: ...in the list of Interactive PC Doctor tests available for selection from the Diagnostic Test Tree see Figure 7 99 below and press Set The Test name is displayed under Execute the relevant system compon...

Page 487: ...t 5 When in the displayed screen press any key to return to PC Dr for Windows Monitor Test screen 6 When done trackball to the Close button and press Set to exit the test When the Monitor Test sequenc...

Page 488: ...errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and F...

Page 489: ...ite ExC is only available in English at this time there is no multilingual platform available 7 7 1 1 Service Log In 1 From the Scanning Screen click the Global Service icon in the status bar at the b...

Page 490: ...I N AND VIVID Q N SERVICE MANUAL 7 138 Section 7 7 InSite ExC The lower section next page of the GEMS Service Home Page is illustrated in Figure 7 104 Figure 7 103 GEMS Service Home Page upper section...

Page 491: ...urity The service interface has various different security and user levels Users are only granted access to the tools they are authorized to use as described in the following table 7 7 2 Error Logs Pa...

Page 492: ...7 107 Click a category to navigate to the desired log category Figure 7 106 Log Viewer 7 7 2 1 Examples and Brief Description of Selected Categories When selecting System in the Logs drop down menu t...

Page 493: ...7 2 1 1 System Logs Displays all the system logs including errors and additional details see Figure 7 109 Figure 7 109 System Logs Note When a log category has been selected it is possible to select...

Page 494: ...form For example select Power or Temperature from the Log option on the Logs Viewer menu and then select Plot Log to display the results as shown in Figure 7 110 Figure 7 110 Plot Log Window Plot Page...

Page 495: ...s Viewer menu enables you to enter case sensitive text that you want to find from the log page that you are currently viewing as shown below Figure 7 112 Search for Text in Log or Page 7 7 2 1 4 Exit...

Page 496: ...d without the user s intervention 7 7 3 1 Diagnostics Page Overview The left pane of the Diagnostics page is shown in Figure 7 114 below There are five main categories Calibration Full System Test Fro...

Page 497: ...ally grey Upon completion of a diagnostic the background color changes to indicate the updated status either Red Fail Green Pass or Grey Abort Loop Count Field Enables you to enter the number of times...

Page 498: ...on on page 7 25 DC Offset Verification on page 7 28 7 7 3 2 2 Front End The Front End directory tree showing all the Front End diagnostic tests is shown in Figure 7 116 The Front End diagnostic tests...

Page 499: ...Test on page 7 12 ACW Matrix Test on page 7 13 ACW Delay Range Test on page 7 13 ACW Pre Mixer Gain Test on page 7 13 ACW BPF Test on page 7 13 ACW HPF Test on page 7 13 Pencil Pencil Rx Test on page...

Page 500: ...sts is shown in Figure 7 117 The BEP diagnostic tests are described in the following sections Extended Keyboard Test on page 7 14 ECG Test on page 7 14 PC Doctor Automatic Full Back End Test w o Hard...

Page 501: ...detected the address of the block in which the error was detected is shown During testing these diagnostics will also look for parity errors and other exceptions If encountered these are added to the...

Page 502: ...e or DSL modems The Network Communications Test checks if your PC s network card works and if your PC properly communicates over a network It makes a list of all the remote PCs on the network that the...

Page 503: ...Quality Page 7 7 5 Calibration Page Figure 7 119 Calibration Page Note This page is not populated in this version When the Calibration page is selected as shown in Figure 7 119 although not populated...

Page 504: ...en the Configuration page is selected you can access the Software Options Interface and install software options for the specific Vivid i n Vivid q N system No changes should be made in this page with...

Page 505: ...ptive Mode when the OLC or field engineer requests the user to invoke the Disruptive mode the user will be asked to confirm Disruptive Mode by clicking the Yes button as shown in Figure 7 122 Disrupti...

Page 506: ...E DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 154 Section 7 7 InSite ExC Disk Usage displays available disks and their utilization in the system see Figure 7 123 Figure 7 12...

Page 507: ...ION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 155 IP Configuration displays IP information required by the FE or OLC see Figure 7 124 Figure 7...

Page 508: ...Section 7 7 InSite ExC Network Status displays information concerning the system s communication with the network see Figure 7 125 Windows Services information concerning Windows applications invoked...

Page 509: ...er 7 Diagnostics Troubleshooting 7 157 User Accounts states those users registered on the system see Figure 7 127 Shared Resources indicates those system resources accessible from the network see Shar...

Page 510: ...1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 158 Section 7 7 InSite ExC System Shutdown see Figure 7 129 Disk Defragmenter to invoke Windows disk defragmenter utility see Figure 7 130 Figure 7 129 Syste...

Page 511: ...hooting 7 159 Gather Logs Utility accumulates the logs available in the system in D Export from where the OLC can download them see Figure 7 131 Image Viewer Utility enables the user to save images fo...

Page 512: ...OLC to compress images for downloading and deleting redundant files from D Export see Figure 7 133 Scanner Documentation Interface enables the system to read a DOC CD if present in a CD DVD device Fi...

Page 513: ...ID Q N SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 161 Virtual Console Observation this is started once the Disruptive mode has been invoked See Figure 7 135 Telnet Server Control this feat...

Page 514: ...pane of the Invoke Software screen Figure 7 136 above 3 From your PC open Grub Challenger MFC see Figure 7 137 4 Type the Challenge string in the Challenge field A Response code is displayed below the...

Page 515: ...ld in the pane on the right refer to Figure 7 136 The following screen Figure 7 138 displays a message stating that the Reload Setup was successful 7 Reboot the system as explained in the User Manual...

Page 516: ...N 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 164 Section 7 7 InSite ExC 7 7 7 2 Scanner Utilities Resource File Utility see Figure 7 139 7 7 7 3 DICOM Verify Refer to Figure 7 140 Figure 7 139 Resourc...

Page 517: ...replacement details and ordering information is displayed as described in Chapter 9 Renewal Parts Figure 7 141 Replacement Page Note This page is not populated in this version 7 7 9 PM Page When the...

Page 518: ...prompt line as shown in the example in Figure 7 143 The bookmark will allow users to place a time stamp where a particular problem has occurred while continuing with their current activity with minimu...

Page 519: ...drop down menu select the required export destination DVD CD or Memory Stick then click Save and Export A Zip file named logfile_ date _ time zip is copied onto the selected removable media Note Previ...

Page 520: ...xternal position as shown in Figure 7 145 below 2 Insert the blank DVD CD media into the DVD CD RW drive tray 3 On the Vivid i n Vivid q N system press the Config key 4 Select the Connectivity Tab and...

Page 521: ...ides recommendations for activating a specific Extensive Log category Table 7 4 Extensive Log Categories and Failure Scenarios Extensive Log Category Type of Failure Archive Loss of patient informatio...

Page 522: ...the appropriate checkboxes to activate the categories for which detailed analysis is required refer to the recommendations in Table 7 4 Note The impact of activating a checkbox would be to generate la...

Page 523: ...ting the Log File Ranges Note By default the log files may be very large as they are not limited by time date The user is able to limit the size timeframe of the log files by choosing the required set...

Page 524: ...press Ctrl E to eject the media Table 7 5 Attachment Categories and Recommendations for Use Category Type of Information and Recommended Use Diagnostic Reports Should be used whenever a diagnostic tes...

Page 525: ...GE ftp site NOTE Successful upload of logs to the GE ftp site can only be achieved when the designated PC is connected to the Internet and when all security measures such as firewalls taken at the lo...

Page 526: ...6 Inform the local GE personnel that you have uploaded files to the GE ftp site Note In the event that the upload process fails check your local network security settings to make sure that they did no...

Page 527: ...During this procedure a log file which contains information about the integrity of the database will automatically be created and stored in the Logs directory in the folder name Log If required this...

Page 528: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 176 Section 7 8 Automatic Error Log This page was intentionally left blank...

Page 529: ...3 Battery Replacement Procedure 8 5 Hard Disk Replacement Procedure 8 7 Control Panel and Keyboard Replacement Procedure 8 9 Rear Cover Latch Assembly Replacement Procedure 8 19 LCD Display Cover Hin...

Page 530: ...RVICE MANUAL 8 2 LCD Flex Cable Replacement Procedure 8 58 8 4 SafeLock Cart Components Replacement 8 61 8 5 Software Loading 8 97 8 6 Performing a Complete System Back up 8 141 8 7 Peripherals 8 148...

Page 531: ...handle 8 2 1 1 Tools Use the appropriate flat and Phillips type screw drivers as indicated in the Bearing Handle replacement procedure 8 2 1 2 Time Required 10 minutes 8 2 1 3 Preparations None 8 2 1...

Page 532: ...ng Handle in position using both hands to pull the Bearing Handle slightly apart in order to slot the ends into the handle holes on the Vivid i n casing 2 Carefully fasten each captive screw on either...

Page 533: ...ltrasound scanner 2 Close the Vivid i n Vivid q N LCD cover making sure both front latches are securely fastened 3 Carefully turn the Vivid i n Vivid q N over placing it face down 4 While holding the...

Page 534: ...id i n Vivid q N lying face down refer to Figure 8 5 on page 8 5 hold the unit steady with one hand while sliding the battery sideways into the battery compartment 2 Make sure the battery is properly...

Page 535: ...2 3 3 Preparation Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 2 3 4 Hard Disk Removal Procedure 1 Make sure the Vivid i n Vivid q N LCD Display cover is securely...

Page 536: ...ard disk module holder see Figure 8 7 on page 8 7 2 Return and fasten the two screws previously removed to secure the Hard Disk in position refer to Figure 8 6 on page 8 7 Note When a new hard disk ar...

Page 537: ...rd Removal Procedure 1 Remove the bearing handle as described in the Bearing Handle Removal Procedure on page 8 3 2 Place the Vivid i n Vivid q N portable ultrasound scanner face down and remove the b...

Page 538: ...4 NOTE When removing the Control Panel and Keyboard during other replacement procedures Step 6 will only be required if indicated 6 Lift the Control Panel and Keyboard up to gain access to the cable c...

Page 539: ...in the Bottom Assembly 3 Close the Vivid i n Vivid q N LCD Display cover making sure both front latches are securely fastened 4 Carefully turn the Vivid i n Vivid q Nover placing it face down 5 Return...

Page 540: ...iver between the keycap you wish to remove and its neighboring keycap 2 Use the screw driver to gently lever off the defective keycap until it is completely loosened from its base see Figure 8 16 belo...

Page 541: ...of the bracket that fasten the trackball assembly to the base of the Control Panel as shown in Figure 8 17 below 3 Disconnect the three trackball cable connectors from the trackball assembly carefully...

Page 542: ...nd remove the defective keycap see Figure 8 18 below 8 2 4 5 6 ATGC Slider Keycap Installation Procedure 1 Carefully place a new slider keycap in position on the slider base taking care to place the p...

Page 543: ...the defective knob see Figure 8 19 below 8 2 4 5 8 ATGC Rotary Knob Installation Procedure 1 Carefully place a new ATGC Rotary Knob in position on the keyboard taking care to place the plastic alignm...

Page 544: ...res as shown below NOTE In the event that the traces res file is not deleted prior to installing the new ECG Respiratory board the following error message will later be displayed 2 Shut down the Vivi...

Page 545: ...location on the base of the Control Panel 3 Re connect the ECG cable see Figure 8 21 above 4 Re connect the Flex Cable previously removed from the ECG Board to the new ECG Respiratory Board 5 Return...

Page 546: ...s shown in Figure 8 22 11 Proceed to perform the following functionality tests Isolated Patient Lead Source Leakage Lead to Ground on page 10 27 Isolated Patient Lead Source Leakage Lead to Lead on pa...

Page 547: ...nd Keyboard Removal Procedure on page 8 9 4 Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on page 8 27 5 Remove the LCD Display see the LCD Display Removal Proce...

Page 548: ...Return the Bearing Handle as described in the Bearing Handle Installation Procedure on page 8 4 7 Reconnect the AC adapter and all cables to the Vivid i n Vivid q N portable ultrasound scanner 8 Proc...

Page 549: ...nutes 8 2 6 3 Preparations Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 2 6 4 LCD Display Cover Hinges Removal Procedure 1 Remove the Control Panel and Keyboard as...

Page 550: ...n and remove the 2 screws that secure the left hinge to the LCD Display cover Figure 8 26 then carefully remove the hinge 6 Loosen and remove the 2 screws that secure the right hinge to the LCD Displa...

Page 551: ...over by raising and lowering it while exerting pressure to slide the LCD Display Cover until the hinge holes are aligned with the Bottom Assembly screw holes see right hinge in Figure 8 29 on page 8 2...

Page 552: ...e 8 22 Do not over tighten the screws 9 Reconnect the LCD flex cable to the BEP refer to Figure 8 24 on page 8 21 10 Refit the Control Panel and Keyboard as described in the Control Panel and Keyboard...

Page 553: ...he HVPS as described in the HVPS Removal Procedure on page 8 46 9 Remove the fans as described in the Fan Removal Procedure on page 8 49 10 Remove the LCD Display Frame as described in the LCD Display...

Page 554: ...31 10 Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedure on page 8 11 11 Return the hard disk as described in the Hard Disk Installation Proced...

Page 555: ...vid i n Vivid q N ultrasound unit as described in System Setup 8 2 8 4 LCD Display Frame Removal Procedure 1 With the LCD Display in the open upright position use a flat screwdriver to remove the scre...

Page 556: ...on return the LCD display frame to the correct position and using both hands press the frame under the securing clips on all four sides to fasten it against the LCD display Refer to Figure 8 32 on pag...

Page 557: ...n tightening screws do not overtighten or use excessive force CAUTION DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN AR...

Page 558: ...edure on page 8 3 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen and remove the four screws that secure the P M Board to...

Page 559: ...into position on the Bottom Assembly Make sure it is properly seated before returning and fastening the two screws previously removed Refer to Figure 8 35 and Figure 8 34 above 2 Return and fasten the...

Page 560: ...the board as shown in Figure 8 36 below CAUTION When removing the boards ESD may cause damage to a board Always have the ESD hand strip connected to the machine chassis and to your hand 8 3 2 1 Tools...

Page 561: ...the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen the eight screws that secure the TR32 and RFI Boards to the Bottom Assembly as sho...

Page 562: ...DC Power Supply as shown in Figure 8 40 below 7 Gently lift one side of the P M Board upwards and hinge it open to gain access to the RFI and TR32 Board assembly beneath as shown in Figure 8 42 below...

Page 563: ...rd assembly upwards and remove it from the Bottom Assembly as shown in Figure 8 42 below 9 Remove the RFI Board from the TR32 Boards separate by gently pulling them apart as shown in Figure 8 43 below...

Page 564: ...rnal Component Replacement Procedures 10 Separate the two TR32 Boards from each other as shown in Figure 8 44 below Note Replace the defective board as described in the appropriate steps of the follow...

Page 565: ...embly While performing this step gently lift one side of the P M Board upwards and hinge it open in order to insert the TR32 and RFI Board assembly underneath Refer to Figure 8 41 and Figure 8 42 on p...

Page 566: ...dure on page 8 3 3 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and place the scanner face down Remove the battery as described in the Battery Removal Procedure on page 8 5 4...

Page 567: ...e BEP to RFI Flex cable c Disconnect the HVPS to BEP power cable d Disconnect the Fan Speaker to BEP power cable 8 Unscrew and remove the four screws that secure the BEP to the Bottom Assembly as show...

Page 568: ...tab illustrated in Figure 8 48 that serves to seat the BEP firmly in position It is necessary to loosen and release this tab lift it up using a screw driver if necessary in order to facilitate complet...

Page 569: ...ow NOTE When removing the BEP for replacement with the SATA BEP disconnect the control panel keyboard cable Figure 8 50 from the parallel BEP you have just removed and keep this cable for installing o...

Page 570: ...tion The bracket is shown on the left in Figure 8 51 in the picture on the right the bracket is shown in position in the Vivid i n Vivid q N scanner Notice that the bracket sits on the PCMCIA slot fla...

Page 571: ...to fit the access codes already registered in the Hard Disk Only qualified personnel should perform these tasks 2 Check that the BEP is seated correctly then return and fasten the four screws previous...

Page 572: ...n page 8 118 Note If both the Back End and the Hard Disk are being replaced simultaneously continue to the next step If only the Back End is being replaced skip step 12 and proceed to step 13 below 12...

Page 573: ...up 8 3 4 4 BIOS Battery Removal Procedure 1 Remove the BEP as described in the BEP Removal Procedure on page 8 38 2 Take out the BIOS battery located on the bottom of the BEP as shown in Figure 8 54 8...

Page 574: ...HVPS Removal Procedure 1 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and place the scanner face down Remove the battery as described in the Battery Removal Procedure on page...

Page 575: ...Bottom Assembly as shown in Figure 8 57 below 6 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currentl...

Page 576: ...8 46 3 Reconnect each of the two Power Supply cables to the DC Power Supply HVPS to BEP and HVPS to RFI as shown in Figure 8 55 on page 8 46 4 Return the bearing handle as described in the Bearing Han...

Page 577: ...utes 8 3 6 3 Preparation Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 3 6 4 Fan Removal Procedure 1 Remove the bearing handle as described in the Bearing Handle Rem...

Page 578: ...upwards from the front and then the rear to remove it from the Bottom Assembly lay it down adjacent to the system ready for the next step as shown in Figure 8 60 below Figure 8 60 Removing the Fan Ass...

Page 579: ...an Assembly is seated in the correct position Return and fasten the two screws previously removed to secure the Fan Assembly to the Bottom Assembly as shown in Figure 8 58 on page 8 49 4 Reconnect the...

Page 580: ...bed in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen and remove the two screws on each of the speaker housings four in total 4 Disconnect the speaker cables from the fan assemb...

Page 581: ...as described in the Bearing Handle Removal Procedure on page 8 3 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Remove the LCD...

Page 582: ...8 3 Internal Component Replacement Procedures 6 Using both hands carefully lift the LCD Display upwards and remove it from the Rear Cover Latch Assembly as shown in Figure 8 64 below Take care not to...

Page 583: ...hen performing this step 2 Use both hands to re connect the LCD Flex ribbon cable to the BEP refer to Figure 8 63 on page 8 53 3 Return and fasten the four screws to secure the LCD Display to the Rear...

Page 584: ...3 Internal Component Replacement Procedures 9 From Windows Explorer go to the directory C Marlin Target bin see Figure 8 65 below 10 Double click on SetLCDType exe to run this utility The Set LCD Type...

Page 585: ...ion procedure proceed to calibrate the newly installed LCD Display by making any adjustments required to the Brightness Contrast and or Blue Tint as described in the LCD Display Adjustments and Calibr...

Page 586: ...lex Cable Removal Procedure 1 Disconnect the AC cable and battery 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Remove the LCD...

Page 587: ...nel and disconnect the flex cable see Figure 8 69 Figure 8 69 Removing Tape from LCD Rear Panel 2 Connect the new flex cable to LCD rear panel connector as shown in Figure 8 70 3 Connect the flex cabl...

Page 588: ...me as described in the LCD Display Frame Installation Procedure on page 8 28 9 Check that the LCD Display can be tilted upwards and downwards and that it holds its set position 10 Return the Control P...

Page 589: ...ment must not be operated within 15 feet of any magnetic equipment DANGER Electrical hazards exist at several points in the system Familiarize yourself with all hazardous voltages and high current lev...

Page 590: ...ock Cart Components Label Item Label Item 1 Top Shelf Assembly with handle 8 Rear Lower Cover 2 ECG Cable Protector 9 Bottom Cover Front 3 Peripheral Power Outlet 10 Bottom Cover Rear AC Assembly Cove...

Page 591: ...Cable Hook see page 8 65 AC Power Cable Bumper Protector see page 8 66 ECG Cable Protector see page 8 68 Lower Cover Rear see page 8 71 AC Distribution Assembly see page 8 72 Peripheral Power Cable se...

Page 592: ...er and while supporting the cover with one hand carefully remove the cover see Figure 8 73 8 4 3 4 Upper Cover Rear Installation Procedure 1 Return the cover to the correct position on the rear of the...

Page 593: ...m the Upper Cover rear unscrew the four AC Cable Hook retaining screws see Figure 8 74 and remove the cable hook holder 8 4 4 3 AC Cable Hook Installation Procedure 1 Position the AC Cable Hook on the...

Page 594: ...ctor upwards to release it from the shoulder screws refer to Figure 8 77 and Figure 8 75 2 Remove the Bumper Protector from the Cart 8 4 5 3 AC Power Cable Bumper Protector Installation Procedure NOTE...

Page 595: ...rotector onto the shoulder screws by aligning the mounting tabs ears with the screws then pushing the bumper downwards until it is seated securely See Figure 8 77 4 Proceed to perform the following fu...

Page 596: ...tector Installation Procedure If this is a first time installation of the ECG Cable Protector it is necessary to remove the two plastic guide holders from the Top Shelf Assembly as shown in Figure 8 7...

Page 597: ...the rear securing screw then the front securing screw 3 Return and fasten the rear securing screw Figure 8 80 then the front securing screw Figure 8 81 illustrates the ECG Cable Protector secured in...

Page 598: ...t 4 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 SafeLock Cart Chassis Current Leakage Test on page 10 25 Figure 8 81 ECG Cable Protector Secur...

Page 599: ...screws of the lower cover rear and while supporting the cover with one hand gently remove the cover see Figure 8 82 8 4 7 4 Lower Cover Rear Installation Procedure 1 Return the cover to the correct p...

Page 600: ...of the cart rear 2 Remove the Upper Cover Rear as described in the Upper Cover Rear Removal Procedure on page 8 64 3 Remove the Lower Cover Rear as described in the Lower Cover Rear Removal Procedure...

Page 601: ...asten the Ground screw to the AC Distribution Assembly Figure 8 84 above 3 Reconnect the peripheral power cable to the AC Distribution Assembly Figure 8 84 above 4 Secure all cables in position with n...

Page 602: ...r of the SafeLock Cart as described in Upper Cover Rear Removal Procedure on page 8 64 and Lower Cover Rear Removal Procedure on page 8 71 8 4 9 3 Peripheral Power Cable Removal Procedure 1 Disconnect...

Page 603: ...able and power socket in position in the SafeLock s Upper Cover front see item 4 in Figure 8 72 page 8 62 2 With the AC power socket plugged into the cart Figure 8 86 left picture connect the ground c...

Page 604: ...downwards Figure 8 88 Figure 8 88 Routing of Cable Downwards inside Rear Upper Cover 7 Fasten the Peripheral AC Outlet cable to the Main Harness cable as shown in Figure 8 89 Figure 8 89 Peripheral A...

Page 605: ...g it into a u shape and fasten together with tie wraps as shown in Figure 8 91 11 Check that the routing procedure has been implemented correctly lower the cart s top shelf and make sure that the cabl...

Page 606: ...sion 3 Chapter 8 section 8 4 3 4 13 Return the Upper Cover rear and Lower Cover rear of the SafeLock Cart as described in Upper Cover Rear Installation Procedure on page 8 64 and Lower Cover Rear Inst...

Page 607: ...rd as shown in Figure 8 93 4 From above the Top Shelf carefully push the Main Cable Harness Vivid i n Vivid q N power supply cable through the top shelf aperture and gently pull it through from below...

Page 608: ...pper Cover Rear Removal Procedure on page 8 64 3 Disconnect the P6 cable and USB connector from the USB board as shown in Figure 8 93 4 Remove the upper and lower bolts that attach the USB board to th...

Page 609: ...t Procedures 8 81 5 Reconnect the AC power cable to the AC input assembly at the base of the cart rear and turn ON power to the SafeLock Cart 6 Proceed to perform the following functionality tests Saf...

Page 610: ...re on page 8 64 3 Remove the Lower Cover Rear as described in the Lower Cover Rear Removal Procedure on page 8 71 4 Disconnect the Main Harness cable from the AC unit cut both tie clips and free the M...

Page 611: ...Gas Spring locking nut counter clockwise until it rises about a quarter way upwards on the thread 3 Insert the Gas Spring handle into the Gas Spring Housing Assembly until it is flush with the outer...

Page 612: ...round the Gas Spring and then reconnect it to the AC unit 8 Secure all cables in position with new cable ties 9 Return the Lower Cover Rear as described in the Lower Cover Rear Installation Procedure...

Page 613: ...the DVD Holder with one hand remove the four retaining screws as shown in Figure 8 98 2 Remove the DVD Holder from the base of the Top Shelf 8 4 13 3 DVD Holder Installation Procedure 1 Working from...

Page 614: ...ve the two LAN Holder retaining screws see Figure 8 99 and remove the LAN Holder 3 Remove the LAN Isolation Box 8 4 14 3 LAN Isolation Box and Holder Installation Procedure 1 Position the LAN Isolatio...

Page 615: ...8 64 4 Remove the Lower Cover Rear as described in the Upper Cover Rear Removal Procedure on page 8 64 5 Remove the USB board as described in the USB Board Removal Procedure on page 8 80 6 Remove the...

Page 616: ...8 82 4 Return the USB Box as described in the USB Board Installation Procedure on page 8 80 5 Fasten all cables in position using new cable ties 6 Return the Lower Cover Rear as described in the Lower...

Page 617: ...ve all peripherals from the Vivid i n Vivid q N portable ultrasound scanner 2 Working from the front of the SafeLock Cart and below the main shelf loosen and remove the three screws that secure the ha...

Page 618: ...id q N ultrasound unit System Setup 8 4 17 3 Probe Shelf Assembly Removal Procedure 1 Remove all peripherals from the Vivid i n Vivid q N portable ultrasound scanner 2 Working from the front of the Sa...

Page 619: ...hand return and fasten the two upper screws previously removed 3 Return and fasten the two lower screws one on either side making sure that the shelf is firmly secured in position 4 Place the require...

Page 620: ...2 Probe Cable Hook Removal Procedure 1 From below the Probe Shelf remove the Probe Cable Hook retaining screw see Figure 8 105 on page 8 92 2 Gently pull the Probe Cable Hook upwards to remove it 8 4...

Page 621: ...i n Vivid q N ultrasound unit as described in System Setup 8 4 19 3 Front Wheel Removal Procedure 1 Disconnect mains power from the AC adapter 2 Disconnect the AC adapter and all cables and peripheral...

Page 622: ...previously removed and make sure that wheel is secured firmly to the chassis 3 Stand the SafeLock Cart upright 4 Return the Vivid i n Vivid q N and all peripherals to their appropriate positions on th...

Page 623: ...Removal Procedure 1 Disconnect mains power from the AC adapter 2 Disconnect the AC adapter and all cables and peripherals from the Vivid i n Vivid q N portable ultrasound scanner 3 Remove the Vivid i...

Page 624: ...y removed see Figure 8 107 and make sure that wheel is secured firmly to the chassis 3 Stand the SafeLock Cart upright 4 Return the Vivid i n Vivid q N and all peripherals to their appropriate positio...

Page 625: ...eady installed In some cases it is necessary to re install the software for example after replacing the Back end Processor by following the software re installation procedure shown in Table 8 6 on pag...

Page 626: ...iate ordering number HCat as listed in Table 8 3 Table 8 3 Software BT10 to BT11 Upgrade Ordering Number HCat Description H45032BF Vi q German Upgr BT11 H45032BG Vi q Italian Upgr BT11 H45032BH Vi q S...

Page 627: ...11 H45032CH Vi q Serbian Upgr BT11 H45032CJ Vi q Korean Upgr BT11 H45032CK Vi q Estonian Upgr BT11 H45032CL Vi q Hebrew Upgr BT11 H45032CM Vi q Croatian Upgr BT11 H45032CN Vi q Slovenian Upgr BT11 H45...

Page 628: ...ireless network settings if present Trackball speed InSite ExC settings NOTE The following upgrade procedure is based on the Vivid q N system The procedure is the same as for the Vivid i n system 8 5...

Page 629: ...assword use lower case letters A message is displayed The DVD then begins the software installation process and a progress indicator is displayed as shown below This stage of the process takes approxi...

Page 630: ...version corresponds with the System SW version you have installed check against the CD label on the installation disc and that it was correctly installed 14 Restore all Presets as per user manual Not...

Page 631: ...ion 8 5 1 1 1 on page 8 100 Verify proper Connectivity functionality refer to Connectivity Setup on page 3 85 18 Proceed to perform the following functionality tests Software Configuration Checks on p...

Page 632: ...he Eject button of the DVD CD RW drive and remove any media if present 3 Insert the CD provided with the software kit into the DVD CD RW The DVD CD RW drive loads the software CD 4 Hold down the ON OF...

Page 633: ...ALT DELETE The system re boots 11 Follow the instructions on the screen clicking OK or Cancel when prompted The machine performs a number of reboots before completing the installation As the installa...

Page 634: ...10 seconds then re connect the battery and the AC adapter Boot up the system by pressing the On Off button If this is a new installation the system will prompt for the software licence Password as sho...

Page 635: ...rol panel A message is displayed prompting for system re boot 19 Click OK and re boot the system 20 If this is a new software version an adhesive Rating Plate is supplied with the software CD Adhere t...

Page 636: ...ersions 1 While booting press F2 Config Diag on alphanumeric keyboard 2 Type the password marlin04 3 Exit and save presets 4 Select Exit section and load Setup Defaults 5 Select Main section 6 Set cor...

Page 637: ...e Advanced menu and press Enter Advanced Chipset Control is highlighted BIOS Parameter Corresponding Figure Chipset Control Figure 8 110 on page 8 110 PCI PNP Configuration Figure 8 111 on page 8 110...

Page 638: ...ove in Figure 8 110 4 Press Esc to return to the Advanced menu home screen Figure 8 109 5 Scroll to PCI PNP Configuration and press Enter to open the PCI PNP Configuration screen 6 Set the parameters...

Page 639: ...in Figure 8 112 10 Press Esc to return to the Advanced menu home screen Figure 8 109 11 Scroll to I O Device Configuration and press Enter to open the I O Device Configuration screen 12 Set the parame...

Page 640: ...reen as shown above in Figure 8 114 16 Press Esc to return to the Advanced menu home screen Figure 8 109 17 Scroll to Hardware Monitor and press Enter to open the Hardware Monitor screen 18 Set the pa...

Page 641: ...g Settings as shown above in Figure 8 116 22 Press Esc to return to the Advanced menu home screen Figure 8 109 23 Scroll to Display Control and press Enter to open the Display Control screen 24 Set th...

Page 642: ...ection 8 5 Software Loading 26 Scroll to Miscellaneous and press Enter to open the Miscellaneous screen 27 Set the parameters in the Miscellaneous screen as shown above in Figure 8 118 28 Proceed to A...

Page 643: ...shown in this section 1 From the BIOS main menu see Figure 8 108 select the Advanced menu and press Enter 2 Select the Security menu and press Enter 3 Set the parameters in the Security screen as sho...

Page 644: ...he Power screen as shown above in Figure 8 120 7 Press Esc to return to the main BIOS screen Figure 8 108 on page 8 108 8 Select the Boot menu and press Enter 9 Set the parameters in the Boot paramete...

Page 645: ...ss F10 to Exit Saving Changes as shown in Figure 8 122 12 Reboot the system and check that the Supervisor PASSWORD was installed properly 13 Proceed to perform the following functionality tests Softwa...

Page 646: ...ble damage to the system 1 Make sure the Vivid i n Vivid q N scanner is connected to both battery and AC power 2 Insert the Vivid i n Vivid q N software installation CD in the DVD CD R and close the t...

Page 647: ...he BIOS update process When the BIOS Update process is complete the following message is displayed 7 Press any key to restart the system The boot up screen is displayed as shown in Figure 8 124 on pag...

Page 648: ...SERVICE MANUAL 8 120 Section 8 5 Software Loading 8 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on p...

Page 649: ...t the DVD CD RW drive to the system s USB and connect the power supply of the drive to power outlet 2 Place the GHOST CD inside the DVD CD RW Tray and close tray 3 Turn ON power to the system and allo...

Page 650: ...ace i e complete hard disk size minus 11GB for partitions C D and F IMPORTANT Do not stop the operation during this process 8 When complete continue with the software installation procedure Refer to t...

Page 651: ...ee page 8 139 it will be DELETED after performing the following procedure 1 Insert the Vivid i n Vivid q N software installation CD in the DVD CD R and close the tray 2 Hold down the ON OFF button on...

Page 652: ...he Service Tools Options menu opens 6 Do not select an option but press Ctrl C 7 When the message EMM386 is running Available memory is limited to 32MB is displayed click OK 8 When the message Termina...

Page 653: ...n click OK 12 Repeat step 10 and step 11 to select and delete each of the remaining partitions in turn including the partition type Extended The non partitioned hard disk will now be shown on the Part...

Page 654: ...d checkbox 26 When done click OK The User partition and allocated space is now listed in the Partition Magic Screen 27 With Unallocated selected in the Partition Magic Screen repeat step 15 and step 1...

Page 655: ...1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 8 Replacement Procedures 8 127 36 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on...

Page 656: ...f this is first installation format the hard disk as described in Formatting and Partitioning the Hard Disk Automatic Procedure on page 8 121 When done proceed to the next step below 6 Select option 1...

Page 657: ...pe in the password provided for the specific software and approve by clicking OK 10 Restart the system and commence scanning 11 Proceed to perform the following functionality tests Software Configurat...

Page 658: ...into the DVD CD RW drive and close the tray 4 Turn ON the Vivid i n Vivid q N system and allow it to boot up from the DVD The system powers ON and when system boot up from the DVD is complete the fol...

Page 659: ...Procedures 8 131 When done the Start Application screen is displayed as shown in Figure 8 127 6 Click on Install Software and follow the on screen instructions A warning message is displayed as shown...

Page 660: ...oading 7 Click OK The following message is displayed 8 Click OK to continue The following warning message is displayed WARNING DO NOT INTERRUPT THE INSTALLATION PROCESS Interruption will result in a n...

Page 661: ...T button 11 Press any key to re boot the system The system re boots into Normal Scanning mode 12 Press Config and then About The newly installed software version is now listed under the System Version...

Page 662: ...stallation Complete message is displayed refer to Figure 8 129 on page 8 133 7 Move the cursor over the displayed message window and press the SET button 8 Press any key to re boot the system The syst...

Page 663: ...as ADM The default password is ulsadm 4 Press the About button and activate the FwVersion tab verify that PMC version is 19 or above as shown in the example in Figure 8 131 below 5 Turn the system OF...

Page 664: ...low Figure 8 132 Run PMCbootloader exe A window is displayed showing the current PMC version and the path of the PMCJPS2 hex file 10 Click Burn The burning process will take approximately 30 seconds W...

Page 665: ...Wait approximately 5 seconds then boot up the system again Allow the system to boot up into Normal Scanning mode 15 Press Config The system will prompt you to login 16 Login as ADM The default passwo...

Page 666: ...be repeated Other indications of failure are that the system does not shut down automatically step 11 above or there is a strange pattern on the display In the event that failure has occurred perform...

Page 667: ...DVD and the PCMCIA 2 Press the Eject button of the DVD CD RW drive and remove any media if present The DVD CD RW drive tray ejects 3 Insert the CD provided with the software kit and press the Eject b...

Page 668: ...ns the software installation process and a progress indicator is displayed as shown below This stage of the process takes approximately 5 mins A message is then displayed on screen prompting you to pr...

Page 669: ...ll down menu 2 At the command prompt type Echoloader service and click OK The default scanning screen opens 8 6 1 2 2 Running Disk Management NOTE See Disk Management in the Vivid i n or Vivid q N Use...

Page 670: ...ssword NOTE The Operator Logon screen only appears after the first boot up of the system 4 Insert media e g CD DVD or connect to a network repository 5 Click the Admin tab at the bottom of the screen...

Page 671: ...nd follow the on screen wizard instructions The system stores the images 8 6 1 3 Running the Export Procedure 1 Insert the required storage media into the system 2 From the keyboard press the Patient...

Page 672: ...oft key at the bottom of the screen see Figure 8 140 NOTE If the Export button is not displayed click More consecutively until the Export button is displayed Figure 8 140 Export and More Soft Keys The...

Page 673: ...e required destination media Figure 8 142 and click OK The Export Patient List screen opens Figure 8 143 Figure 8 143 Export Patient Archive 7 Click the soft keys at the bottom of the screen consecuti...

Page 674: ...4 Click the Import soft key at the bottom of the screen see Figure 8 145 NOTE If the Import button is not displayed click More consecutively until the Import button is displayed NOTE A warning is disp...

Page 675: ...order as the system processes the actions see Figure 8 148 NOTE Wait while the system processes each action before pressing the next soft key 8 Press Alt E to eject the CD 9 Press 2D on the keyboard t...

Page 676: ...in Figure 8 149 below 8 7 1 Connecting and Removing Peripherals For detailed instructions on connecting peripherals to the Vivid i n Vivid q N portable ultrasound scanner refer to the following proce...

Page 677: ...removal of peripherals perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock cart Safe...

Page 678: ...up 8 7 2 4 External USB Hard Disk Removal Procedure NOTE Ensure that there is no activity on the drive before you turn it off If power is turned off while files are being written to the drive the file...

Page 679: ...at the rear of the disk unit 2 Connect the cable from the power supply to the power connector on the rear of the disk unit 3 Verify that the volume configuration is set to RAID 1 The External USB Har...

Page 680: ...ernal USB Hard Disk to the Vivid i n Vivid q N scanner at a time Follow these steps to set the Volume Name 1 Connect the USB cable from the External USB Hard Disk to a laptop or PC 2 Power up the Exte...

Page 681: ...nt Procedures 8 153 Figure 8 153 Removable Disk Shown under My Computer 4 Right click the drive name Removable Disk and select Rename from the menu Figure 8 154 Select Rename from the Drop down Menu 5...

Page 682: ...he Safely Remove Hardware tool in the Task Bar before continuing with the next step 9 Disconnect the USB cable from the laptop or PC 8 7 2 7 Connecting the External USB Hard Disk to the Vivid i n or V...

Page 683: ...iew of replacement parts for the Vivid i n and Vivid q N Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 2 9 3 Mechanical Hardware Parts 9...

Page 684: ...2 List of Abbreviations Assy Assembly BEP Back End Processor Ctrl Control FEP Front End Processor FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer l...

Page 685: ...6 2410799 BT 06 1 Y 4 HINGES concealed 45 MM 52 MM 2381805 2 BT 06 and BT 09 1 Y 5 Rubber inserts kit 2410804 BT 06 and BT 09 1 Y 6 Screw Kit FL200035 BT 06 and BT 09 1 Y 7 Internal Battery Pack 23789...

Page 686: ...board Assy 5376323 1 Y 2 Vi Operator Panel w o ECG Keyboard Assy BT 09 S2423357 1 Y 3 RoHS Keyboard Assy Vivid I BT 06 R2409723 1 Y 4 Trackball without bracket 2405164 2 BT 06 1 Y 5 Trackball S2423360...

Page 687: ...SEMBLY R2403248 4 1 Y 10 AC adaptor wo cable S2422706 BT 06 and BT 09 1 Y 11 Hard Disk Module 2376515 2 BT 06 and BT 09 Excludes software 1 Y 12 HARD DISK ASSY VI Q NEW SATA BB R2424374 Hard Disk Supp...

Page 688: ...PLAY MODULE ASSY Vivid q R2423283 1 Y 16 ECG Module Assy NR2421861 For Vivid q N 1 Y 17 TR32 v12 Board NS2423369 For Vivid q N 1 Y 18 Probe and MUX Assy N5420885 For Vivid q N 1 Y 19 RFI Board N241829...

Page 689: ...tion Box Cable Kit 2409727 BT 06 and BT 09 1 Y 2 LEAD WIRE KIT EUROPE YELLOW RED GREEN S2424558 For Vivid q N 1 Y 3 Cable ECG Cable Europe 2418832 2 BT 06 and BT 09 Replaces FL200039 1 Y 4 Cable Exter...

Page 690: ...tware Table 9 5 Software ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 Vivid q N BT11 System SW 6 1 9 and Application SW 11 2 0 build 40 FQ200002 Software for Vivid q N BT 11 1 Y 2 Service Software...

Page 691: ...T 09 1 Y 11 Probe 10S RS 2377040 H4000PF BT 06 and BT 09 1 Y 12 Probe M4S RS H40452LH 1 5 3 6 MHz 1 Y 13 Probe 9L RS H40442LL 4 0 13 0 MHz 1 Y 14 Probe 6T RS KN100093 H45531MZ BT 06 and BT 09 Must hav...

Page 692: ...On Bite Guard Adult H45511EE BT 06 and BT 09 1 Y 30 TEE Clip On Bite Guard Adult OR H45521CB BT 06 and BT 09 1 Y 31 TEE Conventional Bite Guard Pediatric H45521JG BT 06 and BT 09 1 Y 32 TEE Convention...

Page 693: ...eo Scan Converter H45531XE 2412280 BT 06 and BT 09 1 Y 7 Modem USB 5114058 BT 06 and BT 09 1 Y 8 USB Flash Card 1GB R2422220 BT 06 and BT 09 1 Y 9 USB Flash Card 512MB H45531XG 2411544 BT 06 and BT 09...

Page 694: ...le 9 8 SafeLock Cart Parts 1 of 2 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 AC Distribution Assy 220V 240V S2423305 1 Y 2 Handrest S2423250 1 Y 3 Lower Shelf Package assy S2422977 1 Y 4 Main Co...

Page 695: ...12 Vivid i Cart Gel Holder Cup S2422982 1 Y 13 Probe Cable Hooks S2422983 1 Y 14 AC Cable Hook S2422984 1 Y 15 Pencil Probe Holder Kit S2422985 1 Y 16 Pheripherals Power Out Cable S2422986 1 Y 17 Vi...

Page 696: ...obe and gel holder peripherals shelves hand rest and handles Y 2 Cable Clip M Cart KTZ220885 BT 06 and BT 09 Cable Clip M Cart 2 Y 3 Isolation Transformer M Cart KTZ220714 BT 06 and BT 09 Isolation Tr...

Page 697: ...ncludes Back end Assy FL200055 TR32 Board FL200069 1 Y 1 Spare Part Kit 1 for Vivid i n BT06 FL200087 BT 06 1 Y 2 Spare Part Kit 2 FL200115 BT 09 This kit includes Back end Assy FL200055 RFI Board FL2...

Page 698: ...1 Y 9 FANS ASSY VIVID I R2376696 1 Y 10 ECG BOARD V3 Vivid i q N Natural N2419020 1 Y 11 Spare Part Kit 3 FL200111 This kit includes Keyboard assy FL200059 Fans and Speakers Kit FL200040 Cable and Fle...

Page 699: ...oc CD FL200028 SW CD FL200103 SW ver 1 36 19 Appl SW 6 1 0 build 122 FL200107 Application SW Patch for Vivid i n and Vivid q N 1 Y 17 VI COVER KIT BT09 S2423361 This kit includes BOTTOM BASIC ASSY VIV...

Page 700: ...BOTTOM BASIC ASSY VIVID I DISPLAY RIGHT LATCH ASSY VIVID I DISPLAY LEFT LATCH ASSY VIVID I REAR DISPLAY COVER VIVID I DISPLAY FRONT FRAME ASSY DISPLAY SCREW CAP VIVID I MAGNET PERMANENT 4X20MM DISPLAY...

Page 701: ...ual Turkish FQ092006 1 N 6 BT 11 Vivid q N Reference manual English FQ092007 1 N 7 Vivid i n SafeLock Cart User manual EN FR TR FL092077 English French and Turkish 1 N 8 Vivid i n Cart User manual EN...

Page 702: ...t Manuals BT 06 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 Vivid i n Service Manual FL091021 This Manual 1 N 2 Vivid i n Unpack Pack Procedure FL094023 1 N 3 BT06 Vivid i n User manual English F...

Page 703: ...ound scanner and its peripherals These procedures are designed to protect your investment in the system and are intended to maintain the quality of system performance After delivery and installation o...

Page 704: ...aker before you remove any parts Proceed with caution whenever power is ON and covers are removed DANGER Do not pull out or insert circuit boards while mains power to the system is ON CAUTION Practice...

Page 705: ...tion from organizations such as the American College of Radiology USA it is the customer s responsibility to have a Quality Assurance program in place for each scanner The program must be directed by...

Page 706: ...vels of safety dependability and system performance Your GE Medical Systems Service Representative has an in depth knowledge of your Vivid i n Vivid q N ultrasound scanning system and can best provide...

Page 707: ...kage Checks Also after corrective maintenance or as required by your facility s QA program Surface Probe Current Leakage Checks Also after corrective maintenance or as required by your facility s QA p...

Page 708: ...cuts or abrasions Equipment for loose or missing hardware The keyboard for defects Brakes for defective operation DANGER To avoid electrical shock hazard do not remove the panels or covers from the ma...

Page 709: ...atic Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for electrical tests SVHS VCR Cassette E7010GG E7010GF 60 minutes 120 minutes SVHS VCR...

Page 710: ...N SERVICE MANUAL 10 8 Section 10 4 Tools Required System Service Code Required in order to enter system diagnostics User Manual Current revision Service Manual Current revision Table 10 3 Overview of...

Page 711: ...customer if they have any problems or questions about the equipment 2 Paperwork Fill in the appropriate details at the top of the Vivid i n Vivid q N System Inspection Certificate see page 10 34 Note...

Page 712: ...ic system controls that affect this mode of operation 5 Applicable Software Optionsa a Some software may be considered as standard depending upon system model configuration Verify the basic operation...

Page 713: ...3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 InSite iL...

Page 714: ...nt CAUTION When performing Cleaning Procedures to prevent the risk of system damage always observe the following precautions Use only cleaning materials and solutions such as Isopropyl Alcohol IPA as...

Page 715: ...t contain impurities Wring out as much of the liquid as possible then wipe the LCD Display again Do not let any liquid drip into the computer Be sure to dry the LCD Display before closing the cover 3...

Page 716: ...l alcohol and water Never allow liquid to flow into the keyboard Use a 70 30 mixture of isopropyl alcohol and water Staff users doctors Monitor display surface Weekly In event of contamination clean i...

Page 717: ...ge 10 13 Inspect the Probe Holders for cracks or damage 7 SafeLock Cart Check all four SafeLock Cart wheels and castors for wear Check the brakes on each of the two front wheels Make sure that pressin...

Page 718: ...anning 13 Peripherals Check and clean the peripherals according to the manufacturer s directions To prevent EMI or system overheating dress the peripheral cables inside the peripheral cover 14 Externa...

Page 719: ...a visual check of the probe pins and system sockets before plugging in a probe Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the prob...

Page 720: ...m sockets before plugging in a probe When handling probes always observe the precautions listed in Probe Handling on page 10 17 The TEE and Interoperative probes often have special usage consideration...

Page 721: ...rasive soap 3 Follow the manufacturer s cold sterilization instructions to disinfect the probe including concentration and soaking time 10 minutes is a typical disinfecting period 4 Follow step 3 for...

Page 722: ...must be properly removed from any part or equipment prior to shipment GEMS employees as well as customers are responsible for ensuring that parts equipment have been properly decontaminated prior to...

Page 723: ...leaning and disinfecting procedures before handling the equipment Test the system peripherals and probes for current leakage WARNING EXCESSIVE CURRENT LEAKAGE CAN CAUSE FATAL INJURY EXCESSIVE CURRENT...

Page 724: ...ntry Normal Condition Open Ground Reverse Polarity Open Neutral Mains Applieda a Mains Applied is the Sink Leakage Test In this test mains supply voltage is supplied to the part or equipment to determ...

Page 725: ...et tester can detect the condition in which the Neutral grounded supply conductor and the Grounding protective earth conductor have been reversed If later tests indicate high current leakages this sho...

Page 726: ...ower cable should be tested with a substitute Safety Tester 1 Connect one of the Ohmmeter probes to the tested AC wall outlet cable side Ground pin 2 Using the other Ohmmeter probe touch the exposed m...

Page 727: ...Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch Otherwise the unit may be damaged 10 7 5 2 Generic Procedure The test verifies the isolati...

Page 728: ...wn in Table 10 14 Record all data on the Vivid i n Vivid q N System Inspection Certificate see page 10 34 Table 10 17 Typical Data Sheet for Chassis Source Current Leakage Test Unit Power Tester Polar...

Page 729: ...wered ON Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch 10 7 6 2 Generic Procedure Measurements should be made with the Ground Open and Cl...

Page 730: ...solated Patient Lead Sink Leakage Isolation Test NOTE Refer to the procedure in the IEC 60601 1 CAUTION Line voltage is applied to the ECG leads during this test To avoid the risk of possible electric...

Page 731: ...Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Current Leakage Test and Patient Lead to Lead Current Leakage Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 19 Maximu...

Page 732: ...ould be made with the Ground OPEN and CLOSED with power line polarity Normal and Reversed and with the unit OFF and ON For each combination the probe must be active to find the worst case condition NO...

Page 733: ...TERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area of the probe into the...

Page 734: ...ge 10 34 CAUTION When power to the unit is ON never switch the Polarity and the status of Neutral Be sure to turn OFF power to the unit before switching them using the POLARITY switch and or the NEUTR...

Page 735: ...ith the probe or the scanner NOTE Each probe will have some amount of leakage depending on its design Small variations in probe current leakages are normal from probe to probe Other variations will re...

Page 736: ...requency Scan Format Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe 8 Frequency Scan Format Model Nu...

Page 737: ...Caster Chassis Source Current Leakage CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Current Leakage Periphe...

Page 738: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 10 36 Section 10 9 Vivid i n Vivid q N Inspection Certificates This page was intentionally left blank...

Page 739: ......

Page 740: ...2011 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...

Reviews: