Chapter 7 - Diagnostics/Troubleshooting
7-9
NOTE:
Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also evokes
the display of the following information relative to that specific point: Average, Deviation (standard
deviation from the Average), Max, and Min.
The signal should be within the limits marked in the white area (the dotted line represents the Average),
as shown in
Figure 7-4 ADC Test - Viewing Report (Test Signal Amplitude 2nd Harmonic)