Chapter 3 - System Setup
3-65
Section 3-8
Mounting the Vivid
i
n/ Vivid q N on the SafeLock Cart (optional)
NOTE:
The Vivid i n and Vivid q N portable ultrasound scanner may be mounted on the SafeLock Cart
while power to the system is either ON or OFF. When mounting the unit on the cart, it may be
necessary to adjust the locking mechanism, as described in
3-8-2 "Adjusting the SafeLock Cart
Locking Mechanism" on page 3-68
.
3-8-1
Mounting the Vivid
i
n and Vivid q N on the SafeLock Cart
1.) Working from the front of the SafeLock Cart, place the Vivid
i
n and Vivid q N scanner flat down on
the main shelf (towards the front) then slide it backwards between the sliding guides on either side
of the shelf, as shown below.
NOTICE
Only the SafeLock Cart supplied by GE Ultrasound and/or any medical-grade cart which has been
approved as suitable for use with the system by GE Ultrasound should be used with the Vivid
i
n and
Vivid q N portable ultrasound scanner.
NOTICE
The Vivid
i
n and Vivid q N system and the SafeLock Cart should be operated at a distance of more
than 15 feet from any magnetic field.
NOTICE
When scanning with the Vivid
i
n/ Vivid q N portable ultrasound scanner mounted on the SafeLock Cart,
the Vivid
i
n/ Vivid q N should be connected to the Cart via the USB port.
WARNING
WARNING
Before mounting the Vivid i n and Vivid q N portable ultrasound scanner on the
SafeLock Cart make sure there are no foreign objects on the main shelf surface or in
front of the locking tooth.
Figure 3-62 Placing the Vivid
i
n and Vivid q N on the SafeLock Cart Main Shelf
Sliding Guides