Error Codes and Diagnostics
6-3
March 2004
E6: Cycle Skip, Accel.
Problem:
The acceleration exceeds the limits programmed in the
Error Limits
option of the
User Program
.
Cause:
This condition is usually caused by poor flow conditions
or improper transducer alignment.
Action:
Refer to the sections on
Fluid and Pipe Problems
(page
6-6) and on
Transducer Problems
(page 6-8) to correct
any problems.
E7: Analog Out Error
Problem:
The current setting is outside the programmed limits.
Cause:
The calculated output value exceeds the programmed
limits.
Action:
Verify that the 4-20 loop configuration base and span
settings are correct for your process.
E30: Channel Disabled
Problem:
The channel is not available.
Cause:
The channel has been turned off.
Action:
Enter the
PROGRAM
menu and enable the channel (see
page 2-4).
E31: Invalid Calibration
Problem:
The calibration is invalid.
Cause:
Improper calibration for the application has been entered.
Action:
Consult GE Panametrics.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...