Error Codes and Diagnostics
6-1
March 2004
Introduction
The Model UTX878 flow transmitter is a reliable, easy to maintain
instrument. When properly installed and operated, as described in
Chapter 1,
Installation
, the meter provides accurate flow rate
measurements with minimal user intervention. However, if a problem
should arise with the electronics enclosure or transducers, this chapter
explains how to troubleshoot the Model UTX878. Indications of a
possible problem include:
•
display of an error message on the LCD screen
•
erratic flow readings
•
readings of doubtful accuracy (i.e., readings that are not consistent
with readings from another flow measuring device connected to
the same process).
If any of the above conditions occurs, proceed with the instructions
presented in this chapter.
Error Codes
If a problem occurs with the electronics or transducers, a built-in error
code message system greatly simplifies the troubleshooting process.
All of the possible UTX878 error code messages are discussed in this
chapter, along with the possible causes and the recommended actions.
When an error code is generated, it will appear in the lower right
corner of the LCD screen, as discussed in Chapter 3.
If an error message appears on the display screen during operation of
the Model UTX878, refer to the appropriate section of this chapter for
instructions on how to proceed.
E0: No Error
Problem:
No error condition currently exists.
Cause:
This message appears briefly to confirm that the response
to another error message has corrected the problem.
Action:
No action is required.
E1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the
limits entered via the
User Program
.
Cause:
Poor signal strength may be caused by a defective cable,
a flowcell problem, a defective transducer or a problem
in the electronics console. A signal that exceeds the
programmed limits is probably caused by the entry of an
improper value in the Error Limits option of the
User
Program
.
Action:
Using the procedures in Chapter 1, check the components
listed above. Also, check the value entered into the
Signal Strength
Error Limits
option, as described on page
2-19.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...