MA-014 - TRANSFIX DGA 500 User Guide – Rev 1.1
13-Oct-15
Page 10 of 28
5.3
Battery
The product uses a non-rechargeable lithium coin cell battery (Panasonic® CR2450 3 V
620 mAh) as shown in Figure 5—3. In the event that the battery needs to be replaced,
data from the product must be backed up. Failure to do so may result in historical data
loss.
Figure 5—3: Coin cell battery
The following steps describe how to change the battery:
1.
Back up the product data – contact your GE representative.
2.
Open the inner door to locate the battery on the system board (as
shown in Figure 5—3).
3.
Slide the battery out of its housing.
4.
Replace with a new Panasonic CR2450 3 V 620 mAh coin cell.
5.
Close the inner door.
There is a danger of a new battery exploding if installed incorrectly.
Dispose of the used battery in accordance with local regulations — not in a fire
or with household waste. Contact your local waste disposal agency for the
address of the nearest battery deposit site. Perchlorate material — special
handling may apply. See: