MAINTENANCE: Cleaning
Revision J
Solar 7000/8000/View Patient Monitor
3-7
414993-001
Cleaning Inside the
Solar 7000 Color
Monitor
The Solar 7000 color monitor uses a forced-air cooling system that draws
appreciable quantities of air through the unit. As a result there can be a
buildup of lint and other debris inside the unit. Accumulations of lint and
debris can lead to thermal failures or short-circuit failures if not removed
regularly.
It is recommended that the inside of all Solar monitors should be cleaned
every 12 months.
To clean the inside of the monitor, follow this procedure:
1.
Power down the monitor. Disconnect the monitor from the power
source and the network.
2.
Remove the screws attaching the housing to the chassis at the rear of
the unit.
3.
Pull the chassis out of the housing.
4.
Using a source of clean, dry compressed air blow all lint and debris
from between the circuit boards and the other assemblies.
CAUTION
Do not use a vacuum device unless suitable static-control
procedures are followed. Many vacuum systems generate
appreciable static electricity which could damage
semiconductor circuits in the monitor.
5.
Make sure all circuit boards are seated firmly. Make sure all cables
are connected.
6.
Install the chassis back into the housing and attach with screws.
7.
Perform leakage tests.
8.
Connect the monitor to the network and to the power source.
9.
Apply power to the monitor and verify operation.
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......