DIMENSIONS OF THE MAIN REPLACEMENT PARTS
EQUIPMENT
DIMENSIONS
LxWxH
WEIGHT
NOTE
mm
in
kg
lbs
Replacement BRM gradient coil assembly
on a shipping cradle/cart
902x2444x1420
35.5x96.1x55.9
1491
3287
Initial gradient coil assembly is shipped
installed in the magnet.
Shipping/installation cart is used to install
re-placement coil assembly only.
The weight bearing structure of the site should support any additional weight of the main replacement parts
occurring during maintenance of the magnet, throughout the whole lifecycle of the MR.
Front view of the BRM Gradient
Side view of the BRM Gradient
2444
[96.2 in]
902
[35.5 in]
ROUTING
The customer is solely liable for routing of components from dock to final site.
GE
must be able to move system components in or out with no need to uncrate or disassemble any of the
components. The entire passageway must be cleared, adequately lighted and free from dust.
The floor and it surfacing must be able to withstand the live load of components and handling equipment.
Floor surfacing must be continuous.
The customer must protect any fragile flooring surfaces.
MINIMUM SPECIFICATIONS FOR MAGNET ROUTING
Floor must be able to withstand a moving load of 5320 daN
Minimum Opening: Height: 2.5m (98.5 in), width: 2.5m (98.5 in)
Maximum slope: 30 degree
STRAIGHT PATH
(Rigging wheels required)
PATH WITH 90 DEGREE TURN
INSTALLATION AND DELIVERY ACCEPTANCE
A survey of the site established by the customer and
GE
Healthcare will
make the decision for the delivery time.
This survey of the site (a form is made available by
GE
) is only to check if the
apparent conditions of the site allow the equipment to be delivered.
If the site is not ready,
GE
can delay the delivery time.
DELIVERY
Recommended minimum opening for side (wall) delivery : 2500 (width) x 2500 (height)
FRONT VIEW OF MAGNET
RIGHT SIDE VIEW OF MAGNET
2467
[97.1 in]
2388
[94 in]
1953
[76.9 in]
2500
[98.4 in]
2700
[106.3 in]
This is only a partial list of items required for delivery of the magnet. For a complete checklist refer to the
pre-installation manual referenced on cover sheet.
24/7 chilled water and 480v power for shield/cryo cooler
24/7 120v power for the magnet monitor
Phone lines for magnet monitoring and emergency use
Magnet room exhaust fan
Cryogen venting (if roof hatch, completed within 24 hrs)
Magnet anchors installed and tested
CRITICAL ITEMS FOR MAGNET DELIVERY
Date
Rev
/24
EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG
A
03/Aug/2018
SIGNA EXPLORER /SIGNA CREATOR
Typical
A8 - Delivery
10