GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG M028 - E24 084/008 Error Troubleshooting
Page no. 1087
Chapter 9
JC-TSG-M-028.fm
5
PREVENTIVE ACTIONS THAT MAY REDUCE TUBE ARCING
During tube lifecycle, it is normal to observe tube arcing, which generates exposure spit/interruption.
Note:
Do not plan to replace the tube if tube arcing occurrence is low (1 tube arcing every 400 exams).
Nevertheless, it is recommended to perform the calibrations below.
Tube arcing may:
•
Occur more frequently for brand new and old tubes
•
Increase with tube usage as a result of wrong or inaccurate generator calibrations.
It may be possible that
Job Card TSG M022 - HV Troubleshooting on page 1403
does not report any
corrective action. Intermittent arcing failure may result from a calibration or a heater current that is not
adjusted to the current tube state. For information on heater current regulation, see
Since tube usage may change the tube characteristics, it may be useful to perform all the calibrations
that are required at tube change. See the
Job Card D/R A261 - X-ray Tube
.
Perform the following calibrations:
Considering that tube arcing may result from a tube instability regarding its heater state, if the generator/
tube calibration doesn't reduce the tube arcing occurrence, you may ask the operator to perform the
heating procedure for the selected focal spot size (large or small) when the system has not been used
Calibrations - Checks
Job Card
Acquisition Image Chain
Bias Calibration
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
Heating Current Calibration
Job Card CAL A001 - Heater Current Scale Factor
mAs X-ray Calibration
Job Card CAL A012 - X-Ray Tube Heater Current
Tube Tilt Calibration
Job Card CAL A027 - Tube Tilt Calibration
Collimator Calibration
Job Card CAL A029 - Collimator Format Calibration
Collimation Check
Job Card ELE A028 - Collimation Checks
If Collimation Check not OK, repeat Collimator Calibration
Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Gain calibration
Job Card CAL A043 - Detector Gain Calibration
flat field tests
Job Card ELE A032 – IQ Tools Flat Field Test
If IQ Tests not OK, repeat all calibrations from Bad Pixel.
Half Value Layer
Job Card CAL A046 - Half-Value Layer Measurement
Conversion Factor
Job Card CAL A045 - Conversion Factor Measurement
Use the same dosimeter as that used for CAL A046
AOP: Automatic Optimization of Parameters
AOP Calibration
Job Card CAL A051 - AOP Calibration
mAs non linearity
Job Card CAL A053 - mAs Non-linearity Calibration
AOP/SNR Check
Job Card ELE A038 - AOP Mode and SNR Check
If AOP/SNR Test not OK, repeat all calibrations from AOP Calibration