RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–42
REV 21
3–10 GERMAN ROM
Purpose
Installation of the German ROM on the MST Board.
Supply Parts List :
There are two kinds of German ROM. The one used depends on the MST Boards Parts number. Install German
EPROMs up carefully. See the Table below.
NO
Version (Style)
MST-E Assy
German ROM
LOCATION
1
Version 4 (style D)
P9514ZZ–00
P9514QV
U9A
P9514QW
U11A
2
Version 5 (style E)
2113953–3
2113971–3
U9A
2113972–3
U11A
Man Power
1.5~4 h 1 SVE Engineer (Depending on whether the system has already been used by customer or not; refer to Il-
lustration 3–30).
Required Tool
Standard Tolls
Procedure
This German ROM Installation Instruction consists of the following sections.
SECTION (A) : Data Recording
SECTION (B) : Installation of the German ROM
SECTION (C) : System Initialization
SECTION (D) : Confirm the System is German Software.
SECTION (E) : Preset All Parameters reset by Initialization
If the system has already been used by the customer, it is necessary to record the user’s defined preset parameters
prior to changing the EPROM(s). In this case, refer to the Operator Manual to input the system data.