5
Pressure Rating RHM 30
The maximum pressure (pmax) of a sensor is determined by its weakest part. The weakest part can be the measuring
loops (pmax indicated below) or the construction type (pmax indicated in the Basic Order Code section, last page) or the
selected flanges / fittings (for pmax please see respective standard).
pmax of P1 measuring loops
M3 optional material - 2.4602 (Alloy C22)
OD x WT 33.4 x 2.8 mm (1.32 x 0.11 in)
bar
°C
psi
°F
271
50
3931
122
239
120
3466
248
204
210
2959
410
170
350
2466
662
pmax of P1 measuring loops, standard
M1 standard material - 1.4571 (316Ti)
OD x WT 33.7 x 2 mm (1.33 x 0.08 in)
bar
°C
psi
°F
137
50
1987
122
123
120
1784
248
106
210
1537
410
89
350
1291
662
pmax of P2 measuring loops
M1 standard material - 1.4571 (316Ti)
OD x WT 33.7 x 4.55 mm (1.33 x 0.18 in)
bar
°C
psi
°F
334
50
4844
122
299
120
4337
248
258
210
3742
410
217
350
3147
662
pmax of P3 measuring loops
M1 standard material - 1.4571 (316Ti)
OD x WT 33.7 x 6.35 mm (1.33 x 0.25 in)
bar
°C
psi
°F
490
50
7107
122
438
120
6353
248
378
210
5482
410
319
350
4627
662