Managing patients and monitoring
●
Patient is not available
appears: another Hub has already been assigned to this
patient, or the patient has been discharged remotely. Check the patient and Hub
assignments and patient admission status on the Central Viewer.
●
No wireless network
appears: there is no wireless network connection. Make sure
that the Hub is within the network coverage range. If that does not help, contact
service personnel.
●
No connection to patient monitoring services
appears: the Hub is not connected
to the Core Services. Make sure that the Hub is within the network coverage range.
If that does not help, contact service personnel.
●
No license available
appears: there are no available monitoring licenses. Unassign
one Hub or wait until a license becomes available again.
Attaching measurement devices to the patient
1.
Attach the Wearable SpO
2
Parameter and/or the Wearable RR Parameter to the
patient:
For more detailed information on each measurement, refer to the parameter-specific
instructions.
Pairing the first sensor
You need to create a wireless connection between the Hub and sensor to start
monitoring a patient. This is done by pairing them.
To pair the sensor, follow the animation on the Hub screen.
5815148-01
Portrait
™
Mobile Monitoring Solution
67
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...